Covidien Llc recalls LigaSure Blunt Tip Laparoscopic Sealer/Divider, Nano-Coated, Product Number LF1837
Customers reported that the device jaws were difficult to open or would not open following application on tissue. Use of a device with this potential assembly issue could result in the potential for bleeding,…
Read the official notice →- Date
- 2022-01-19
- Country
- United States
- Brand
- Covidien
- Product
- LigaSure Blunt Tip Laparoscopic Sealer/Divider, Nano-Coated, Product Number LF1837
- Classification
- Class II
- Units
- Worldwide Distribution in the countries of Australia, Austria, Belgium, Czech…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Covidien recalls
- 2022-07-13
US
Covidien Llc recalls Palindrome Precision Chronic Catheter, Sterile, Single Use, Rx Only. For vascular access for acute and chronic hemodialysis, apheresis,…
There is a potential leaking condition within the hub of specific chronic dialysis catheters, due to an inter-lumen void in the hub component, which may result in unanticipated fluid return during use.
FDA · DeviceCovidien LlcWorldwide distribution - US Nationwide distribution in the states of AK, AL,…Official notice →- 2022-02-09
US
Covidien Llc recalls Chameleon PTA Balloon Dilation Catheter 8 mm x 40 mm, Chameleon PTA Balloon Catheter with Injection Port 8 mm x 40 mm; Cardiovascular.…
The firm has received six (6) complaint reports from customers indicating difficulty removing the balloon protector sleeve from the catheter (e.g., excessive force required to remove the protector sleeve). If the…
FDA · DeviceCovidien LlcWorldwide distribution - US Nationwide and the countries of Germany, Israel,…Official notice →- 2022-01-26
US
Covidien, LP recalls SonarMed AirWave Monitor Kit (Monitor, Sensor Cable, Power Cord), Item Code AW-M0001. used with endotracheal tube (ETT)
Prior to use on a patient, the device alarm speaker failed to annunciate on power up and the device falsely detected a sensor cable disconnect and reconnect during setup. The monitor detection of a false sensor cable…
FDA · DeviceCovidien, LPThe devices were distributed to the following US states: CA, CO, FL, IL, KY,…Official notice →- 2021-12-15
US
Covidien Llc recalls Puritan Bennett 980 Series Ventilator, 980X2ESDIEC
Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use.
US
Covidien Llc recalls Covidien Sonicision Battery Charger-intended to charge the Sonicision Reusable Battery Pack and to provide information on the battery…
Intermittent risk of electric shock to users if an affected charger is used in conjunction with a faulty or bypassed institutions Mains Protective Earth (Ground) circuit
- 2021-11-17
US
Covidien Llc recalls Covidien Emprint Ablation Catheter with Thermosphere Technology Bronchoscopic Ablation, CFN CA108L1
The firm identified an error in the device temperature monitoring circuitry which is designed to interrupt energy delivery in the event of a primary cooling system failure. Use of an affected device in conjunction with…
FDA · DeviceCovidien LlcNo domestic distribution. Foreign distribution to Hong Kong and UK.Official notice →
Other drugs & medical recalls in the United States
- 2022-01-26
US
Teva Pharmaceuticals USA recalls Brinzolamide Ophthalmic Suspension, USP 1%, packaged as a)10 ml dropper bottles (NDC 0591-2127-79), and b) 15 ml dropper bottles (NDC…
Defective Container: The notch in the cap that fits into the nozzle of the dropper could break off and block the dropper possibly resulting in the product not dispensing.
US
Geri-Care Pharmaceuticals Recalls Over-the-Counter Drugs Due to Failure to Meet Child Resistant Packaging Requirement; Risk of Poisoning
The over-the-counter drug products contain regulated substances which must be in child resistant packaging as required by the Poison Prevention Packaging Act (PPPA). The packaging of the products is not child…
- 2022-01-19
US
Gentherm Medical, LLC recalls Electri-Cool II Portable Cold Therapy Unit, Model Number 767, Part Number 86193
This recall has been initiated to provide a labeling update with revised instruction how to operate the unit with Cold Therapy pads.
FDA · DeviceGentherm Medical, LLCWorldwide distribution - US Nationwide and the countries of Canada, Chile,…Official notice →US
Canon Medical System, USA, INC. recalls Flat Panel Detector, Model Number: TFP-800A, used with the Alphenix Systems, Item Numbers: INFX-8000V, INFX-8000C, INFX-8000F,…
Abnormal images such as horizontal stripes may be displayed and recovery may not not be possible by restarting the system. The worst case scenario is an operator who is working without a workable image may remove a…
FDA · DeviceCanon Medical System, USA, INC.US Nationwide distribution in the states of DE, FL, LA, MT, NJ, OH, and WV.Official notice →US
Medtronic Perfusion Systems recalls Medtronic Custom Perfusion kits, with the following product descriptions and corresponding model numbers: 1. CUSTOM PACK BB10Q85R…
Product is labeled as non-pyrogenic but endotoxin testing was not performed.
FDA · DeviceMedtronic Perfusion SystemsWorldwide distribution - US Nationwide distribution in the states of AL, AR,…Official notice →US
DePuy Orthopaedics, Inc. recalls GLOBAL UNITE REV STEM SZ 10 Product Code: 110010600
Stems may have an undersized spigot bore diameter therefore separating the proximal body from the stem may not be possible without surgical delay and/or bone damage
FDA · DeviceDePuy Orthopaedics, Inc.Worldwide distribution - US Nationwide and the countries of Austria,…Official notice →