Recall Atlas
Drugs & medicalClass II

Microgenics Corporation recalls MAS Omni IMMUNE PRO Liquid Assayed Integrated Immunoassay Control, Level 1, REF: OPRO-101; MAS Omni IMMUNETM PRO Liquid Assayed…

Vial-to-vial variability in MAS¿ Omni"IMMUNE PRO Liquid Assayed Integrated Immunoassay Control, Level 1, Lot OPRO29061A results in low Parathyroid Hormone (PTH) recovery, which may lead to a delay in reporting patient…

Read the official notice →
Date
2026-10-21
Country
United States
Brand
Microgenics
Product
MAS Omni IMMUNE PRO Liquid Assayed Integrated Immunoassay Control, Level 1, REF: OPRO-101; MAS Omni IMMUNETM PRO Liquid Assayed Integrated…
Classification
Class II
Units
Worldwide - US Nationwide distribution in the states of UT, MI, OK, NY, TX,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Microgenics recalls

  1. 2025-11-05
  2. Drugs & medicalClass II

    Microgenics Corporation recalls Thermo scientific MAS Omni CARDIO, Liquid Assayed Integrated Cardiac Control, Model/Catalog Number: OCRD-L, OCRD-101, OCRD-202, OCRD-303,…

    The firm received complaints from customers reporting vial-to-vial variability when using the control with high sensitivity Troponin I assays. This defect may lead to as delay in patient results for hs Troponin I assays.

    FDA · DeviceMicrogenics CorporationWorldwide - US Nationwide distribution in the states of AL, OK, NH, NY, WI,…Official notice →
  3. 2022-07-06
  4. Drugs & medicalClass II

    Microgenics Corporation recalls QMS Tacrolimus Immunoassay REF 10015573 CALIBRATOR SET QMS TACROLIMUS CAL A 10015566-3 QMS TACROLIMUS CAL B 10015567-3 QMS TACROLIMUS CAL…

    Due to QMS Tacrolimus Calibrator recovering between 82 to 94% of their target values, resulting in over-quantitation in tacrolimus patient result values when used with the QMS Tacrolimus Immunoassay.

    FDA · DeviceMicrogenics CorporationU.S.: AL, AZ, CA, DC, FL, IA, ID, IL, KY, LA, MA, NE, NJ, NM, OH, OK, PA, SC,…Official notice →
All 3 Microgenics recalls →

Other drugs & medical recalls in the United States

  1. 2026-10-21
  2. Drugs & medicalClass II

    Cook Incorporated recalls Williams Cystoscopic Injection Needles, used for cystoscopic injection into the urethra, bladder neck, and bladder wall, Reference Part…

    The female Luer lock adapter component in the affected device lots may not have been correctly seated due to misalignment and inadequate torquing during manufacturing. Potential events that may occur as a result of…

    FDA · DeviceCook IncorporatedUS States: FL, IL, NM, TX, WA. Canada, France, Germany, Italy, Spain,…Official notice →
  3. Drugs & medicalClass III

    Magellan Diagnostics, Inc. recalls LeadCare II Blood Lead Test Kit. Model/Catalog Number: 70-6762.

    In kit sub-lot 2530M-01 the date of expiration printed on the outer labeling may indicate the wrong expiration date of July 22, 2026 while the correct expiration date is July 13, 2026. If used the system will display…

    FDA · DeviceMagellan Diagnostics, Inc.US Nationwide distribution in the states of CA, IN, PA, TX, NE, IL, MA, MO, KY.Official notice →
  4. Drugs & medicalClass II

    Beckman Coulter, Inc. recalls DxU 840m Iris, REF: C76947; DxU 850m Iris, REF: C49513

    The rinse-line check valve has a shorter service lifespan than originally expected and requires annual replacement to prevent lamina fluid leakage; however, the check valve replacement was excluded from the annual…

    FDA · DeviceBeckman Coulter, Inc.Worldwide - US Nationwide distribution including in the states and territory…Official notice →
  5. Drugs & medicalClass II

    W L Gore & Associates, Inc. recalls Generic Name: GORE VIABIL Short Wire Biliary Endoprosthesis; Model/Catalog Number: VSWVH0808 and VSWVN1006 Product Description: The…

    The potential for the devices to be mislabeled with the incorrect length.

    FDA · DeviceW L Gore & Associates, Inc.US Nationwide distribution in the states of GA, RI.Official notice →
  6. Drugs & medicalClass II

    In2bones USA, LLC recalls REF: P65 ST862 CoLag SCREW, FT, 6.7X62.5 mm

    Due to labeling error, bone screws specified as fully threaded, are actually partially threaded screws.

    FDA · DeviceIn2bones USA, LLCUS Nationwide distribution in the states of FL, GA, HI, NC, OH, PA, SD, TX, VA.Official notice →
  7. Drugs & medicalClass II

    Steris Corporation recalls RAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020115

    If the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the cap is twisted…

    FDA · DeviceSteris CorporationWorldwide - US Nationwide distribution.Official notice →
See all drugs & medical recalls →