Drugs & medicalClass II

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Azurion Interventional Fluoroscopic X-ray Systems, specifically the Disk Bay of the PC component of the affected…

Framegrabber card may not perform as intended. Monitor may show no viewports, or show one or more viewports may show no image, a distorted image, or a frozen image. Switching between viewports on the monitor may not be…

Read the official notice →
Date
2024-02-28
Country
United States
Brand
Philips Medical Systems Nederland
Product
Philips Azurion Interventional Fluoroscopic X-ray Systems, specifically the Disk Bay of the PC component of the affected system. Affected…
Classification
Class II
Units
Worldwide - US Nationwide distribution.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Philips Medical Systems Nederland recalls

  1. 2024-02-28
  2. US
    Drugs & medicalClass II

    PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Azurion 7 with a Certeray generator-To perform image guidance in diagnostic, interventional, and minimally invasive…

    Generator may fail due to a potential short circuit in the Printed Circuit Board Assembly (PCBA) in PoInt EVR (Power Invertor). If a short circuit occurs, it will cause the fuses to trip and the system will become…

    FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Worldwide distribution - US Nationwide distribution in the states of NY, PA,…Official notice →
  3. US
    Drugs & medicalClass II

    PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Allura Xper Interventional Fluoroscopic X-ray Systems, specifically the DIMMs (Dual In-line Memory Modules) used…

    Dual In-Line Memory Modules (DIMMs) may not perform as intended. System may stop functioning and imaging may not be possible, resulting in delay of procedure.

    FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Worldwide Distribution: US nationwide including states of: AL, AK, AR, AZ, CA,…Official notice →
  4. US
    Drugs & medicalClass II

    PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Azurion Interventional Fluoroscopic X-ray Systems, specifically the Disk Bay of the PC component of the affected…

    Disk Bay may not perform as intended. System may stop functioning and imaging may not be possible, resulting in delay of procedure.

    FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Worldwide Distribution: US nationwide including states of: AL, AK, AR, AZ, CA,…Official notice →
  5. 2024-02-07
  6. US
    Drugs & medicalClass II

    PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Azurion 7 M20 FlexArm system (Software version R2.x) with Maquet Magnus Table (Software version 05.02.12)

    Loss of connectivity between the FlexArm and the Table due to a software issue related to the date/time format. This connection issue leads to a loss of movement of the arm or table.

    FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Worldwide - US Nationwide distribution in the states of CO, FL, IL, MS, NV and…Official notice →
  7. US
    Drugs & medicalClass II

    PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Azurion 7 M20 FlexArm system (Software version R2.x) with Maquet Magnus Table (Software version 05.02.12)

    Loss of connectivity between the FlexArm and the Table due to a software issue.

    FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Worldwide - US Nationwide distribution in the states of CO, FL, IL, MS, NV and…Official notice →
  8. US
    Drugs & medicalClass II

    PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Azurion system, Interventional fluoroscopic x-ray system. System Product Name - Model Numbers Azurion 3M12…

    System may exhibit a loss of imaging functionality and data due to an issue which can cause the system to continuously restart. If the issue occurs during a procedure, there may be a delay and/or abortion of the…

    FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Worldwide - US Nationwide distribution.Official notice →
All 59 Philips Medical Systems Nederland recalls →

Other drugs & medical recalls in the United States

  1. 2024-02-28
  2. US
    Drugs & medicalClass II

    B. Braun Medical, Inc. recalls Epidural tray. PERIFIX 17 Ga. x 3-1/2 in. (90 mm) Tuohy, 5 mL Glass Luer Lock LOR Tray (Kit) A collection of devices intended for the…

    Kits should contain Filter Straws with Standard Luer Connections however, the kits contain Filter Straws with Neuraxial connectors preventing interface with the other components of the kit.

    FDA · DeviceB. Braun Medical, Inc.US NationwideOfficial notice →
  3. US
    Drugs & medicalClass II

    B. Braun Medical, Inc. recalls Epidural tray. PERIFIX 17 Ga. x 3-1/2 in. (90 mm) Tuohy, PERIFIX FX 19 Ga. Springwound Closed Tip Catheter, 5 mL Glass Luer Lock LOR Tray…

    Kits should contain Filter Straws with Standard Luer Connections however the kits contain Filter Straws with Neuraxial connectors preventing interface with the other components of the kit.

    FDA · DeviceB. Braun Medical, Inc.US NationwideOfficial notice →
  4. US
    Drugs & medicalClass II

    AMD Medicom Inc. recalls PATTERSON¿ PROCEDURE EARLOOP MASKS- ANTI-FOG, ASTM LEVEL 3 (BLUE) (item code 127-7573)

    Level 3 masks were produced on alternate non-validated production equipment that were not included as part of the initial qualification of the products.

    FDA · DeviceAMD Medicom Inc.US Nationwide distribution in the states of FL, IA, IN, PA, NV, TX.Official notice →
  5. US
    Drugs & medicalClass II

    Voco GmbH recalls Ionolux, Catalog (REF) Numbers: a) 2115, b)2117, c) 2118, d) 2119, e) 2120, f) 2122; dental cement

    Due to an error in the manufacturing process, it is possible that an incorrect plunger has been fitted to some capsules. This plunger does not contain the correct reactive component for Ionolux. When these capsules are…

    FDA · DeviceVoco GmbHWorldwide distribution: US distribution to states of: IA, NY, PA, WA; and OUS…Official notice →
  6. US
    Drugs & medicalClass I

    Windstone Medical Packaging, Inc. recalls Aligned Medical Solutions, In House Ocular Pack, AMS12947, STERILEEO

    Sterility assurance for saline included in surgical kits cannot be guaranteed.

    FDA · DeviceWindstone Medical Packaging, Inc.US: FL, IL, CA OUS: NoneOfficial notice →
  7. US
    Drugs & medicalClass I

    Cardinal Health 200, LLC recalls Dover" IRRIGATION 1200 CC TRAY, 60 CC PISTON SYRINGE, STERILE SALINE, TIP PROTECTOR, ALCOHOL, LIDDED

    Convenience kits were manufactured and distributed with components which were subsequently recalled by Nurse Assist.

    FDA · DeviceCardinal Health 200, LLCUS and EMEA, Japan, Latin AmericaOfficial notice →