Covidien Llc recalls Covidien Emprint Ablation Catheter with Thermosphere Technology Bronchoscopic Ablation, CFN CA108L1
The firm identified an error in the device temperature monitoring circuitry which is designed to interrupt energy delivery in the event of a primary cooling system failure. Use of an affected device in conjunction with…
Read the official notice →- Date
- 2021-11-17
- Country
- United States
- Brand
- Covidien
- Product
- Covidien Emprint Ablation Catheter with Thermosphere Technology Bronchoscopic Ablation, CFN CA108L1
- Classification
- Class II
- Units
- No domestic distribution. Foreign distribution to Hong Kong and UK.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Covidien recalls
- 2022-07-13
US
Covidien Llc recalls Palindrome Precision Chronic Catheter, Sterile, Single Use, Rx Only. For vascular access for acute and chronic hemodialysis, apheresis,…
There is a potential leaking condition within the hub of specific chronic dialysis catheters, due to an inter-lumen void in the hub component, which may result in unanticipated fluid return during use.
FDA · DeviceCovidien LlcWorldwide distribution - US Nationwide distribution in the states of AK, AL,…Official notice →- 2022-02-09
US
Covidien Llc recalls Chameleon PTA Balloon Dilation Catheter 8 mm x 40 mm, Chameleon PTA Balloon Catheter with Injection Port 8 mm x 40 mm; Cardiovascular.…
The firm has received six (6) complaint reports from customers indicating difficulty removing the balloon protector sleeve from the catheter (e.g., excessive force required to remove the protector sleeve). If the…
FDA · DeviceCovidien LlcWorldwide distribution - US Nationwide and the countries of Germany, Israel,…Official notice →- 2022-01-26
US
Covidien, LP recalls SonarMed AirWave Monitor Kit (Monitor, Sensor Cable, Power Cord), Item Code AW-M0001. used with endotracheal tube (ETT)
Prior to use on a patient, the device alarm speaker failed to annunciate on power up and the device falsely detected a sensor cable disconnect and reconnect during setup. The monitor detection of a false sensor cable…
FDA · DeviceCovidien, LPThe devices were distributed to the following US states: CA, CO, FL, IL, KY,…Official notice →- 2022-01-19
US
Covidien Llc recalls LigaSure Blunt Tip Laparoscopic Sealer/Divider, Nano-Coated, Product Number LF1837
Customers reported that the device jaws were difficult to open or would not open following application on tissue. Use of a device with this potential assembly issue could result in the potential for bleeding,…
FDA · DeviceCovidien LlcWorldwide Distribution in the countries of Australia, Austria, Belgium, Czech…Official notice →- 2021-12-15
US
Covidien Llc recalls Puritan Bennett 980 Series Ventilator, 980X2ESDIEC
Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use.
US
Covidien Llc recalls Covidien Sonicision Battery Charger-intended to charge the Sonicision Reusable Battery Pack and to provide information on the battery…
Intermittent risk of electric shock to users if an affected charger is used in conjunction with a faulty or bypassed institutions Mains Protective Earth (Ground) circuit
Other drugs & medical recalls in the United States
- 2021-11-17
US
Philips North America Llc recalls Philips Azurion Interventional Fluoroscopic X-Ray System, Software version 2.1.x, Model numbers 722063, 722064, 722067, 722068, 722078,…
When the user presses or releases both the APC (accept) button and the Float Tabletop (panning) button at the same time, the geometry may stop reacting on movement requests (table lock-up) and X-ray imaging becomes…
FDA · DevicePhilips North America LlcDomestic distribution nationwide. International distribution worldwide.Official notice →US
Therakos, Inc. recalls THERAKOS CELLEX Photopheresis Procedural Kit PRODUCT CODE: CLXUSA
Product was released prior to all testing being completed. If used and a drive tube leak were to occur, may cause illness or injury.
US
Z-Medica, LLC recalls QuikClot Combat Gauze-For Temporary External Use To Control Traumatic Bleeding Model / Part# 200
Lack of Packaging seal integrity may result in a sterile barrier breach.
US
Zhejiang Kindly Medical Devices Co., Ltd. recalls HENRY SCHEIN Safety Blood Collection Needle 21Gx 1 1/4 Inch, REF 570-2067, 100pcs/box 1000pcs/ctn 350*210*350mm
Potential for difficulty in safety mechanism activation.
US
ELLUME LTD recalls Ellume COVID-19 Home Test
Due to a higher rate of false positive test results.
US
SUN PHARMACEUTICAL INDUSTRIES INC recalls Tadalafil Tablets, USP 5 mg, 30 count bottle, Rx Only, Manufactured for: Northstar Rx LLC Memphis, TN 38141, Manufactured by:…
Incorrect Product Formulation: An incorrect grade of Crospovidone was used to manufacture the product.
FDA · DrugSUN PHARMACEUTICAL INDUSTRIES INCDistributed to one distributor who may have further distribute the product…Official notice →