Kico Knee Innovation Company recalls Arvis Hip & Knee 2.0 Instrument Set, REF: IN-21200 containing one of the following components: IN-10500, Impactor V-Block Assembly;…
The Impactor V-Block Assembly and Trackers A, B, C and E may exhibit signs of early wear and/or corrosion on the exposed magnet face.
Read the official notice →- Date
- 2026-06-10
- Country
- United States
- Brand
- Kico Knee Innovation
- Product
- Arvis Hip & Knee 2.0 Instrument Set, REF: IN-21200 containing one of the following components: IN-10500, Impactor V-Block Assembly;…
- Classification
- Class II
- Units
- Worldwide - US Nationwide distribution in the states of MS, CA, IA, IN, MI,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Kico Knee Innovation recalls
- 2026-08-26
US
Kico Knee Innovation Company recalls Brand Name: Kico Knee Innovation Product Name: ARVIS SURGICAL NAVIGATION SYSTEM GEN 2 Model/Catalog Number: IN-29200 Software…
Due to an internal sensor issue, image displayed with eyepiece may appear tilted relative to the eyepiece housing
FDA · DeviceKico Knee Innovation CompanyConsignee Consignee Number Region Device Quantity RFK MEDICAL CE102 RICHMOND…Official notice →- 2026-06-10
US
Kico Knee Innovation Company recalls Arvis Hip & Knee 3.0 Instrument Set, REF: IN-28000 containing one of the following components: IN-10500, Impactor V-Block Assembly;…
The V-Block Assembly and Trackers A, B, C and E may exhibit signs of early wear and/or corrosion on the exposed magnet face.
FDA · DeviceKico Knee Innovation CompanyWorldwide - US Nationwide distribution in the states of MS, CA, IA, IN, MI,…Official notice →- 2025-10-22
US
Kico Knee Innovation Company recalls Brand Name: ARVIS Product Name: ARVIS" Shoulder Model/Catalog Number: IN-27300, Arvis Surgical Planning Software Software Version:…
Complaint identified issue with AI surgical planning software that may result in implant malalignment and/or decrease range of motion.
FDA · DeviceKico Knee Innovation CompanyUS Nationwide distribution in the states of IN, MN, MO, OH, FL, IL, RI, KY,…Official notice →
Other drugs & medical recalls in the United States
- 2026-06-10
US
Bayer Medical Care, Inc. recalls Avanta Multi-Patient Administration Tube Sets (AVA 500 MPAT (Part 2 of 2)) used with the Medrad¿ Avanta Injection System; Catalog…
Supplier changes resulted in a fit issue with the Contrast Flow Control Valve (CFCV) interface on the Avanta Injector System. Multiple customer reports indicate that the stopcock cannot be reliably engaged with the…
FDA · DeviceBayer Medical Care, Inc.Worldwide - US Nationwide distribution including in the states of AL, AR, AZ,…Official notice →US
Daavlin Distributing Company recalls Phothera 200, Phototherapy Devices indicated for use to treat diagnosed skin disorders such as, but not limited to, Psoriasis,…
Some of the Calibration/Output Certificates had the outputs reversed.
US
Shimadzu Medical Systems recalls Digital Angiography System Trinias:UD150G-40 / D150GC-40; Digital Angiography System BRANSIST safire:UD150G-40 / D150GC-40; Digital…
Electromagnetic contactors within high voltage generator may weld together, resulting in current being supplied to power supply even if power is off.
FDA · DeviceShimadzu Medical SystemsWorldwide - US Nationwide distribution including in the states of AK, AL, AR,…Official notice →US
Kico Knee Innovation Company recalls Arvis Hip & Knee 3.0 Instrument Set, REF: IN-28000 containing one of the following components: IN-10500, Impactor V-Block Assembly;…
The V-Block Assembly and Trackers A, B, C and E may exhibit signs of early wear and/or corrosion on the exposed magnet face.
FDA · DeviceKico Knee Innovation CompanyWorldwide - US Nationwide distribution in the states of MS, CA, IA, IN, MI,…Official notice →US
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Azurion 3M15, Model Numbers: 722064, 722222 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with…
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without…
US
Advanced Bionics, LLC recalls Brand Name: HiResolution" Bionic Ear System Product Name: HiRes Ultra 3D CI HiFocus Mid Scala Model/Catalog Number: CI-1601-04 Software…
Due to incorrect shelf-life expiration date.
FDA · DeviceAdvanced Bionics, LLCU.S. Nationwide distribution in the states of CA, CO, GA, IL, IN, KS, MA, MD,…Official notice →