Ion Beam Applications S.A. recalls IBA Proton Therapy System - PROTEUS 235
It was identified that eh Universal Beam Triggering Interface (UBTI) is disabled in the Therapy Safety System (TSS) configuration. In this condition, the TSS does not interrupt the beam delivery if the Healthy Signal…
Read the official notice →- Date
- 2026-03-18
- Country
- United States
- Brand
- Ion Beam Applications
- Product
- IBA Proton Therapy System - PROTEUS 235
- Classification
- Class II
- Units
- Worldwide - US Nationwide distribution in the state of FL and the countries of…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Ion Beam Applications recalls
- 2024-09-18
US
Ion Beam Applications S.A. recalls IBA Proton Therapy System - PROTEUS 235 ProteusPLUS and ProteusONE Software versions: PTS-8 versions, PTS-10 versions, PTS-11…
Safety Parameters Verification Mechanisms can be deactivated in clinical runtime
FDA · DeviceIon Beam Applications S.A.Worldwide - US Nationwide distribution in the states of AK, NJ, PA, VA,OK,FL,…Official notice →- 2024-08-28
US
Ion Beam Applications S.A. recalls IBA Proton Therapy System - PROTEUS 235- Designed to produce and deliver a proton beam for the treatment of patients with localized…
Temporary deactivation of collision detection for Forte Robotic Patient Positioning System occasionally experiences issues
FDA · DeviceIon Beam Applications S.A.IL,LA, NJ, MI,OK, TN, TX, VA, WA Foreign: Czech Republic France Germany Italy…Official notice →- 2024-06-12
US
Ion Beam Applications S.A. recalls IBA Proton Therapy System - PROTEUS 235 - Designed to produce and deliver a proton beam for the treatment of patients with localized…
Irradiation is not interrupted when some C230 Accelerator Control Unit critical parameters are outside of the defined tolerances
FDA · DeviceIon Beam Applications S.A.Worldwide - US Nationwide distribution in the state of Virginia and the…Official notice →- 2024-05-01
US
Ion Beam Applications S.A. recalls IBA Proton Therapy System - PROTEUS 235, PTS-10 versions, PTS-11 versions, and PTS-12 versions.
There is a risk of mistreatment as irradiation is not prevented when some safety parameters are out of tolerance.
FDA · DeviceIon Beam Applications S.A.US Nationwide distribution in the states of AR, FL, LA, KS, MI, TX, VA.Official notice →- 2023-11-22
US
Ion Beam Applications S.A. recalls IBA Proton Therapy System - PROTEUS 235. PTS-11 versions, PTS-12 versions with Orion PPS version OSS6.0. Configured with Orion Patient…
The patient positioning system can perform motions in unexpected directions, creating the potential for crushing of the patient between the patient positioning devices.
FDA · DeviceIon Beam Applications S.A.Worldwide - US Nationwide distribution in the states of Arkansas, Florida,…Official notice →- 2022-06-08
US
Ion Beam Applications S.A. recalls Proteus235- Proton Therapy System: to produce and deliver a proton beam for the treatment of patients with localized tumors and other…
First layer of Uniform Scanning treatment fields is sometimes irradiated with incorrect Scanning Magnet setpoints resulting in the field size may be smaller than expected and the irradiated area may receive more dose…
FDA · DeviceIon Beam Applications S.A.Worldwide distribution - US Nationwide distribution in the states of IL, VA,…Official notice →
Other drugs & medical recalls in the United States
- 2026-03-18
US
Exactech, Inc. recalls Brand Name: Equinoxe Ergo Product Name: Ergo Modular Impactor Handle Model/Catalog Number: 321-09-05
Impactor handle may be missing cross-pin
FDA · DeviceExactech, Inc.US: CA CO FL GA HI ID IL IN KS KY LA MA MD ME MN NV NY OH OK PR SC TN TX VA WI…Official notice →US
Penner Patient Care, Inc. recalls Penner Pacific Bathing Spa, Model Numbers 360020-1P
The device does not bear a unique device identifier.
US
Raz Design Inc recalls Raz-AT (Attendant Tilt) Mobile Shower Commode Chair, Catalog Number Z300 UDI-DI code: (01)00628758003007 Raz-ART (Attendant Rotational…
Due to a modification by a vendor and may be reinstalled without properly engaging the outside rear seat brackets that are used to keep the seat in place.
FDA · DeviceRaz Design IncU.S. Nationwide distribution in the states of CO, FL, IL, MA, MI, MN, MO, NC,…Official notice →US
Becton Dickinson & Company recalls 10mL Non-Sterile BD Luer-Lok Tip Syringe Only. Catalog Number: 301029. 850 bulk non-sterile units per case.
During the bulk packaging process, some Luer Slip syringes were incorrectly included in a batch designated for Luer Lok syringes.
FDA · DeviceBecton Dickinson & CompanyUS Nationwide distribution in the states of California and Texas.Official notice →US
KOB GmbH recalls MEDLINE UNNA-Z STRETCH ZINC PASTE BANDAGE, REF NONUNNAS40 KOB article #40307010
Due to labeling error. Incomplete ingredients listed. Labeling did not include castor oil or calamine.
FDA · DeviceKOB GmbHU.S.: IL O.U.S.: Not provided as product is manufactured O.U.S.Official notice →US
Zydus Pharmaceuticals (USA) Inc recalls Icosapent Ethyl Capsules, 1 gram, 120 Capsules per bottle, Rx only, Manufactured by: Softgel Healthcare Pvt. Ltd., India;…
Failed Tablet/Capsule specifications: Red dots inside capsule and melted capsule caused by oxidized Icosapent ethyl, the active ingredient.