Fresenius Medical Care Holdings, Inc. recalls Blood Pressure Module BPM3 TM-2917B (Spare Part Numbers 191378 or 191389) with firmware version S005 installed in 2008…
Machines equipped with BPM3 Blood Pressure Modules (Blood pressure monitor TM-2917B with firmware version S005) will not record Blood Pressure measurements and does not alarm if the cuff tubing is kinked close to the…
Read the official notice →- Date
- 2024-09-18
- Country
- United States
- Brand
- Fresenius Medical Care Holdings
- Product
- Blood Pressure Module BPM3 TM-2917B (Spare Part Numbers 191378 or 191389) with firmware version S005 installed in 2008 Series Hemodialysis…
- Classification
- Class II
- Units
- US Nationwide. Canada.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Fresenius Medical Care Holdings recalls
- 2025-07-02
US
- 2025-06-25
US
Fresenius Medical Care Holdings, Inc. recalls 5008X HDF Hemodialysis Blood Tubing Set, 24 units per case, with the following models: a. 5008X Standard HDF Bloodline,…
Potential of blood leak occurring between the lower port of the venous chamber and the tubing include: hemodiafiltration (HDF), hemodialysis (HD), hemofiltration (HF), and isolated ultrafiltration.
FDA · DeviceFresenius Medical Care Holdings, Inc.US Nationwide distribution in the state of Massachusetts.Official notice →- 2025-01-15
US
Fresenius Medical Care Holdings, Inc. recalls Dialyzer Optiflux 160NRe
Potential for internal blood leaks due to cracked polyurethane
- 2024-07-24
US
Fresenius Medical Care Holdings, Inc. recalls DELFLEX PERITONEAL DIALYSIS SOLUTION With LOW MAGNESIUM / LOW CALCIUM 4.25% DEXTROSE and attached stay "safe Exchange Set,…
This product is being recalled due to the tube weld failure presents itself as a slow leak and can be difficult to detect.
- 2024-07-17
US
Fresenius Medical Care Holdings, Inc. recalls RTL190395: 2008K@HOME MACHINE, SHORT CAB,OLC/DP,HP - Blood Pump Rotor spare part, Model Number: F40015481 Rev A
An increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
FDA · DeviceFresenius Medical Care Holdings, Inc.Domestic: Nationwide Distribution. International: Canada.Official notice →- 2024-06-19
US
Fresenius Medical Care Holdings, Inc. recalls US AquaA 900H-Typ USA-3 x 208v-Intended for the purification of water to be used for hemodialysis Vivonic Part Numbers:…
Concentrate Flow was exceeding the maximum 19.9 bar set point. The standard default set point range is between 4 and 18.Error message (F020104, "concentrate pressure alarm limit exceeded") concentrate pressure is…
FDA · DeviceFresenius Medical Care Holdings, Inc.US Nationwide distribution in the states of AL, IN, AK, WI.Official notice →
Other drugs & medical recalls in the United States
- 2024-09-18
US
Medical Action Industries, Inc. 306 recalls Foam Protective Pouches, model numbers 6010 (50 each/case), 6010-1 (350 each /case), 6060 (8 rolls/case), 6070 (75…
The protective pouches were identified to have been improperly labeled with UDI that is associated with an incorrect GMDN (Global Medical Device Nomenclature) code. These blue foam pouches were misused due to…
FDA · DeviceMedical Action Industries, Inc. 306Nationwide distribution. International distribution to Canada.Official notice →US
Medtronic Perfusion Systems recalls Medtronic DLP Single Stage Venous Cannulae with right angle metal tip, REF 67312
During manufacturing, DLP Single Stage Venous Cannulae - straight tip (Model 66118) was incorrectly placed into a product labeled as DLP Single Stage Venous Cannulae - right angle metal tip (Model 67312).
FDA · DeviceMedtronic Perfusion SystemsWorldwide distribution - US Nationwide and the countries of Cambodia, Greece,…Official notice →US
Hitachi America, Ltd., Particle Therapy Division recalls Brand Name: PROBEAT Product Name: PROBEAT-FR Model/Catalog Number: N/A Software Version: N/A Product…
During positioning within the system's 3D to 3D matching mode while transferring CBCT images, when the Float Image Selection button is pressed, it may display CBCT images that are insufficient in number.
FDA · DeviceHitachi America, Ltd., Particle Therapy DivisionU.S. Nationwide distribution in the states of TN, TX, and WA.Official notice →US
BioFire Diagnostics, LLC recalls FILMARRAY 2.0 INSTRUMENT, REF: FLM2-ASY-0001, FLM2-ASY-0001-W
Some automated in vitro diagnostic modules and instruments manufactured or serviced between May 2022 and May 2024 were calibrated with internal equipment that may lead modules and instruments to report higher melting…
FDA · DeviceBioFire Diagnostics, LLCWorldwide - US Nationwide distribution including in the states of NC, TX, OH,…Official notice →US
MEDLINE INDUSTRIES, LP - Northfield recalls Medline convenience kits containing semi-rigid suction liners labeled as follows: a) AMB VITRECTOMY, Pack Number DYNJ59921B;…
XXX
FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldWorldwide distribution - US Nationwide and the countries of Canada, Panama.Official notice →US
MEDLINE INDUSTRIES, LP - Northfield recalls Medline convenience kits containing semi-rigid suction liners labeled as follows: MTHZ GYN LAPAROSCOPY, Pack Number DYNJ909739
Due to reported adverse events, Medline Industries, LP. is issuing a recall on specific SKUs and lots of 1500 cc semi-rigid liners manufactured from 09/2023-11/2023 due to the potential loss of suction or low suction…
FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldWorldwide distribution - US Nationwide and the countries of Canada, Panama.Official notice →