FUJIFILM Healthcare Americas Corporation recalls Synapse Cardiology PACS V7.3.0, V7.2, V7.1, V7.0. A web-based application as the primary user interface for the…
The LV Mass (2D Bullet) equation may be calculated incorrectly, resulting in a variance in the mass of the left ventricle (LV Mass). If used as a main factor for diagnosis, there is a risk of misdiagnosis or incorrect…
Read the official notice →- Date
- 2024-02-28
- Country
- United States
- Brand
- FUJIFILM Healthcare Americas
- Product
- Synapse Cardiology PACS V7.3.0, V7.2, V7.1, V7.0. A web-based application as the primary user interface for the processing of medical…
- Classification
- Class II
- Units
- US domestic distribution nationwide. International distribution to Israel.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More FUJIFILM Healthcare Americas recalls
- 2024-06-05
US
FUJIFILM Healthcare Americas Corporation recalls FDR Visionary Suite - Intended to generate digital or conventional radiographic images of the skull, spinal column,…
Bolts on the CH-200 tube may rupture due to fatigue failure, causing the tube support to fall, could result in patient injury
FDA · DeviceFUJIFILM Healthcare Americas CorporationUS Nationwide distribution in the states of AK, CA, FL, IL, NE, NH, OH, TN, WI.Official notice →- 2024-05-29
US
FUJIFILM Healthcare Americas Corporation recalls Noblus AC Adapter used with the Noblus Ultrasound Imaging System Model : AHM250PS24
AC adapters of Noblus can heat up, and cause smoke in the worst case, when an excessive mechanical stress is applied to the socket for power cable (frequent insertion/removal of the power cable).
- 2024-03-13
US
FUJIFILM Healthcare Americas Corporation recalls Synapse PACS - Version 7.2.100
Measurements on a Secondary Capture 2D image, that does not have pixel spacing in the DICOM header, when combined with a Breast Tomo Series results in incorrect measurements.
FDA · DeviceFUJIFILM Healthcare Americas CorporationWorldwide distribution - US Nationwide, including Puerto Rico and the…Official notice →- 2023-05-03
US
FUJIFILM Healthcare Americas Corporation recalls FujiFilm FDR AQRO (Model DR-XD1000)- A digital mobile X-ray system intended for use in general purpose radiography for…
Arm which holds the tube head in place is susceptible to breaking, if the device breaks it could potentially cause serious harm or injury to the patient and/or user.
- 2022-09-21
US
FUJIFILM Healthcare Americas Corporation recalls Surpria 64: Software Version V3.11, V3.22
Error 00003050 or 00003052 may occur during scanning, the captured images will not be reconstructed and the Raw Data will not be displayed in the work list.result, may result in the need for additional scanning to…
- 2021-11-10
US
FujiFilm Healthcare Americas Corporation recalls Arietta 65 Ultrasound Software Versions: V1.0 through V4.0.0
When the ultrasound system is used in combination with a Fujifilm (formally Hitachi) transducer, there is an error in the calculation formula of Effective Regurgitant Orifice Area (EROA) when you perform MR Flow…
Other drugs & medical recalls in the United States
- 2024-02-28
US
Abiomed, Inc. recalls Automated Impella Controller (AIC), Product No. 0042-0000-US, 0042-0010-US, 0042-0040-US. Used with Impella Ventricular Support Systems…
Retrospective reporting for the release of Technical Bulletin IMP-2643 AIC Version 8.5 Software Update Available. The update resolved an issue in version 8.4 where a pump was not recognized by the AIC.
US
MRP, LLC dba Aquabiliti recalls AQUABILITI AQUASTAT 0.9% Sodium Chloride Injection syringe, USP, 5mL in 12mL, REF 2T0805; and 3mL in 12mL, REF 2T0807, intermediate…
The saline flush syringes may fail to meet USP criteria for stated sodium chloride content after 12 months.
FDA · DeviceMRP, LLC dba AquabilitiUS Nationwide distribution in the states CO, FL, NC, NY, OH, PA, TN, TX, and VA.Official notice →US
IMPULSE DYNAMICS (USA) INC recalls OPTIMIZER model CCM X11 implantable pulse generator (IPG) devices - Smart Mini and Lite
OPTIMIZER devices may cease to deliver CCM therapy if the device incorrectly detects a charging error. This may cause patients to experience heart failure symptoms similar to before implantation of the device.
FDA · DeviceIMPULSE DYNAMICS (USA) INCDomestic US distribution nationwide. International distribution to Austria…Official notice →US
3M Company - Health Care Business recalls 3M Specialty Blade Assembly, REF 9660,70-2011-8462-2, 7100213114, used with used with 3M Surgical Clipper Catalog Number 9661L
3M has added the following cautionary statement regarding the use of the product: "Caution - To reduce the risk of burns, which if not avoided may result in minor injury: " Do not use on injured skin. " Do not leave…
US
W&H DENTALWERK BUERMOOS GMBH recalls W&H Surgical handpiece S-12. For Surgical treatment of dental hard tissue Model Number: 30061000
1:2 speed increasing surgical handpieces have been incorrectly laser marked during production. The specific laser marking should read 1:2, but instead it reads 1:1 on the body of the handpiece
FDA · DeviceW&H DENTALWERK BUERMOOS GMBHWorldwide distribution - US Nationwide and the countries of Canada, Australia,…Official notice →US
Olympus Corporation of the Americas recalls ESG PK CUTTING FORCEPS, 5MM, 33CM
The jaw may fracture prior to the procedure during the inspection instructed per the IFU, or during the procedure.
FDA · DeviceOlympus Corporation of the AmericasWorldwide distribution - US Nationwide and the countries of Australia, Canada,…Official notice →