Drugs & medicalClass II

Kico Knee Innovation Company recalls Brand Name: Kico Knee Innovation Product Name: ARVIS SURGICAL NAVIGATION SYSTEM GEN 2 Model/Catalog Number: IN-29200 Software…

Due to an internal sensor issue, image displayed with eyepiece may appear tilted relative to the eyepiece housing

Read the official notice →
Date
2026-08-26
Country
United States
Brand
Kico Knee Innovation
Product
Brand Name: Kico Knee Innovation Product Name: ARVIS SURGICAL NAVIGATION SYSTEM GEN 2 Model/Catalog Number: IN-29200 Software Version:…
Classification
Class II
Units
Consignee Consignee Number Region Device Quantity RFK MEDICAL CE102 RICHMOND…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Kico Knee Innovation recalls

  1. 2026-06-10
  2. US
    Drugs & medicalClass II

    Kico Knee Innovation Company recalls Arvis Hip & Knee 2.0 Instrument Set, REF: IN-21200 containing one of the following components: IN-10500, Impactor V-Block Assembly;…

    The Impactor V-Block Assembly and Trackers A, B, C and E may exhibit signs of early wear and/or corrosion on the exposed magnet face.

    FDA · DeviceKico Knee Innovation CompanyWorldwide - US Nationwide distribution in the states of MS, CA, IA, IN, MI,…Official notice →
  3. US
    Drugs & medicalClass II

    Kico Knee Innovation Company recalls Arvis Hip & Knee 3.0 Instrument Set, REF: IN-28000 containing one of the following components: IN-10500, Impactor V-Block Assembly;…

    The V-Block Assembly and Trackers A, B, C and E may exhibit signs of early wear and/or corrosion on the exposed magnet face.

    FDA · DeviceKico Knee Innovation CompanyWorldwide - US Nationwide distribution in the states of MS, CA, IA, IN, MI,…Official notice →
  4. 2025-10-22
  5. US
    Drugs & medicalClass II

    Kico Knee Innovation Company recalls Brand Name: ARVIS Product Name: ARVIS" Shoulder Model/Catalog Number: IN-27300, Arvis Surgical Planning Software Software Version:…

    Complaint identified issue with AI surgical planning software that may result in implant malalignment and/or decrease range of motion.

    FDA · DeviceKico Knee Innovation CompanyUS Nationwide distribution in the states of IN, MN, MO, OH, FL, IL, RI, KY,…Official notice →
All 4 Kico Knee Innovation recalls →

Other drugs & medical recalls in the United States

  1. 2026-08-26
  2. US
    Drugs & medicalClass II

    Boston Scientific Corporation recalls OverStitch Suture Cinch, 1pk. Endoscopic tissue approximation device.

    Device may fail to deploy the cinch implant and/or fail to cut the suture as intended.

    FDA · DeviceBoston Scientific CorporationWorldwide distribution - US Nationwide and the countries of Austria,…Official notice →
  3. US
    Drugs & medicalClass II

    Tobii Dynavox Llc recalls TD Navio Midi Speech Generating Device Model/Catalog Number: 1000798

    Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and…

    FDA · DeviceTobii Dynavox LlcUS Nationwide distribution.Official notice →
  4. US
    Drugs & medicalClass II

    Arrowhead DE LLC recalls Arrow-LOK Implant, REF: 21-253519, 21-253522, 21-253525, 21-253528, 21-253530, 21-253533, 21-253535, 21-253540, 21-253545, 21-253550,…

    Implant did not undergo required gamma radiation dose audit. A potential reduction in sterility assurance could lead to patient infection.

    FDA · DeviceArrowhead DE LLCUS Nationwide distribution in the states of VT, FL, TN, NY, GA, TX, HI.Official notice →
  5. US
    Drugs & medicalClass II

    Philips North America Llc recalls Cardiac Workstation 5000. Product Number: 860439.

    Device will intermittently provide inaccurate data related to native heart rate.

    FDA · DevicePhilips North America LlcWorldwide distribution - US Nationwide and the countries of Australia,…Official notice →
  6. US
    Drugs & medicalClass II

    SDI Limited recalls Dentsply Sirona Capsule Mixer Model No. 5546068

    Potential for electrical discharge from unit when unplugged.

    FDA · DeviceSDI LimitedUS: AK AL AR AZ CA CO CT DE FL GA HI IA ID IL IN KS KY LA MA MD ME MI MN MO MS…Official notice →
  7. US
    Drugs & medicalClass II

    Intersurgical Inc recalls One-Piece Guedel Airway, size 000, ISO 3.5, Light Green. Oropharyngeal Airway Device.

    Devices could potentially be occluded or partially occluded.

    FDA · DeviceIntersurgical IncUS Nationwide distribution in the states of CT, FL, IN, MI, OH & VA.Official notice →
See all drugs & medical recalls →