Drugs & medicalClass II

W.L. Gore & Associates, Inc. recalls GORE VIABIL Short Wire Biliary Endoprosthesis, Product labeled as (1) Catalog Number VSWVH1008, 10 mm x 8 cm; with holes; and (2)…

Devices without transmural drainage holes may be mislabeled as devices with transmural drainage holes, or vice versa. Potential health impact includes extending procedure time, unplanned device removal, secondary…

Read the official notice →
Date
2024-02-28
Country
United States
Brand
W L Gore & Associates
Product
GORE VIABIL Short Wire Biliary Endoprosthesis, Product labeled as (1) Catalog Number VSWVH1008, 10 mm x 8 cm; with holes; and (2) Catalog…
Classification
Class II
Units
US Nationwide distribution in the state of Georgia.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More W L Gore & Associates recalls

  1. 2025-10-08
  2. US
    Drugs & medicalClass II

    W L Gore & Associates, Inc. recalls GORE ACUSEAL Vascular Graft, REF: ECH050020J, ECH050020W, ECH050050J, ECH050050W, ECH060010A, ECH060020A, ECH060020J, ECH060020W,…

    Reports of vascular graft delamination, which may lead to reduced access flow; cannulation difficulties; bleeding or bruising; and harms related to reintervention procedures, so Instructions for Use will be updated to…

    FDA · DeviceW L Gore & Associates, Inc.US: OR, AL, OH, MD, VA, CA, FL, NC, OK, MA, MO, AZ, TX, IL, GA, PA, NH, HI,…Official notice →
  3. 2024-09-11
  4. US
    Drugs & medicalClass II

    W L Gore & Associates, Inc. recalls Heparin, Gore Viabahn VBX, Balloon Expandable Endoprosthesis, for the following Reference numbers: United States: BXB051501A…

    Due to packaging defects, sterility assurance and heparin activity may be compromised.

    FDA · DeviceW L Gore & Associates, Inc.Worldwide - US Nationwide distribution including in the states of AK, AL, AR,…Official notice →
  5. 2024-08-14
  6. US
    Drugs & medicalClass II

    W.L. Gore & Associates, Inc. recalls GORE SEAMGUARD Bioabsorbable Staple Line Reinforcement Configured for Circular Staplers, 25mm; Catalog number: 1BSGC25;

    Devices were labeled with an expiration date of four years and ten days rather than the validated two years.

    FDA · DeviceW.L. Gore & Associates, Inc.US Nationwide distribution in the states of CA, GA, IL, NC, OH, OK, WA.Official notice →
  7. 2023-02-08
  8. US
    Drugs & medicalClass II

    W L Gore & Associates, Inc. recalls GORE CARDIOFORM Septal Occluder, REF: GSX0030A

    Septal Occluders released with incomplete quality test documentation that assesses the release of the device from the delivery system during the index procedure. A potential failure of this quality test could…

    FDA · DeviceW L Gore & Associates, Inc.US Nationwide distribution in the states of NY, CA, FL, CT, TX, AR, IN, MA,…Official notice →
  9. 2022-11-02
  10. US
    Drugs & medicalClass II

    W L Gore & Associates, Inc. recalls GORE CARDIOFORM ASD Occluder. cardiovascular implant.

    Due to manufacturing records (Release Test Results) indicating "Failed".

    FDA · DeviceW L Gore & Associates, Inc.Worldwide distribution - U.S. Nationwide distribution including in the states…Official notice →
  11. 2021-11-03
  12. US
    Drugs & medicalClass II

    W L Gore & Associates, Inc. recalls REF/Catalogue Number GSX0030H, GORE CARDIOFORM SEPTAL OCCLUDER 30 mm Diameter, Catheter Working Length 75 cm, Delivery Profile 10 Fr,…

    Septal Occluders contain the incorrect expiration date. Product is labeled with a 3 yr. expiration date instead of 2 yrs.

    FDA · DeviceW L Gore & Associates, Inc.US: AL,AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD,…Official notice →
All 10 W L Gore & Associates recalls →

Other drugs & medical recalls in the United States

  1. 2024-03-06
  2. US
    Drugs & medicalClass II

    Golden State Medical Supply Inc. recalls Fluticasone Propionate Nasal Spray USP, 50 mcg per spray, 120 metered sprays, 16 g net fill, Rx only, Manufactured by: Apotek…

    CGMP Deviations:Suspected potential presence of Burkholderia cepacia complex

    FDA · DrugGolden State Medical Supply Inc.Nationwide in the USAOfficial notice →
  3. 2024-02-28
  4. US
    Drugs & medicalClass II

    PTW-FREIBURG recalls Software VERIQA (S070031), version 2.0 and 2.1- Software package for display, evaluation and digital processing of medical image data sets…

    Software: If the user excludes voxels from the Gamma calculation that are below a dose threshold (Suppress gamma calculation), the Gamma Passing Rate (GPR) calculated for individual ROIs is not correct. The GPR…

    FDA · DevicePTW-FREIBURGGAOfficial notice →
  5. US
    Drugs & medicalClass II

    MEDLINE INDUSTRIES, LP - Northfield recalls These wipes are saturated with 0.65% bleach solution. These wipes are antimicrobial, disinfectant with sporicidal, virucidal,…

    There is stability concern for products dated with a 24-month shelf life. Product is safe and effective up to 12 months shelf life.

    FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldUS and territories, S. Korea, Panama, and DubaiOfficial notice →
  6. US
    Drugs & medicalClass II

    PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Azurion 7 with a Certeray generator-To perform image guidance in diagnostic, interventional, and minimally invasive…

    Generator may fail due to a potential short circuit in the Printed Circuit Board Assembly (PCBA) in PoInt EVR (Power Invertor). If a short circuit occurs, it will cause the fuses to trip and the system will become…

    FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Worldwide distribution - US Nationwide distribution in the states of NY, PA,…Official notice →
  7. US
    Drugs & medicalClass II

    Abiomed, Inc. recalls Automated Impella Controller (AIC), Product No. 0042-0000-US, 0042-0010-US, 0042-0040-US. Used with Impella Ventricular Support Systems…

    Retrospective reporting for the release of Technical Bulletin IMP-2643 AIC Version 8.5 Software Update Available. The update resolved an issue in version 8.4 where a pump was not recognized by the AIC.

    FDA · DeviceAbiomed, Inc.US Nationwide distribution.Official notice →
  8. US
    Drugs & medicalClass II

    MRP, LLC dba Aquabiliti recalls AQUABILITI AQUASTAT 0.9% Sodium Chloride Injection syringe, USP, 5mL in 12mL, REF 2T0805; and 3mL in 12mL, REF 2T0807, intermediate…

    The saline flush syringes may fail to meet USP criteria for stated sodium chloride content after 12 months.

    FDA · DeviceMRP, LLC dba AquabilitiUS Nationwide distribution in the states CO, FL, NC, NY, OH, PA, TN, TX, and VA.Official notice →
See all drugs & medical recalls →