PTW-FREIBURG recalls Software VERIQA (S070031), version 2.0 and 2.1- Software package for display, evaluation and digital processing of medical image data sets…
Software: If the user excludes voxels from the Gamma calculation that are below a dose threshold (Suppress gamma calculation), the Gamma Passing Rate (GPR) calculated for individual ROIs is not correct. The GPR…
Read the official notice →- Date
- 2024-02-28
- Country
- United States
- Brand
- Ptw-Freiburg
- Product
- Software VERIQA (S070031), version 2.0 and 2.1- Software package for display, evaluation and digital processing of medical image data sets…
- Classification
- Class II
- Units
- GA
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Ptw-Freiburg recalls
- 2026-09-23
US
PTW FREIBURG recalls Brand Name: VERIQA; Product Name: RT MonteCarlo 3D v3; Model/Catalog Number: S070031.520; Software Version: 3.1.18;
When a bolus is included in a treatment plan and sent for secondary dose calculation, the software accounts for it but does not display this clearly, which may lead users to incorrectly omit the bolus during treatment…
FDA · DevicePTW FREIBURGWorldwide - US Nationwide distribution in the states of GA, NY and the…Official notice →- 2023-09-20
US
PTW-FREIBURG recalls BeamAdjust software Version 2.3.3 Ref: S080032 The following devices are affected when a relative calibration is performed using ArrayCal…
Generated calibration file will not be correct in case of a relative calibration of an OCTAVIUS Detector 1000 or OCTAVIUS Detector 1600, if a Tiff file or a DICOM data set is used as reference matrix. Therefore, the…
Other drugs & medical recalls in the United States
- 2024-03-06
US
US
SUN PHARMACEUTICAL INDUSTRIES INC recalls Mesalamine Extended-Release Capsules, USP 500mg, Rx Only, 120 Capsules per bottle, Manufactured by: Sun Pharmaceutical…
Failed Dissolution Specifications: Out of specification for dissolution.
US
Golden State Medical Supply Inc. recalls Fluticasone Propionate Nasal Spray USP, 50 mcg per spray, 120 metered sprays, 16 g net fill, Rx only, Manufactured by: Apotek…
CGMP Deviations:Suspected potential presence of Burkholderia cepacia complex
- 2024-02-28
US
MEDLINE INDUSTRIES, LP - Northfield recalls These wipes are saturated with 0.65% bleach solution. These wipes are antimicrobial, disinfectant with sporicidal, virucidal,…
There is stability concern for products dated with a 24-month shelf life. Product is safe and effective up to 12 months shelf life.
FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldUS and territories, S. Korea, Panama, and DubaiOfficial notice →US
W.L. Gore & Associates, Inc. recalls GORE VIABIL Short Wire Biliary Endoprosthesis, Product labeled as (1) Catalog Number VSWVH1008, 10 mm x 8 cm; with holes; and (2)…
Devices without transmural drainage holes may be mislabeled as devices with transmural drainage holes, or vice versa. Potential health impact includes extending procedure time, unplanned device removal, secondary…
FDA · DeviceW.L. Gore & Associates, Inc.US Nationwide distribution in the state of Georgia.Official notice →US
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Azurion 7 with a Certeray generator-To perform image guidance in diagnostic, interventional, and minimally invasive…
Generator may fail due to a potential short circuit in the Printed Circuit Board Assembly (PCBA) in PoInt EVR (Power Invertor). If a short circuit occurs, it will cause the fuses to trip and the system will become…
FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Worldwide distribution - US Nationwide distribution in the states of NY, PA,…Official notice →