Fresenius Medical Care Holdings, Inc. recalls DELFLEX PERITONEAL DIALYSIS SOLUTION With LOW MAGNESIUM / LOW CALCIUM 4.25% DEXTROSE and attached stay "safe Exchange Set,…
This product is being recalled due to the tube weld failure presents itself as a slow leak and can be difficult to detect.
Read the official notice →- Date
- 2024-07-24
- Country
- United States
- Brand
- Fresenius Medical Care Holdings
- Product
- DELFLEX PERITONEAL DIALYSIS SOLUTION With LOW MAGNESIUM / LOW CALCIUM 4.25% DEXTROSE and attached stay "safe Exchange Set, 2500mL (Approx.…
- Classification
- Class II
- Units
- Nationwide
- Category
- Drugs & medical
- Source
- FDA · Drug
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Fresenius Medical Care Holdings recalls
- 2025-06-25
US
Fresenius Medical Care Holdings, Inc. recalls 5008X HDF Hemodialysis Blood Tubing Set, 24 units per case, with the following models: a. 5008X Standard HDF Bloodline,…
Potential of blood leak occurring between the lower port of the venous chamber and the tubing include: hemodiafiltration (HDF), hemodialysis (HD), hemofiltration (HF), and isolated ultrafiltration.
FDA · DeviceFresenius Medical Care Holdings, Inc.US Nationwide distribution in the state of Massachusetts.Official notice →- 2025-01-15
US
Fresenius Medical Care Holdings, Inc. recalls Dialyzer Optiflux 160NRe
Potential for internal blood leaks due to cracked polyurethane
- 2024-09-18
US
Fresenius Medical Care Holdings, Inc. recalls Blood Pressure Module BPM3 TM-2917B (Spare Part Numbers 191378 or 191389) with firmware version S005 installed in 2008…
Machines equipped with BPM3 Blood Pressure Modules (Blood pressure monitor TM-2917B with firmware version S005) will not record Blood Pressure measurements and does not alarm if the cuff tubing is kinked close to the…
- 2024-07-17
US
Fresenius Medical Care Holdings, Inc. recalls RTL190395: 2008K@HOME MACHINE, SHORT CAB,OLC/DP,HP - Blood Pump Rotor spare part, Model Number: F40015481 Rev A
An increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
FDA · DeviceFresenius Medical Care Holdings, Inc.Domestic: Nationwide Distribution. International: Canada.Official notice →- 2024-06-19
US
Fresenius Medical Care Holdings, Inc. recalls US AquaA 900H-Typ USA-3 x 208v-Intended for the purification of water to be used for hemodialysis Vivonic Part Numbers:…
Concentrate Flow was exceeding the maximum 19.9 bar set point. The standard default set point range is between 4 and 18.Error message (F020104, "concentrate pressure alarm limit exceeded") concentrate pressure is…
FDA · DeviceFresenius Medical Care Holdings, Inc.US Nationwide distribution in the states of AL, IN, AK, WI.Official notice →- 2024-05-29
US
Fresenius Medical Care Holdings, Inc. recalls Blood Pump Rotor, Model Number: F40015481 Rev A
An increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
Other drugs & medical recalls in the United States
- 2024-07-24
US
RemedyRepack Inc. recalls Potassium Chloride Micro 10mEq K (750 mg) Extended Release Capsules, 30-count blister card, Rx only, MFG: Glenmark, Mahwah, NJ 07430,…
CGMP Deviations: Out of specification for dissolution
US
Homeocare Laboratories, Inc. recalls STELLALIFE ADVANCED FORMULA Peppermint, VEGA Oral Care, Rinse, 16 fl oz (473 ml), Distributed by: StellaLife, 2875 NE 191 Street,…
Microbial Contamination of Non-Sterile Products: presence of Terribacillus species organism
US
Homeocare Laboratories, Inc. recalls STELLALIFE VEGA Oral Care, Spray, Unflavored, 1 fl oz (30 ml) bottles, Distributed by: StellaLife, 22875 NE 191 Street, Suite 500,…
Microbial Contamination of Non-Sterile Products: multiple Bacillus species organisms
US
US
US
Viatris Inc recalls Cardura XL (doxazosin) extended release tablets 8mg, 30-count bottles, Rx Only Made in Singapore, Distributed by Roerig Division of Pfizer…
Failed Impurities/Degradation Specifications