B. Braun Medical, Inc. recalls Epidural tray. PERIFIX 17 Ga. x 3-1/2 in. (90 mm) Tuohy, 5 mL Glass Luer Lock LOR Tray (Kit) A collection of devices intended for the…
Kits should contain Filter Straws with Standard Luer Connections however, the kits contain Filter Straws with Neuraxial connectors preventing interface with the other components of the kit.
Read the official notice →- Date
- 2024-02-28
- Country
- United States
- Brand
- B. Braun Medical
- Product
- Epidural tray. PERIFIX 17 Ga. x 3-1/2 in. (90 mm) Tuohy, 5 mL Glass Luer Lock LOR Tray (Kit) A collection of devices intended for the…
- Classification
- Class II
- Units
- US Nationwide
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More B. Braun Medical recalls
- 2024-06-19
US
B Braun Medical Inc recalls Stimuplex A, 30 DEG, 24GX1", 0.55x25mm, Catalogue Number: 4894251.
The devices are labeled "not made with DEHP" or the DEHP free symbol; however, the glue used contains traces of DEHP.
FDA · DeviceB Braun Medical IncWorldwide distribution - US Nationwide and the country of Canada.Official notice →- 2024-04-10
US
B. Braun Medical Inc recalls Lactated Ringer's Injection USP, 1000mL, EXCEL CONTAINER, Rx only, B.Braun Medical Inc., Bethlehem, PA 18018, NDC 0264-7750-00
Lack of assurance of sterility: bags have the potential to leak..
- 2024-03-20
US
B. Braun Medical Inc recalls Potassium Chloride for Injection Concentrate USP, 500 mEq/250 mL (2 mEq/mL), 250 mL Bag, Rx only, B. Braun Medical Inc., Bethlehem, PA…
Lack of assurance of sterility: pinholes, within the blue label characters on the EXCEL bag, specifically within the dotted characters on the label, resulting in leaks.
- 2024-02-28
US
B. Braun Medical, Inc. recalls Epidural tray. PERIFIX 17 Ga. x 3-1/2 in. (90 mm) Tuohy, PERIFIX FX 19 Ga. Springwound Closed Tip Catheter, 5 mL Glass Luer Lock LOR Tray…
Kits should contain Filter Straws with Standard Luer Connections however the kits contain Filter Straws with Neuraxial connectors preventing interface with the other components of the kit.
- 2023-12-13
US
B. Braun Medical, Inc. recalls Infusomat Administration Sets as listed below: 362031 Infusomat Space Pump Set with Caresite Injection Site; 362032 15 drops/mL, priming…
In affected products, the tubing outer diameter may vary in size or the length of tubing may be too short. Both issues may lead to false air-in-line alarms, which will stop the infusion and interrupt therapy.…
- 2023-11-15
US
B. Braun Medical, Inc. recalls TERUMO 60" TUBE SET W/HIGH CRACK CHK VLV, Article No. 7S3070. Singapore only. IV injection set.
Devices may leak as a result of a manufacturing issue. Leakage may result in a delay of therapy, increase risk of bloodstream infection (BSI), and may expose both patients and healthcare workers to hazardous…
FDA · DeviceB. Braun Medical, Inc.Worldwide distribution - US Nationwide and the countries of Canada and…Official notice →
Other drugs & medical recalls in the United States
- 2024-02-28
US
Canon Medical System, USA, INC. recalls The INFX-8000F is designed to take advantage of the latest technological innovations to reduce dose and to save time. A…
VAREX changed the manufacturing of the internal board, and the detectors displayed abnormal images. Internal board of the X-ray flat panel detector may display abnormal images, such as horizontal stripes, which may…
US
Boston Scientific Corporation recalls EndoVive Replacement Button Kit w/ ENFIT 28F 2.8 CM, Outer Box UPN M00509500 or M00509510.
Affected batches of the kit contain Stoma Measuring Devices sized to measure for an 18/24F replacement button, but are labeled as being for a 28F. Use of defective product may lead to undersizing the button, which may…
FDA · DeviceBoston Scientific CorporationInternational distribution to the country of Japan only.Official notice →US
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Azurion Interventional Fluoroscopic X-ray Systems, specifically the Disk Bay of the PC component of the affected…
Disk Bay may not perform as intended. System may stop functioning and imaging may not be possible, resulting in delay of procedure.
FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Worldwide Distribution: US nationwide including states of: AL, AK, AR, AZ, CA,…Official notice →US
B. Braun Medical, Inc. recalls Epidural tray. PERIFIX 17 Ga. x 3-1/2 in. (90 mm) Tuohy, PERIFIX FX 19 Ga. Springwound Closed Tip Catheter, 5 mL Glass Luer Lock LOR Tray…
Kits should contain Filter Straws with Standard Luer Connections however the kits contain Filter Straws with Neuraxial connectors preventing interface with the other components of the kit.
US
AMD Medicom Inc. recalls PATTERSON¿ PROCEDURE EARLOOP MASKS- ANTI-FOG, ASTM LEVEL 3 (BLUE) (item code 127-7573)
Level 3 masks were produced on alternate non-validated production equipment that were not included as part of the initial qualification of the products.
FDA · DeviceAMD Medicom Inc.US Nationwide distribution in the states of FL, IA, IN, PA, NV, TX.Official notice →US
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Azurion Interventional Fluoroscopic X-ray Systems, specifically the Disk Bay of the PC component of the affected…
Framegrabber card may not perform as intended. Monitor may show no viewports, or show one or more viewports may show no image, a distorted image, or a frozen image. Switching between viewports on the monitor may not be…
FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Worldwide - US Nationwide distribution.Official notice →