Drugs & medicalClass II

Daavlin Distributing Company recalls Phothera 200, Phototherapy Devices indicated for use to treat diagnosed skin disorders such as, but not limited to, Psoriasis,…

Some of the Calibration/Output Certificates had the outputs reversed.

Read the official notice →
Date
2026-06-10
Country
United States
Brand
Daavlin Distributing
Product
Phothera 200, Phototherapy Devices indicated for use to treat diagnosed skin disorders such as, but not limited to, Psoriasis, Vitiligo,…
Classification
Class II
Units
US Nationwide distribution.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Daavlin Distributing recalls

  1. 2026-07-22
  2. US
    Drugs & medicalClass II

    Daavlin Distributing Company recalls Photera 600 3-D, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders

    Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).

    FDA · DeviceDaavlin Distributing CompanyUS Nationwide distribution.Official notice →
  3. 2022-05-18
  4. US
    Drugs & medicalClass II

    Daavlin Distributing Company recalls 1 Series CX phototherapy units equipped with Daavlin's ClearLink Control System software versions v3.03h and 3.04v. Models: [1 Ser…

    Software issue, resuming an interrupted treatment will result in swap of utilized calibration treatment distance 0" inches to "9. Resulting in over or under dosing.

    FDA · DeviceDaavlin Distributing CompanyWorldwide distribution - US Nationwide distribution in the states of AL, AR,…Official notice →
All 3 Daavlin Distributing recalls →

Other drugs & medical recalls in the United States

  1. 2026-06-10
  2. US
    Drugs & medicalClass II

    DeRoyal Industries Inc recalls DeRoyal Industries Inc. Ref 46-406, Covaderm Plus, Vascular Access Dressing 6"x6" (15cmx15cm)

    Sterile wound dressing, lacks sterility assurance

    FDA · DeviceDeRoyal Industries IncUS Nationwide distribution in the states of FL & NC.Official notice →
  3. US
    Drugs & medicalClass II

    CareFusion 303, Inc. recalls BD Pyxis MedBank Tower REF: 1145-00 Medication cabinet

    Fingerprint scanner may overheat to a temperature to cause 1st degree burn.

    FDA · DeviceCareFusion 303, Inc.US: Nationwide OUS: AE, AU, BE, BH, BR, CA, CL, ES, GB, IT, KW, MA, MX, NZ,…Official notice →
  4. US
    Drugs & medicalClass II

    Bayer Medical Care, Inc. recalls Avanta Multi-Patient Administration Tube Sets (AVA 500 MPAT (Part 2 of 2)) used with the Medrad¿ Avanta Injection System; Catalog…

    Supplier changes resulted in a fit issue with the Contrast Flow Control Valve (CFCV) interface on the Avanta Injector System. Multiple customer reports indicate that the stopcock cannot be reliably engaged with the…

    FDA · DeviceBayer Medical Care, Inc.Worldwide - US Nationwide distribution including in the states of AL, AR, AZ,…Official notice →
  5. US
    Drugs & medicalClass II

    Shimadzu Medical Systems recalls Digital Angiography System Trinias:UD150G-40 / D150GC-40; Digital Angiography System BRANSIST safire:UD150G-40 / D150GC-40; Digital…

    Electromagnetic contactors within high voltage generator may weld together, resulting in current being supplied to power supply even if power is off.

    FDA · DeviceShimadzu Medical SystemsWorldwide - US Nationwide distribution including in the states of AK, AL, AR,…Official notice →
  6. US
    Drugs & medicalClass II

    Kico Knee Innovation Company recalls Arvis Hip & Knee 2.0 Instrument Set, REF: IN-21200 containing one of the following components: IN-10500, Impactor V-Block Assembly;…

    The Impactor V-Block Assembly and Trackers A, B, C and E may exhibit signs of early wear and/or corrosion on the exposed magnet face.

    FDA · DeviceKico Knee Innovation CompanyWorldwide - US Nationwide distribution in the states of MS, CA, IA, IN, MI,…Official notice →
  7. US
    Drugs & medicalClass II

    Kico Knee Innovation Company recalls Arvis Hip & Knee 3.0 Instrument Set, REF: IN-28000 containing one of the following components: IN-10500, Impactor V-Block Assembly;…

    The V-Block Assembly and Trackers A, B, C and E may exhibit signs of early wear and/or corrosion on the exposed magnet face.

    FDA · DeviceKico Knee Innovation CompanyWorldwide - US Nationwide distribution in the states of MS, CA, IA, IN, MI,…Official notice →
See all drugs & medical recalls →