Hamilton Medical AG recalls Hamilton C6 REF 160021 (ESM Board PN 10144626 UDI-DI code: 76300028PN160021ZW Hamilton C6 is a ventilator intended to provide positive…
A mechanical defect on a printed circuit board (PCB) as a result of PCB separating process. This defect might cause the ventilator to either fail at startup (no patient involvement) or during the use of the device…
Read the official notice →- Date
- 2025-08-27
- Country
- United States
- Brand
- Hamilton Medical
- Product
- Hamilton C6 REF 160021 (ESM Board PN 10144626 UDI-DI code: 76300028PN160021ZW Hamilton C6 is a ventilator intended to provide positive…
- Classification
- Class I
- Units
- US Nationwide distribution.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Hamilton Medical recalls
- 2026-07-08
US
Hamilton Medical AG recalls IntelliCuff, Portable automatic cuff pressure controller REF: 951001
Due to reported complaints, the cuff device may alarm with a "Cuff System Leakage" error. As a result, the motor may pump continuously and fail to maintain the cuff pressure set by the user.
FDA · DeviceHamilton Medical AGWorldwide - US Nationwide distribution in the states of AZ, CA, CO, FL, GA,…Official notice →US
Hamilton Medical AG recalls Breathing circuit set, coaxial: REF: 260127, 260128, 260167, 260168 used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.
Ventilator coaxial breathing circuit may have, expiratory valve sets with membranes sticking to valve body, not detectable during pre-operative test, which will result in "exhalation obstructed " alarm, which may…
FDA · DeviceHamilton Medical AGUS Nationwide distribution including in the states of KS, WA, VT, TX, WI, AR,…Official notice →- 2025-08-06
US
Hamilton Medical AG recalls Breathing circuit set, coaxial, 1.80 m, with flow sensor and expiratory valve set, REF: 260128, used with HAMILTON-C1, HAMILTON-T1 and…
Ventilator coaxial breathing circuit may have a crack in blue tubing, which is used to deliver fresh gas to the patient. The crack will compromise inspiratory and expiratory flows, leading to partial or complete…
FDA · DeviceHamilton Medical AGUS Nationwide distribution in the states of NH, NY, CT, PA, IL, ME, CO, NC,…Official notice →- 2024-06-26
US
Hamilton Medical AG recalls Ventilator HAMILTON-C6, PN: 160021
Ventilator may enter sensor fail mode, ventilation may not re-initiate, after patient is reconnected, which may cause hypoxia, if this sequence occurs: 1)User presses O2 enrichment, 2)User disconnects endotracheal tube…
FDA · DeviceHamilton Medical AGUS Nationwide distribution in the states of GA, KY, FL, IN, NV, MI, VA, OR,…Official notice →- 2023-09-27
US
Hamilton Medical, Inc. recalls Hamilton C1/T1/MR1, REF: MSP161502/12, Control Board Spare part for ventilator
Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption…
FDA · DeviceHamilton Medical, Inc.Distribution US nationwide, including Puerto Rico. OUS: Not providedOfficial notice →- 2023-08-02
US
Hamilton Medical, Inc. recalls HAMILTON C3 Ventilator REF 160005 Is intended to provide positive pressure ventilatory support to adults and pediatrics, and optionally…
Due to a software anomaly after approximately 91 days of cumulative use, the ventilator may stop and enter the ambient mode without prior notice.
FDA · DeviceHamilton Medical, Inc.U.S.: Alabama Alaska Arizona Arkansas California Colorado Connecticut Delaware…Official notice →
Other drugs & medical recalls in the United States
- 2025-08-27
US
Spiggle & Theis Mt Gmbh recalls Injection needle, bayonet, Luer-Lock, 1. Model no. 50-353-23, working length 25 cm, total length 31cm, nonsterile; 2. Model no.…
Affected product show elevated cytotoxicity values. Use of affected injection needles may present a risk of intolerance reaction or allergic reaction.
FDA · DeviceSpiggle & Theis Mt GmbhUS Nationwide distribution in the states of California, Connecticut, Missouri,…Official notice →US
Drs Vascular, Inc recalls Brand Name: Q-Stop Product Name: Suture Retention Device Model/Catalog Number: QS Software Version: N/A Product Description: The DRS…
XXX
FDA · DeviceDrs Vascular, IncU.S. Nationwide distribution in the states of AR, CA, and NY.Official notice →US
Pentax of America Inc recalls Pentax Medical Video Processor; Model Number: EPK-i8020c;
During endoscopic procedures using a combination of the video processor EPK-i8020c and i20c series video endoscope, the observed image can become reddish or dark. Users have observed smoke like steam and noted that the…
FDA · DevicePentax of America IncUS Distribution to states of: AZ, CA, DC, FL IA, ID, IL, KS, MA, MD, MI, MO,…Official notice →US
Boston Scientific Corporation recalls ENDOTAK RELIANCE G/SG with 4-SITE Connector, transvenous defibrillation lead, Permanent Defibrillator Electrodes; Model Nos. 0282,…
Affected lead models may experience gradually rising LVSI measurements in association with calcification of the defibrillation lead coils. The issue may lead to early lead replacement. Most serious harm is death or…
FDA · DeviceBoston Scientific CorporationWorldwide Distribution. US nationwide. International distribution worldwide.Official notice →US
US
The Harvard Drug Group LLC dba Major Pharmaceuticals and… recalls Carvedilol Tablets USP, 12.5 mg, 100- Tablets, (10x10) cartons, Rx Only, Packaged and distributed by…
CGMP Deviations: Results for N-Nitroso Carvedilol Impurity-1 (NNCI) impurity observed to be above the FDA-recommended limit of NMT 4.0 ppm
FDA · DrugThe Harvard Drug Group LLC dba Major Pharmaceuticals and…Product was distributed nationwide within the United States.Official notice →