Hamilton Medical AG recalls Breathing circuit set, coaxial: REF: 260127, 260128, 260167, 260168 used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.
Ventilator coaxial breathing circuit may have, expiratory valve sets with membranes sticking to valve body, not detectable during pre-operative test, which will result in "exhalation obstructed " alarm, which may…
Read the official notice →- Date
- 2026-07-08
- Country
- United States
- Brand
- Hamilton Medical
- Product
- Breathing circuit set, coaxial: REF: 260127, 260128, 260167, 260168 used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.
- Classification
- Class I
- Units
- US Nationwide distribution including in the states of KS, WA, VT, TX, WI, AR,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Hamilton Medical recalls
- 2026-07-08
US
Hamilton Medical AG recalls IntelliCuff, Portable automatic cuff pressure controller REF: 951001
Due to reported complaints, the cuff device may alarm with a "Cuff System Leakage" error. As a result, the motor may pump continuously and fail to maintain the cuff pressure set by the user.
FDA · DeviceHamilton Medical AGWorldwide - US Nationwide distribution in the states of AZ, CA, CO, FL, GA,…Official notice →- 2025-08-27
US
Hamilton Medical AG recalls Hamilton C6 REF 160021 (ESM Board PN 10144626 UDI-DI code: 76300028PN160021ZW Hamilton C6 is a ventilator intended to provide positive…
A mechanical defect on a printed circuit board (PCB) as a result of PCB separating process. This defect might cause the ventilator to either fail at startup (no patient involvement) or during the use of the device…
- 2025-08-06
US
Hamilton Medical AG recalls Breathing circuit set, coaxial, 1.80 m, with flow sensor and expiratory valve set, REF: 260128, used with HAMILTON-C1, HAMILTON-T1 and…
Ventilator coaxial breathing circuit may have a crack in blue tubing, which is used to deliver fresh gas to the patient. The crack will compromise inspiratory and expiratory flows, leading to partial or complete…
FDA · DeviceHamilton Medical AGUS Nationwide distribution in the states of NH, NY, CT, PA, IL, ME, CO, NC,…Official notice →- 2024-06-26
US
Hamilton Medical AG recalls Ventilator HAMILTON-C6, PN: 160021
Ventilator may enter sensor fail mode, ventilation may not re-initiate, after patient is reconnected, which may cause hypoxia, if this sequence occurs: 1)User presses O2 enrichment, 2)User disconnects endotracheal tube…
FDA · DeviceHamilton Medical AGUS Nationwide distribution in the states of GA, KY, FL, IN, NV, MI, VA, OR,…Official notice →- 2023-09-27
US
Hamilton Medical, Inc. recalls Hamilton C1/T1/MR1, REF: MSP161502/12, Control Board Spare part for ventilator
Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption…
FDA · DeviceHamilton Medical, Inc.Distribution US nationwide, including Puerto Rico. OUS: Not providedOfficial notice →- 2023-08-02
US
Hamilton Medical, Inc. recalls HAMILTON C3 Ventilator REF 160005 Is intended to provide positive pressure ventilatory support to adults and pediatrics, and optionally…
Due to a software anomaly after approximately 91 days of cumulative use, the ventilator may stop and enter the ambient mode without prior notice.
FDA · DeviceHamilton Medical, Inc.U.S.: Alabama Alaska Arizona Arkansas California Colorado Connecticut Delaware…Official notice →
Other drugs & medical recalls in the United States
- 2026-07-08
US
Insulet Corporation recalls Omnipod 5 Pods. Includes the below Model/REF Numbers: 1. POD-OMNI-I1-6720. 2. POD-OMNI-I1-6220. 3. OMNI-I1-6729 (10-Pack), OMNI-I1-6720…
External soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The primary failure mode is pump…
FDA · DeviceInsulet CorporationWorldwide distribution - US Nationwide and the countries of Austria,…Official notice →US
Baxter Healthcare Corporation recalls Hillrom VOLARA System P.CIRCUIT KIT AC (Single Patient Use Circuit), Model/Catalog Number REF M08473; Oscillation Lung Expansion…
There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization,…
FDA · DeviceBaxter Healthcare CorporationWorldwide distribution - US Nationwide and the country of Canada.Official notice →US
Medline Industries, LP recalls Medline Convenience Kits: 1) SUTURE TRAY, Model Number: SUT12260; 2) CHEST TUBE INSERTION TRAY, Model Number: MNS3390; 3) CONTRACEPTIVE…
The kits containing Huons Lidocaine HCL Injection USP 1% 10mg/ml 5mL Ampules (Lidocaine Ampules)(OEM). A recall notice was issued by Huons Co., Ltd. for Lidocaine Ampules due to quality issues identified during a…
US
Lupin Pharmaceuticals Inc. recalls Glucagon Emergency Kit for Low Blood Sugar, Glucagon for Injection USP, 1mg per vial, Diluent for Glucagon, 1 mL syringe, Rx only,…
CGMP deviation: OOS result observed for the Gliding Force functionality test during 12-month long term stability testing.
US
McKesson recalls Gemcitabine Injection, 1 g per 26.3 mL (38 mg/mL), 26.3 mL Single-Dose Vial, For Intravenous Infusion ONLY, Rx only, Manufactured by:…
Temperature Abuse: Product was stored incorrectly in a controlled room temperature environment instead of a refrigerated environment.
US
Blaine Labs Inc recalls Revitaderm Wound Care, First Aid Antiseptic Gel, Benzalkonium Chloride 0.1%) packaged as (a) 1.0 Fl OZ (16 mL) bottle, long pointy cap,…
CGMP Deviations; product manufactured in the same facility under the same conditions as the product found to be contaminated