Drugs & medicalClass II

Abiomed, Inc. recalls Automated Impella Controller (AIC), Product No. 0042-0000-US, 0042-0010-US, 0042-0040-US. Used with Impella Ventricular Support Systems…

Retrospective reporting for the release of Technical Bulletin IMP-2643 AIC Version 8.5 Software Update Available. The update resolved an issue in version 8.4 where a pump was not recognized by the AIC.

Read the official notice →
Date
2024-02-28
Country
United States
Brand
Abiomed
Product
Automated Impella Controller (AIC), Product No. 0042-0000-US, 0042-0010-US, 0042-0040-US. Used with Impella Ventricular Support Systems…
Classification
Class II
Units
US Nationwide distribution.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Abiomed recalls

  1. 2025-01-22
  2. US
    Drugs & medicalClass I

    Abiomed, Inc. recalls Impella RP Flex with SmartAssist; Product Number: 1000323;

    Optical Sensors have been damaged due to physical interaction between the inlet and another device resulting in a Placement Signal Not Reliable (PSNR) alarm and loss of Central Venous (CV) Placement Signal, Pulmonary…

    FDA · DeviceAbiomed, Inc.Domestic: Nationwide Distribution International: CanadaOfficial notice →
  3. 2024-09-11
  4. US
    Drugs & medicalClass II

    Abiomed, Inc. recalls ImpellaCP SmartAssistSetAPAC, Delo1895; Product Code: 1000302;

    A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential…

    FDA · DeviceAbiomed, Inc.Worldwide distribution - US Nationwide and the countries of Australia,…Official notice →
  5. 2024-07-24
  6. US
    Drugs & medicalClass I

    Abiomed, Inc. recalls Impella CP with SmartAssist, Product No.: 1000080 (REF 0048-0003). left heart support blood pump

    Nine (9) Impella CP pumps failed inspection and were inadvertently released.

    FDA · DeviceAbiomed, Inc.Domestic only: FL, MA, OH TX.Official notice →
  7. 2024-02-21
  8. US
    Drugs & medicalClass II

    Abiomed, Inc. recalls Impella catheters - Intravascular micro axial blood pumps that support a patient's circulatory system. (1) Product Code 005042 - Impella…

    New warnings are being added to the Instructions for Use for all affected products to reduce the possibility of fibers being drawn into the Impella, which may result in low flow of the device.

    FDA · DeviceAbiomed, Inc.Worldwide distribution - US Nationwide and the countries of AU, CA, DE, FR,…Official notice →
  9. US
    Drugs & medicalClass I

    Abiomed, Inc. recalls Impella catheters - Intravascular micro axial blood pumps that support a patient's circulatory system. (1) Product Code 005042 - Impella…

    IFU has been updated to include warnings about the risk of the inlet perforating through the myocardial wall of the left ventricle due to operator handling.

    FDA · DeviceAbiomed, Inc.Worldwide distribution - US Nationwide and the countries of AU, CA, DE, FR,…Official notice →
  10. 2024-01-17
  11. US
    Drugs & medicalClass II

    Abiomed, Inc. recalls Abiomed Kit, 14Fr Introducer, 13cm&25cm, Sterile Part of Pump Set (0048-0040, 0048-0045, 1000080)-Introducer(s) are intended for…

    Hole(s) to the outer pouch of the Introducer Kits (both trayed and non-trayed) may compromise the sterile barrier of the Introducer Kits. Non-sterile product exposes patients to the possibility of the introduction of…

    FDA · DeviceAbiomed, Inc.Nationwide Foreign: Country AT AU BE BR CA CA CH CZ DE DK ES FR FR HK IN IT MX…Official notice →
All 30 Abiomed recalls →

Other drugs & medical recalls in the United States

  1. 2024-02-28
  2. US
    Drugs & medicalClass II

    MEDLINE INDUSTRIES, LP - Northfield recalls These wipes are saturated with 0.65% bleach solution. These wipes are antimicrobial, disinfectant with sporicidal, virucidal,…

    There is stability concern for products dated with a 24-month shelf life. Product is safe and effective up to 12 months shelf life.

    FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldUS and territories, S. Korea, Panama, and DubaiOfficial notice →
  3. US
    Drugs & medicalClass II

    W.L. Gore & Associates, Inc. recalls GORE VIABIL Short Wire Biliary Endoprosthesis, Product labeled as (1) Catalog Number VSWVH1008, 10 mm x 8 cm; with holes; and (2)…

    Devices without transmural drainage holes may be mislabeled as devices with transmural drainage holes, or vice versa. Potential health impact includes extending procedure time, unplanned device removal, secondary…

    FDA · DeviceW.L. Gore & Associates, Inc.US Nationwide distribution in the state of Georgia.Official notice →
  4. US
    Drugs & medicalClass II

    PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Azurion 7 with a Certeray generator-To perform image guidance in diagnostic, interventional, and minimally invasive…

    Generator may fail due to a potential short circuit in the Printed Circuit Board Assembly (PCBA) in PoInt EVR (Power Invertor). If a short circuit occurs, it will cause the fuses to trip and the system will become…

    FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Worldwide distribution - US Nationwide distribution in the states of NY, PA,…Official notice →
  5. US
    Drugs & medicalClass II

    MRP, LLC dba Aquabiliti recalls AQUABILITI AQUASTAT 0.9% Sodium Chloride Injection syringe, USP, 5mL in 12mL, REF 2T0805; and 3mL in 12mL, REF 2T0807, intermediate…

    The saline flush syringes may fail to meet USP criteria for stated sodium chloride content after 12 months.

    FDA · DeviceMRP, LLC dba AquabilitiUS Nationwide distribution in the states CO, FL, NC, NY, OH, PA, TN, TX, and VA.Official notice →
  6. US
    Drugs & medicalClass II

    IMPULSE DYNAMICS (USA) INC recalls OPTIMIZER model CCM X11 implantable pulse generator (IPG) devices - Smart Mini and Lite

    OPTIMIZER devices may cease to deliver CCM therapy if the device incorrectly detects a charging error. This may cause patients to experience heart failure symptoms similar to before implantation of the device.

    FDA · DeviceIMPULSE DYNAMICS (USA) INCDomestic US distribution nationwide. International distribution to Austria…Official notice →
  7. US
    Drugs & medicalClass II

    FUJIFILM Healthcare Americas Corporation recalls Synapse Cardiology PACS V7.3.0, V7.2, V7.1, V7.0. A web-based application as the primary user interface for the…

    The LV Mass (2D Bullet) equation may be calculated incorrectly, resulting in a variance in the mass of the left ventricle (LV Mass). If used as a main factor for diagnosis, there is a risk of misdiagnosis or incorrect…

    FDA · DeviceFUJIFILM Healthcare Americas CorporationUS domestic distribution nationwide. International distribution to Israel.Official notice →
See all drugs & medical recalls →