Drugs & medicalClass II

Therakos, Inc. recalls THERAKOS CELLEX Photopheresis Procedural Kit PRODUCT CODE: CLXUSA

Product was released prior to all testing being completed. If used and a drive tube leak were to occur, may cause illness or injury.

Read the official notice →
Date
2021-11-17
Country
United States
Brand
Therakos, Inc.
Product
THERAKOS CELLEX Photopheresis Procedural Kit PRODUCT CODE: CLXUSA
Classification
Class II
Units
IL
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

Other drugs & medical recalls in the United States

  1. 2021-11-17
  2. US
    Drugs & medicalClass II

    Baxter Healthcare Corporation recalls Hemodialysis Delivery System, Software Version 2.x.

    If the operator initiates therapy with a saved prescription profile and makes a change to the prescription after a disposable filter change using the Same Patient button, the system may suggest values from the original…

    FDA · DeviceBaxter Healthcare CorporationDistributed nationwide to AL, AR, AZ, CA, CO, CT, FL, GA, HI, ID, IL, IN, KY,…Official notice →
  3. US
    Drugs & medicalClass II

    BioCheck, Inc. recalls T4 EIA Kit, Cat No: BC-1007, Store at 2-8C, For in vitro diagnostic use only This product s intended use is For the quantitative…

    Immunoassay TEST Kit do not meet product shelf-life standards resulting in failed controls.

    FDA · DeviceBioCheck, Inc.US Distribution to states of: MI, FL, NJ, OH, CA; and OUS (foreign)…Official notice →
  4. US
    Drugs & medicalClass II

    Philips North America Llc recalls Philips Azurion Interventional Fluoroscopic X-Ray System, Software version 2.1.x, Model numbers 722063, 722064, 722067, 722068, 722078,…

    When the user presses or releases both the APC (accept) button and the Float Tabletop (panning) button at the same time, the geometry may stop reacting on movement requests (table lock-up) and X-ray imaging becomes…

    FDA · DevicePhilips North America LlcDomestic distribution nationwide. International distribution worldwide.Official notice →
  5. US
    Drugs & medicalClass II

    Z-Medica, LLC recalls QuikClot Combat Gauze-For Temporary External Use To Control Traumatic Bleeding Model / Part# 200

    Lack of Packaging seal integrity may result in a sterile barrier breach.

    FDA · DeviceZ-Medica, LLCUS Nationwide distribution in the state of NC.Official notice →
  6. US
    Drugs & medicalClass II

    Covidien Llc recalls Covidien Emprint Ablation Catheter with Thermosphere Technology Bronchoscopic Ablation, CFN CA108L1

    The firm identified an error in the device temperature monitoring circuitry which is designed to interrupt energy delivery in the event of a primary cooling system failure. Use of an affected device in conjunction with…

    FDA · DeviceCovidien LlcNo domestic distribution. Foreign distribution to Hong Kong and UK.Official notice →
  7. US
    Drugs & medicalClass II

    Zhejiang Kindly Medical Devices Co., Ltd. recalls HENRY SCHEIN Safety Blood Collection Needle 21Gx 1 1/4 Inch, REF 570-2067, 100pcs/box 1000pcs/ctn 350*210*350mm

    Potential for difficulty in safety mechanism activation.

    FDA · DeviceZhejiang Kindly Medical Devices Co., Ltd.U.S. Nationwide distribution.Official notice →