Fresenius Medical Care Holdings, Inc. recalls RTL190395: 2008K@HOME MACHINE, SHORT CAB,OLC/DP,HP - Blood Pump Rotor spare part, Model Number: F40015481 Rev A
An increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
Read the official notice →- Date
- 2024-07-17
- Country
- United States
- Brand
- Fresenius Medical Care Holdings
- Product
- RTL190395: 2008K@HOME MACHINE, SHORT CAB,OLC/DP,HP - Blood Pump Rotor spare part, Model Number: F40015481 Rev A
- Classification
- Class II
- Units
- Domestic: Nationwide Distribution. International: Canada.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Fresenius Medical Care Holdings recalls
- 2025-01-15
US
Fresenius Medical Care Holdings, Inc. recalls Dialyzer Optiflux 160NRe
Potential for internal blood leaks due to cracked polyurethane
- 2024-09-18
US
Fresenius Medical Care Holdings, Inc. recalls Blood Pressure Module BPM3 TM-2917B (Spare Part Numbers 191378 or 191389) with firmware version S005 installed in 2008…
Machines equipped with BPM3 Blood Pressure Modules (Blood pressure monitor TM-2917B with firmware version S005) will not record Blood Pressure measurements and does not alarm if the cuff tubing is kinked close to the…
- 2024-07-24
US
Fresenius Medical Care Holdings, Inc. recalls DELFLEX PERITONEAL DIALYSIS SOLUTION With LOW MAGNESIUM / LOW CALCIUM 4.25% DEXTROSE and attached stay "safe Exchange Set,…
This product is being recalled due to the tube weld failure presents itself as a slow leak and can be difficult to detect.
- 2024-06-19
US
Fresenius Medical Care Holdings, Inc. recalls US AquaA 900H-Typ USA-3 x 208v-Intended for the purification of water to be used for hemodialysis Vivonic Part Numbers:…
Concentrate Flow was exceeding the maximum 19.9 bar set point. The standard default set point range is between 4 and 18.Error message (F020104, "concentrate pressure alarm limit exceeded") concentrate pressure is…
FDA · DeviceFresenius Medical Care Holdings, Inc.US Nationwide distribution in the states of AL, IN, AK, WI.Official notice →- 2024-05-29
US
Fresenius Medical Care Holdings, Inc. recalls Blood Pump Rotor, Model Number: F40015481 Rev A
An increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
- 2024-05-22
US
Fresenius Medical Care Holdings, Inc. recalls Novalung sensor box is a component of the Novalung Console (F30000162)-Indicated for long-term (> 6 hours) respiratory/…
Issue related to Novalung sensor box, which is a component of the Novalung Console (Part Number F30000163). In certain instances, error messages #206 (yellow) and #208 (red) technical failure, flow measurement during…
Other drugs & medical recalls in the United States
- 2024-07-24
US
US
Brands International Corporation recalls Oatmeal Daily Moisturizing Body Lotion (1.3% Dimethicone), 8 fl oz (236mL), packaged in an HDPE bottle 12 bottles per case,…
Microbial Contamination of Non-Sterile Products: confirmed presence of mold contamination
US
- 2024-07-17
US
Baxter Healthcare Corporation recalls Braun Thermoscan¿ PRO 6000 Ear Thermometer
The Braun Thermoscan PRO 6000 ear thermometers may have been shipped with a compact disc (CD) containing an outdated version of the Instructions for Use (IFU).
FDA · DeviceBaxter Healthcare CorporationDistributed to United States, Canada, Australia, Malaysia, New Zealand,…Official notice →US
Maquet Cardiovascular, LLC recalls The VASOVIEW HEMOPRO Endoscopic Vessel Harvesting Systems is designed for use in conjunction with the 7 mm Extended Length Endoscope.…
There were 27 complaints between March 22, 2024, and April 30, 2024, related to VH-3500 and VH-4000, reporting that the C-Ring wire was straightened (due to insufficient bend radius) and that the distance between…
FDA · DeviceMaquet Cardiovascular, LLCNationwideForeign: AT DE DK ES GB HK IT NL ZAOfficial notice →US
MICROspecialties, Inc. recalls Synergetics I Pack Injection Kit Custom-Tray for use with Ophthalmic injections Model/Catalog Number: 18069 Product Description:
Could not confirm products were exposed to the proper sterilization cycle can potentially cause post operative wound infection or injection site infection.