WOM World Of Medicine recalls

3 official recalls and safety alerts involving WOM World Of Medicine since 2021-10, from FDA · Device. Newest first. Check the official notice for exact models and batch codes.

This list only includes recalls Recall Atlas has collected since 2021-10. A product not listed here may still be affected; check with WOM World Of Medicine or the relevant agency.

  1. 2024-07-10
  2. US
    Drugs & medicalClass II

    WOM World Of Medicine AG recalls WOM Tube Set for lrrlgatlon, single-use- Intended to transport irrigation fluid to the patient in endoscopic surgeries. Distr. by…

    Incorrect transponder data was written on the RFID tags triggers the error message E03, Tube Set lnvalid, may result in a delay of surgery

    FDA · DeviceWOM World Of Medicine AGUS Nationwide distribution.Official notice →
  3. 2023-10-04
  4. US
    Drugs & medicalClass II

    WOM World Of Medicine AG recalls Aquilex Fluid Control System-Intended to provide fluid distension of the uterus during diagnostic and operative hysteroscopies and to…

    The display of inflow volume can reach its limit of 30,000 ml during long procdeure and the inflow volume display will freeze at the maximum value while the deficit will start counting backwards until 0 ml is reached…

    FDA · DeviceWOM World Of Medicine AGNationwide Foreign: Canada, Australia, New Zealand, Belgium, France, Germany,…Official notice →
  5. 2021-10-06
  6. US
    Drugs & medicalClass II

    WOM World of Medicine AG recalls Covidien HysteroLux Fluid Management System Control Unit, REF: 72205000

    When the display of inflow volume to the uterus reaches 32450 ml, the fluid management system control unit display will freeze because the internal software calculation threshold is reached. The outflow measurement…

    FDA · DeviceWOM World of Medicine AGUS Nationwide distribution in the state of CT.Official notice →