B. Braun Medical, Inc. recalls Infusomat Administration Sets as listed below: 362031 Infusomat Space Pump Set with Caresite Injection Site; 362032 15 drops/mL, priming…
In affected products, the tubing outer diameter may vary in size or the length of tubing may be too short. Both issues may lead to false air-in-line alarms, which will stop the infusion and interrupt therapy.…
Read the official notice →- Date
- 2023-12-13
- Country
- United States
- Brand
- B. Braun Medical
- Product
- Infusomat Administration Sets as listed below: 362031 Infusomat Space Pump Set with Caresite Injection Site; 362032 15 drops/mL, priming…
- Classification
- Class I
- Units
- Distribution US nationwide, Canada and Germany.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More B. Braun Medical recalls
- 2024-03-20
US
B. Braun Medical Inc recalls Potassium Chloride for Injection Concentrate USP, 500 mEq/250 mL (2 mEq/mL), 250 mL Bag, Rx only, B. Braun Medical Inc., Bethlehem, PA…
Lack of assurance of sterility: pinholes, within the blue label characters on the EXCEL bag, specifically within the dotted characters on the label, resulting in leaks.
- 2024-02-28
US
B. Braun Medical, Inc. recalls Epidural tray. PERIFIX 17 Ga. x 3-1/2 in. (90 mm) Tuohy, 5 mL Glass Luer Lock LOR Tray (Kit) A collection of devices intended for the…
Kits should contain Filter Straws with Standard Luer Connections however, the kits contain Filter Straws with Neuraxial connectors preventing interface with the other components of the kit.
US
B. Braun Medical, Inc. recalls Epidural tray. PERIFIX 17 Ga. x 3-1/2 in. (90 mm) Tuohy, PERIFIX FX 19 Ga. Springwound Closed Tip Catheter, 5 mL Glass Luer Lock LOR Tray…
Kits should contain Filter Straws with Standard Luer Connections however the kits contain Filter Straws with Neuraxial connectors preventing interface with the other components of the kit.
- 2023-11-15
US
B. Braun Medical, Inc. recalls TERUMO 60" TUBE SET W/HIGH CRACK CHK VLV, Article No. 7S3070. Singapore only. IV injection set.
Devices may leak as a result of a manufacturing issue. Leakage may result in a delay of therapy, increase risk of bloodstream infection (BSI), and may expose both patients and healthcare workers to hazardous…
FDA · DeviceB. Braun Medical, Inc.Worldwide distribution - US Nationwide and the countries of Canada and…Official notice →- 2023-11-01
US
B. Braun Medical, Inc. recalls 8713051U, Infusomat Large Volume Pump, Wireless; Volumetric Infusion Pump System
Upstream occlusion alarm may sound when no occlusion exists, and the device will stop pumping. Interruption to the infusion of high-risk medications may lead to hemodynamic instability requiring medical intervention to…
FDA · DeviceB. Braun Medical, Inc.Domestic distribution nationwide. Foreign distribution to Canada.Official notice →- 2023-10-04
US
B. Braun Medical, Inc. recalls B.Braun NORM-JECT Luer Lock Solo, 10 ml, Sterile-Intended to be used to deliver (inject) or withdraw an accurate metered dose of a fluid…
The sterile blister packaging may be damaged, and sterility may be compromised
Other drugs & medical recalls in the United States
- 2023-12-13
US
Carestream Health, Inc. recalls DRX Revolution Mobile X-Ray System
Unexpected failure of electrical components within the CPI generator.
FDA · DeviceCarestream Health, Inc.Worldwide Distribution: US (Nationwide Distribution); and OUS (Foreign):…Official notice →US
Stryker Medical Division of Stryker Corporation recalls ProCuity bed series, model number 3009, item number: 3009PX-LE-400.
Patient beds are missing test values for electrical safety tests presenting risk of tissue burn or electrical shock.
FDA · DeviceStryker Medical Division of Stryker CorporationDistribution US Nationwide including Puerto Rico, Australia, Canada, Chile,…Official notice →US
Maquet Medical Systems USA recalls CARDIOHELP Emergency Drive, Part Numbers 701048002 and 701076205. The CARDIOHELP System is a blood oxygenation and carbon dioxide…
Product removal due to possible blocking or impairment of the CARDIOHELP Emergency Drive. This issue can lead to the patient being exposed to inappropriate low blood flow. Potential associated harms are ischemia and…
FDA · DeviceMaquet Medical Systems USAUS Nationwide - Worldwide Distribution: Domestic distribution to: AL, CA, DC,…Official notice →US
STAQ Pharma, Inc. recalls ROPivacaine HCl 0.2% PF in Sodium Chloride 1,000 mg/500 mL (2 mg per mL) Injection, 500 mL bags, Rx only, STAQ Pharma Inc. 14135 E 42nd…
STAQ Pharma Inc. received consumer complaints related to leaking bags and other potential bag concerns. Upon investigation, they determine the source of the issue was with the bag manufacturer who released bags Lot #…
US
US
Novartis Pharmaceuticals Corporation recalls SANDIMMUNE Oral Solution (cyclosporine oral solution, USP) 100 mg/mL, 50 mL bottle, Rx Only, Manufactured by: DELPHARM…
Crystallization: bottles of Sandimmune Oral Solution were determined to contain crystals