Remote Diagnostic Technologies recalls
10 official recalls and safety alerts involving Remote Diagnostic Technologies since 2022-04, from FDA · Device. Newest first. Check the official notice for exact models and batch codes.
This list only includes recalls Recall Atlas has collected since 2021-10. A product not listed here may still be affected; check with Remote Diagnostic Technologies or the relevant agency.
- 2026-01-14
US
Remote Diagnostic Technologies Ltd. recalls Philips Tempus Pro Patient Monitor, REF: 00-1004, 00-1007, 00-1004-R, 00-1007-R, 00-1024-R, 00-1026-R. USB 2-Channel IBP…
Patient monitor measures Intracranial Pressure (ICP) and Bladder Pressure (BDR) but ICP and BDR measurements have not been tested or validated for accuracy. If healthcare providers base clinical judgment on unvalidated…
FDA · DeviceRemote Diagnostic Technologies Ltd.Worldwide - US Nationwide distribution including in the states of AZ, VA, FL,…Official notice →- 2025-08-06
US
Remote Diagnostic Technologies Ltd. recalls Inseego USB8 4G Dongle Kit, Part Number: 01-2298, which contains USB Flash Drive, Part Number: 43-2011. The kit is an…
Kit USB flash drive contains outdated software, which could result in reverting patient monitor to outdated software and reintroducing the video laryngoscope issue (related to an August 2023 recall), which could lead…
FDA · DeviceRemote Diagnostic Technologies Ltd.Worldwide - US Nationwide distribution in the states of SC, KY, NV, IA, MI and…Official notice →- 2025-05-14
US
Remote Diagnostic Technologies Ltd. recalls Philips Tempus Pro Patient Monitor, REF: 00-1004-R, 00-1007-R, 00-1024-R, 00-1026-R, 00-1004, 00-1007
Patient Monitor software update to address issues:1)May freeze at startup with MPM Application error and Attention Restart Required messages,2)Monitor may freeze and reboot at start of 12-lead ECG or after, 3)May…
FDA · DeviceRemote Diagnostic Technologies Ltd.Worldwide - US Nationwide distribution including in the states of AZ, VA, FL,…Official notice →- 2025-04-30
US
Remote Diagnostic Technologies Ltd. recalls Tempus Pro Patient Monitor, REFs: 00-1004-R (Tempus Pro non-printer, IBP, without headset dock), 00-1007-R and 00-1007…
software issue that causes an error screen on a portable vital signs monitor when attaching video Laryngoscope.
FDA · DeviceRemote Diagnostic Technologies Ltd.Worldwide - US Nationwide distribution including in the states of AL, AZ, CA,…Official notice →- 2024-12-04
US
Remote Diagnostic Technologies Ltd. recalls Tempus LS-Manual Defibrillator Model 00-3020 The Tempus LS-Manual is a defibrillator that can deliver a shock in manual or…
Due to incorrect labels applied to devices during servicing. Labels contain the incorrect product name and CAT number.
FDA · DeviceRemote Diagnostic Technologies Ltd.U.S. Nationwide distribution in the states of AL, CA, FL, MO, NY, OH, TX and VA.Official notice →- 2024-06-12
US
Remote Diagnostic Technologies Ltd. recalls Philips Tempus LS-Manual Defibrillator, Part Number: 00-3020 when using Tempus LS-Manual User/Operator Manual
Simultaneous ECG measurements with multiple devices, including a manual defibrillator, on the same patient may cause electrical noise or interference, which may lead to distortions of the ECG signal; distortions may…
FDA · DeviceRemote Diagnostic Technologies Ltd.US Nationwide distribution including in the states of VA, DC, MA, WY, AZ, FL,…Official notice →- 2023-09-27
US
Remote Diagnostic Technologies Ltd. recalls Philips Tempus Pro Patient Monitor, REF: 00-1004-R, 00-1007-R, 00-1024-R, 00-1026-R, 00-1004, and 00-1007, when Used With…
Full screen error message may occur on patient monitors with affected hardware version, either during or after laryngoscopes are unplugged from monitors, informing users a 60-100 second shutdown/restart are required.…
FDA · DeviceRemote Diagnostic Technologies Ltd.US:AZ, AL, CA, CO, DC, DE, FL , GA, HI, IA, ID, IL, IN , KS, KY, MA , MD , ME,…Official notice →- 2023-06-21
US
Remote Diagnostic Technologies Ltd. recalls Philips Tempus LS-Manual Defibrillator, Part Number: 00-3020
Defibrillator/pacemaker module (DPM) may encounter a problem causing communication from the Mainboard to the DPM to fail, leading to no or ineffective pacing with displayed error message "HW failure DPM".
FDA · DeviceRemote Diagnostic Technologies Ltd.US Nationwide distribution in the states of UT, CO, CA, NE, IL, TX, WV, KS, GA.Official notice →- 2023-04-12
US
Remote Diagnostic Technologies Ltd. recalls Tempus Pro Patient Monitor is a portable vital signs monitor intended to be used by clinicians and medically qualified…
During internal testing, Regulatory Compliance issues were identified regarding the Monitor and Power Supplies for fluid ingress and basic safety issues.
FDA · DeviceRemote Diagnostic Technologies Ltd.Worldwide Distribution: U.S.(nationwide): AL, AZ, CA, CO, DC, DE, FL, GA, HI,…Official notice →- 2022-04-13
US
Remote Diagnostic Technologies Ltd. recalls Tempus LS, Type DGE1, CAT 00-3010, REF 3.940590, Software version 1.3.5, UDI: (01)07613365001693
Distribution of Defibrillators that are not approved or cleared for distribution US Market.
FDA · DeviceRemote Diagnostic Technologies Ltd.US Nationwide distribution in the states of IL and KS.Official notice →