Ion Beam Applications S.A. recalls IBA Proton Therapy System - PROTEUS 235 ProteusPLUS and ProteusONE Software versions: PTS-8 versions, PTS-10 versions, PTS-11…
Safety Parameters Verification Mechanisms can be deactivated in clinical runtime
Read the official notice →- Date
- 2024-09-18
- Country
- United States
- Brand
- Ion Beam Applications
- Product
- IBA Proton Therapy System - PROTEUS 235 ProteusPLUS and ProteusONE Software versions: PTS-8 versions, PTS-10 versions, PTS-11 versions,…
- Classification
- Class II
- Units
- Worldwide - US Nationwide distribution in the states of AK, NJ, PA, VA,OK,FL,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Ion Beam Applications recalls
- 2026-03-18
US
Ion Beam Applications S.A. recalls IBA Proton Therapy System - PROTEUS 235
It was identified that eh Universal Beam Triggering Interface (UBTI) is disabled in the Therapy Safety System (TSS) configuration. In this condition, the TSS does not interrupt the beam delivery if the Healthy Signal…
FDA · DeviceIon Beam Applications S.A.Worldwide - US Nationwide distribution in the state of FL and the countries of…Official notice →- 2024-08-28
US
Ion Beam Applications S.A. recalls IBA Proton Therapy System - PROTEUS 235- Designed to produce and deliver a proton beam for the treatment of patients with localized…
Temporary deactivation of collision detection for Forte Robotic Patient Positioning System occasionally experiences issues
FDA · DeviceIon Beam Applications S.A.IL,LA, NJ, MI,OK, TN, TX, VA, WA Foreign: Czech Republic France Germany Italy…Official notice →- 2024-06-12
US
Ion Beam Applications S.A. recalls IBA Proton Therapy System - PROTEUS 235 - Designed to produce and deliver a proton beam for the treatment of patients with localized…
Irradiation is not interrupted when some C230 Accelerator Control Unit critical parameters are outside of the defined tolerances
FDA · DeviceIon Beam Applications S.A.Worldwide - US Nationwide distribution in the state of Virginia and the…Official notice →- 2024-05-01
US
Ion Beam Applications S.A. recalls IBA Proton Therapy System - PROTEUS 235, PTS-10 versions, PTS-11 versions, and PTS-12 versions.
There is a risk of mistreatment as irradiation is not prevented when some safety parameters are out of tolerance.
FDA · DeviceIon Beam Applications S.A.US Nationwide distribution in the states of AR, FL, LA, KS, MI, TX, VA.Official notice →- 2023-11-22
US
Ion Beam Applications S.A. recalls IBA Proton Therapy System - PROTEUS 235. PTS-11 versions, PTS-12 versions with Orion PPS version OSS6.0. Configured with Orion Patient…
The patient positioning system can perform motions in unexpected directions, creating the potential for crushing of the patient between the patient positioning devices.
FDA · DeviceIon Beam Applications S.A.Worldwide - US Nationwide distribution in the states of Arkansas, Florida,…Official notice →- 2022-06-08
US
Ion Beam Applications S.A. recalls Proteus235- Proton Therapy System: to produce and deliver a proton beam for the treatment of patients with localized tumors and other…
First layer of Uniform Scanning treatment fields is sometimes irradiated with incorrect Scanning Magnet setpoints resulting in the field size may be smaller than expected and the irradiated area may receive more dose…
FDA · DeviceIon Beam Applications S.A.Worldwide distribution - US Nationwide distribution in the states of IL, VA,…Official notice →
Other drugs & medical recalls in the United States
- 2024-09-18
US
Zoe Medical Incorporated recalls Spacelabs Ultraview DM3 Vital Signs Monitor (all variations -M, -MT, -N, -NT), Model Number 91330
This product labeling does not adequately provide warning about the risks associated with the use of damaged batteries or batteries past the intended use life.
US
Smith & Nephew Inc. recalls ROTATION MEDICAL TENDON STAPLES (8) (narrow indication)-Intended for fixation of prosthetic material to soft tissues. Product Number:…
Product packaging process may result in an improper or incomplete seal of the outer foil pouch surrounding the inner Tyvek pouch, compromising product sterility
FDA · DeviceSmith & Nephew Inc.Worldwide distribution - US Nationwide and the countries of Australia,…Official notice →US
MEDLINE INDUSTRIES, LP - Northfield recalls Medline ReNewal Reprocessed Covidien LigaSure Maryland Jaw Sealer/Divider, Nano-Coated, Compatible w/ FT10 37 cm FT10 SW…
Potential for an incomplete seal on the packaging tray.
US
Smiths Medical ASD Inc. recalls smiths medical portex BLUselect, tracheostomy tube, labeled as follows: 1) BLUSELECT 6.0, CUFFED, NON-FEN, 10/BX, Item Number…
There is the potential for a disconnection of the pilot balloon from the tracheostomy inflation line within specific lots of the BLUSelect, BLUgriggs and BLUperc products due to a manufacturing defect.
US
Sol-Millennium Medical Inc. recalls Syringes with Eccentric Tips labeled as: 1) SOL-M 10ml Eccentric Tip Syringe w/o Needle, Model No. P180010ET, UPC 81001809551; 2)…
Products in distribution were found by FDA to be to be substantially different from the current cleared 510(k)s with a major change or modification to the device s intended use.
FDA · DeviceSol-Millennium Medical Inc.US Nationwide, Czech Republic, Belgium, Portugal, Canada, Saudi Arabia,…Official notice →US
Medtronic Neuromodulation recalls Medtronic Handset with Communicator, Model TH90Q01, InterStim Smart Programmer and Communicator Kits, containing the Model HH90 Handset…
A subset of the handsets within the kit may not be able to complete the pairing process with the communicator upon initial setup.