Drugs & medicalClass II

W L Gore & Associates, Inc. recalls GORE CARDIOFORM ASD Occluder. cardiovascular implant.

Due to manufacturing records (Release Test Results) indicating "Failed".

Read the official notice →
Date
2022-11-02
Country
United States
Brand
W L Gore & Associates
Product
GORE CARDIOFORM ASD Occluder. cardiovascular implant.
Classification
Class II
Units
Worldwide distribution - U.S. Nationwide distribution including in the states…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More W L Gore & Associates recalls

  1. 2024-09-11
  2. US
    Drugs & medicalClass II

    W L Gore & Associates, Inc. recalls Heparin, Gore Viabahn VBX, Balloon Expandable Endoprosthesis, for the following Reference numbers: United States: BXB051501A…

    Due to packaging defects, sterility assurance and heparin activity may be compromised.

    FDA · DeviceW L Gore & Associates, Inc.Worldwide - US Nationwide distribution including in the states of AK, AL, AR,…Official notice →
  3. 2024-08-14
  4. US
    Drugs & medicalClass II

    W.L. Gore & Associates, Inc. recalls GORE SEAMGUARD Bioabsorbable Staple Line Reinforcement Configured for Circular Staplers, 25mm; Catalog number: 1BSGC25;

    Devices were labeled with an expiration date of four years and ten days rather than the validated two years.

    FDA · DeviceW.L. Gore & Associates, Inc.US Nationwide distribution in the states of CA, GA, IL, NC, OH, OK, WA.Official notice →
  5. 2024-02-28
  6. US
    Drugs & medicalClass II

    W.L. Gore & Associates, Inc. recalls GORE VIABIL Short Wire Biliary Endoprosthesis, Product labeled as (1) Catalog Number VSWVH1008, 10 mm x 8 cm; with holes; and (2)…

    Devices without transmural drainage holes may be mislabeled as devices with transmural drainage holes, or vice versa. Potential health impact includes extending procedure time, unplanned device removal, secondary…

    FDA · DeviceW.L. Gore & Associates, Inc.US Nationwide distribution in the state of Georgia.Official notice →
  7. 2023-02-08
  8. US
    Drugs & medicalClass II

    W L Gore & Associates, Inc. recalls GORE CARDIOFORM Septal Occluder, REF: GSX0030A

    Septal Occluders released with incomplete quality test documentation that assesses the release of the device from the delivery system during the index procedure. A potential failure of this quality test could…

    FDA · DeviceW L Gore & Associates, Inc.US Nationwide distribution in the states of NY, CA, FL, CT, TX, AR, IN, MA,…Official notice →
  9. 2021-11-03
  10. US
    Drugs & medicalClass II

    W L Gore & Associates, Inc. recalls REF/Catalogue Number GSX0030H, GORE CARDIOFORM SEPTAL OCCLUDER 30 mm Diameter, Catheter Working Length 75 cm, Delivery Profile 10 Fr,…

    Septal Occluders contain the incorrect expiration date. Product is labeled with a 3 yr. expiration date instead of 2 yrs.

    FDA · DeviceW L Gore & Associates, Inc.US: AL,AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD,…Official notice →
  11. 2021-10-20
  12. US
    Drugs & medicalClass II

    W L Gore & Associates, Inc. recalls Heparin, GORE VIABAHN VBX, BALLOON EXPANDABLE ENDOPROTHESIS, REF BXAL085902E, 8 MM X 59 MM 8 Fr, LARGE BALLOON EXPANDLBE, Nominal…

    Incorrect labeling, specifically, expandable endoprosthesis labeled as 8 mm x 59mm x 135 cm, however may contain a 9mm x 29 mm x 135 cm device.

    FDA · DeviceW L Gore & Associates, Inc.International distribution to the countries of Italy & Lebanon.Official notice →
All 10 W L Gore & Associates recalls →

Other drugs & medical recalls in the United States

  1. 2022-11-09
  2. US
    Drugs & medicalClass II

    SUN PHARMACEUTICAL INDUSTRIES INC recalls Buprenorphine and Naloxone Sublingual Tablets 8 mg/2 mg, 30-count bottles, Rx only, Distributed by: Sun Pharmaceutical…

    Presence of Foreign Substance

    FDA · DrugSUN PHARMACEUTICAL INDUSTRIES INCNationwide with the United StatesOfficial notice →
  3. 2022-11-02
  4. US
    Drugs & medicalClass II

    Wright Medical Technology, Inc. recalls Stryker INBONE Tibial Tray, Left Plasma Spray Ti6Al4V, CPti, Use w/ 16mm Stem Base, REF 200252903, Size 3

    The tibial tray lock detail is oversized (larger than specification).

    FDA · DeviceWright Medical Technology, Inc.US Nationwide distribution in the state of New York.Official notice →
  5. US
    Drugs & medicalClass II

    Merit Medical Systems, Inc. recalls CorVocet Biopsy System/Catalog Code: CORA1610/B, CORA1610S/B, CORA1615/B, CORA1810/A, CORA1810/B, CORA1810S/B, CORA1815/A,…

    Biopsy Systems have components that are misaligned due to an issue with the manufacturing assembly equipment, which may lead to the needle ejecting when triggered outside the body, which may result in user or patient…

    FDA · DeviceMerit Medical Systems, Inc.Worldwide distribution - US National distribution including in the states of…Official notice →
  6. US
    Drugs & medicalClass III

    LumiraDx recalls LumiraDx SARS-CoV-2 Antigen (Ag) Liquid Quality Control Kit Catalog #: L016080109002 (EUA) The LumiraDx SARS-CoV-2 Quality Controls are…

    Assigned vial lot numbers for the incorrect type of product. SARS-CoV-2 & Flu A/B lot numbers were incorrectly assigned to the positive and negative vial lot numbers for SARS-CoV-2 Ag LQCs within batch 2002204017,…

    FDA · DeviceLumiraDxUS Nationwide distribution.Official notice →
  7. US
    Drugs & medicalClass II

    Luminex Corporation recalls ARIES SARS-CoV-2 Assay, REF: 50-10047

    A part defect with the wash 1 syringe of an SARS-CoV-2 assay could lead to incomplete extraction thereby producing an invalid or false negative assay result.

    FDA · DeviceLuminex CorporationDistribution US nationwide, Indonesia, and Hong Kong.Official notice →
  8. US
    Drugs & medicalClass II

    Intuitive Surgical, Inc. recalls da Vinci X (IS4200) and Xi (IS4000) systems, which consists of finished devices Surgeon Console (SSC), Patient Cart (PSCART), Vision…

    Inadvertent energy delivery from surgical system instrument if 1) Force bipolar and bipolar instruments installed on system 2) Force bipolar connected to Force Triad generator, bipolar not connected to generator 3)…

    FDA · DeviceIntuitive Surgical, Inc.US: MS, PA, WI, KY, TN, MI, FL, GA, NC, NV, LA, OH, IL, MO, IN, CA, AK, MD,…Official notice →