Daavlin Distributing recalls

3 official recalls and safety alerts involving Daavlin Distributing since 2022-05, from FDA · Device. Newest first. Check the official notice for exact models and batch codes.

This list only includes recalls Recall Atlas has collected since 2021-10. A product not listed here may still be affected; check with Daavlin Distributing or the relevant agency.

  1. 2026-07-22
  2. US
    Drugs & medicalClass II

    Daavlin Distributing Company recalls Photera 600 3-D, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders

    Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).

    FDA · DeviceDaavlin Distributing CompanyUS Nationwide distribution.Official notice →
  3. 2026-06-10
  4. US
    Drugs & medicalClass II

    Daavlin Distributing Company recalls Phothera 200, Phototherapy Devices indicated for use to treat diagnosed skin disorders such as, but not limited to, Psoriasis,…

    Some of the Calibration/Output Certificates had the outputs reversed.

    FDA · DeviceDaavlin Distributing CompanyUS Nationwide distribution.Official notice →
  5. 2022-05-18
  6. US
    Drugs & medicalClass II

    Daavlin Distributing Company recalls 1 Series CX phototherapy units equipped with Daavlin's ClearLink Control System software versions v3.03h and 3.04v. Models: [1 Ser…

    Software issue, resuming an interrupted treatment will result in swap of utilized calibration treatment distance 0" inches to "9. Resulting in over or under dosing.

    FDA · DeviceDaavlin Distributing CompanyWorldwide distribution - US Nationwide distribution in the states of AL, AR,…Official notice →