Drugs & medicalClass II

WOM World of Medicine AG recalls Covidien HysteroLux Fluid Management System Control Unit, REF: 72205000

When the display of inflow volume to the uterus reaches 32450 ml, the fluid management system control unit display will freeze because the internal software calculation threshold is reached. The outflow measurement…

Read the official notice →
Date
2021-10-06
Country
United States
Brand
WOM World Of Medicine
Product
Covidien HysteroLux Fluid Management System Control Unit, REF: 72205000
Classification
Class II
Units
US Nationwide distribution in the state of CT.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More WOM World Of Medicine recalls

  1. 2024-07-10
  2. US
    Drugs & medicalClass II

    WOM World Of Medicine AG recalls WOM Tube Set for lrrlgatlon, single-use- Intended to transport irrigation fluid to the patient in endoscopic surgeries. Distr. by…

    Incorrect transponder data was written on the RFID tags triggers the error message E03, Tube Set lnvalid, may result in a delay of surgery

    FDA · DeviceWOM World Of Medicine AGUS Nationwide distribution.Official notice →
  3. 2023-10-04
  4. US
    Drugs & medicalClass II

    WOM World Of Medicine AG recalls Aquilex Fluid Control System-Intended to provide fluid distension of the uterus during diagnostic and operative hysteroscopies and to…

    The display of inflow volume can reach its limit of 30,000 ml during long procdeure and the inflow volume display will freeze at the maximum value while the deficit will start counting backwards until 0 ml is reached…

    FDA · DeviceWOM World Of Medicine AGNationwide Foreign: Canada, Australia, New Zealand, Belgium, France, Germany,…Official notice →
All 3 WOM World Of Medicine recalls →

Other drugs & medical recalls in the United States

  1. 2021-10-06
  2. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls BF-XP60 OES Bronchofiberscope, Model No. BF-XP60

    The firm is issuing validated, new reprocessing instructions according to the FDA 2015 guidance, "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling."

    FDA · DeviceOlympus Corporation of the AmericasDomestic distribution nationwide. Product also distributed globally.Official notice →
  3. US
    Drugs & medicalClass II

    Elekta Inc recalls Elekta Monaco - Product Usage: used to make treatment plans for patients with prescriptions for external beam radiation therapy.

    Contour changes can be saved on an unintended image set. In addition, these contour edits do not cause the dose to be frozen on the plans associated with that image set.

    FDA · DeviceElekta IncWorldwide distribution - US Nationwide distribution in the states of PA, MI,…Official notice →
  4. US
    Drugs & medicalClass II

    Lymol Medical Corporation recalls LYMOL Medical Elite X Class Rigid Bronchoscope System, Part No. BD2410, BB2701, BB2702, BB2401, BB2402, BA2700, BT2000, BT2101, BT2103,…

    Reprocessing instructions were revised November 15, 2020, to reflect new processes and equipment used at consignee user facilities. New instructions were not sent to all consignees of the devices at the time of the…

    FDA · DeviceLymol Medical CorporationUS Nationwide distribution.Official notice →
  5. US
    Drugs & medicalClass II

    B. Braun Medical, Inc. recalls PUMP SET, 15 DR, 3 CARESITE, 127 in.single-use for use with the Outlook Safety Infusion System and/or Horizon NXT Pump for infusion of…

    Leakage of the pump set within the cassette portion of the pump sets, potential for delays in administration of medications, under-delivery or incomplete dosing and bloodstream infections.

    FDA · DeviceB. Braun Medical, Inc.Worldwide distribution - US Nationwide distribution and the country of Canada.Official notice →
  6. US
    Drugs & medicalClass II

    Vascutek, Ltd. recalls The Instructions for Use (IFU) for: Gelsoft Plus Bifurcate, Gelsoft Plus Ax-Fem, GelsoftPlus ERS Ax-Fem, Gelweave Straight, Gelweave…

    The devices contained the IFU for distribution to ROW (rest of world) consignees rather than U.S. consignees.

    FDA · DeviceVascutek, Ltd.Distribution was made to PR and Guam.Official notice →
  7. US
    Drugs & medicalClass I

    Accelerated Care Plus Corporation recalls Omnisound Gel, 250 ml bottle and 5 L container; and EcoGel 200, 250 ml bottle; Manufactured by Eco-Med. For use in ultrasound…

    Ultrasound gels and lotions manufactured by Eco-Med Pharmaceutical, Inc. may contain bacterial contamination. On August 18, 2021, FDA published a Letter to Health Care Providers to further address this issue:…

    FDA · DeviceAccelerated Care Plus CorporationUS nationwide distribution.Official notice →
See all drugs & medical recalls →