Drugs & medicalClass II

GE Healthcare, LLC recalls GE Centricity Universal Viewer Zero Footprint. To view, communicate, process, and display Medical images and data within a computer…

Potential to display inaccurate measurements on images in Centricity Universal Viewer Zero Footprint Client (ZFP)

Read the official notice →
Date
2022-07-13
Country
United States
Brand
GE Healthcare
Product
GE Centricity Universal Viewer Zero Footprint. To view, communicate, process, and display Medical images and data within a computer…
Classification
Class II
Units
Worldwide distribution.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More GE Healthcare recalls

  1. 2022-07-27
  2. US
    Drugs & medicalClass II

    GE Healthcare, LLC recalls SIGNA Premier magnetic resonance scanner, model 5748519.

    Under rare conditions, SIGNA Premier gradient coils could result in elevated acoustic noise during scanning. After prolonged periods of elevated acoustic noise, this could potentially lead to hearing loss.

    FDA · DeviceGE Healthcare, LLCWorldwide distribution - US Nationwide distribution in the states of AL, AZ,…Official notice →
  3. US
    Drugs & medicalClass II

    GE Healthcare, LLC recalls GE Healthcare Avance CS2, Avance CS2 Pro Anesthesia Systems, Part #1009-9050-000; and Aisys CS2 Anesthesia Systems, Part # 1011-9050-000.

    The base of the anesthesia system can have a crack and could potentially fracture resulting in a tip or overbalance of the anesthesia system. This can result in potential injury if it falls on a person.

    FDA · DeviceGE Healthcare, LLCDistribution was made to AZ, CA, CO, CT, FL, IA, ID, IL, IN, KY, LA, MA, MD,…Official notice →
  4. US
    Drugs & medicalClass I

    GE Healthcare, LLC recalls GE Healthcare Aisys CS2 Anesthesia System, Part number 1011-9050-000; and Avance CS2 and Avance CS2 Pro Anesthesia System, Part number…

    There is a potential reversal of the O2 and air cylinder pressure transducer connections.

    FDA · DeviceGE Healthcare, LLCWorldwide Distribution. US nationwide, Australia, Austria, Belgium, Bolivia,…Official notice →
  5. 2022-07-06
  6. US
    Drugs & medicalClass II

    GE Healthcare, LLC recalls Back up batteries in Avance CS2 and Avance CS2 Pro anesthesia devices manufactured on or after April 1, 2019

    Backup batteries can fail earlier than their estimated life.

    FDA · DeviceGE Healthcare, LLCWorldwide distribution - US Nationwide and the countries of Algeria,…Official notice →
  7. 2022-06-22
  8. US
    Drugs & medicalClass II

    GE Healthcare, LLC recalls CARESCAPE Central Station (CSCS) V2, Model numbers 2082278-001 and 2082279-001. Update 3/10/2023: The following FRU and display part…

    The CARESAPE Central Station (CSCS) V2 can shut down due to a potential power supply component failure. This can lead to loss of patient monitoring at the central station.

    FDA · DeviceGE Healthcare, LLCWorldwide distribution - US Nationwide. There was also military/government…Official notice →
  9. 2022-06-15
  10. US
    Drugs & medicalClass II

    GE Healthcare, LLC recalls Centricity Radiology RA600

    Potential for Distance and Area measurements to display inaccurate measurement values when performed on magnified images, and specific to ZFP, Distance and Area measurements can display inaccurate measurement values…

    FDA · DeviceGE Healthcare, LLCWorldwide distribution. Nationwide distribution to AL, AR, AZ, CA, CO, CT, DC,…Official notice →
All 41 GE Healthcare recalls →

Other drugs & medical recalls in the United States

  1. 2022-07-13
  2. US
    Drugs & medicalClass II

    Rayner Intraocular Lenses Ltd recalls RayOne Preloaded Hydrophilic Acrylic IOL Injection Systems containing one MICS injection system with Intraocular Lens: (1) Outer…

    The outer package is mislabeled and the package contains a different IOL strength.

    FDA · DeviceRayner Intraocular Lenses LtdUS Nationwide distribution in the states of AZ, CA, FL, GA, IN, MI, MN, MO,…Official notice →
  3. US
    Drugs & medicalClass II

    Chromsystems Instruments & Chemicals GmbH recalls MassCheck Amino Acid, Acylcarnitines Dried Blood Spot Control Level I

    Too low concentration of glycine was detected in the dried blood controls of the batch 2821 of the products.

    FDA · DeviceChromsystems Instruments & Chemicals GmbHUS: NY, OH, CT.Official notice →
  4. US
    Drugs & medicalClass II

    Luminex Corporation recalls VERIGENE Enteric Pathogens Nucleic Acid Test, Part No. 30-002-23

    There is a potential for false negative results using VERIGENE CDF Stool PREP KIT and VERIGENE EP Stool PREP KIT due to hydrophobic characteristics of the swab.

    FDA · DeviceLuminex CorporationWorldwide distribution - US Nationwide and the countries of Kuwait, Austria,…Official notice →
  5. US
    Drugs & medicalClass II

    Arcoma AB recalls Omnera 400A Digital Radiographic System

    There is the potential for the X-Ray system to short circuit due to an ingress of liquid in the display handles which may result in uncontrolled movement of the Overhead Tube Crane (OTC).

    FDA · DeviceArcoma ABUS Nationwide distribution in the states of NM, DE, TX, VT, FL, MO, UT, PA,…Official notice →
  6. US
    Drugs & medicalClass II

    Ossur Americas recalls OSSUR VARIFLEX SUBA JUNIOR foot and ankle external prosthesis, sized for children and adolescents.

    Due to receiving complaints related to broken (cracked) prosthetic foot with failures occurring in situations where sudden high impact twisting or bending is applied to the foot.

    FDA · DeviceOssur AmericasWorldwide distribution including US nationwide (including Puerto Rico).Official notice →
  7. US
    Drugs & medicalClass II

    DePuy Spine, Inc. recalls EIT CIF cage, H 7mm, 8degree, S- an intervertebral body fusion devices indicated for use with autograft when used as an adjunct to fusion…

    Label on the Outer Carton does not match the labels on the Outer Blister package, Inner Blister package and Patient Labels. The only label that may not be correct is on the Outer Carton

    FDA · DeviceDePuy Spine, Inc.Distribution US NationwideOfficial notice →
See all drugs & medical recalls →