Drugs & medicalClass II

Remote Diagnostic Technologies Ltd. recalls Philips Tempus Pro Patient Monitor, REF: 00-1004-R, 00-1007-R, 00-1024-R, 00-1026-R, 00-1004, and 00-1007, when Used With…

Full screen error message may occur on patient monitors with affected hardware version, either during or after laryngoscopes are unplugged from monitors, informing users a 60-100 second shutdown/restart are required.…

Read the official notice →
Date
2023-09-27
Country
United States
Brand
Remote Diagnostic Technologies
Product
Philips Tempus Pro Patient Monitor, REF: 00-1004-R, 00-1007-R, 00-1024-R, 00-1026-R, 00-1004, and 00-1007, when Used With Tempus Pro Video…
Classification
Class II
Units
US:AZ, AL, CA, CO, DC, DE, FL , GA, HI, IA, ID, IL, IN , KS, KY, MA , MD , ME,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Remote Diagnostic Technologies recalls

  1. 2025-04-30
  2. US
    Drugs & medicalClass II

    Remote Diagnostic Technologies Ltd. recalls Tempus Pro Patient Monitor, REFs: 00-1004-R (Tempus Pro non-printer, IBP, without headset dock), 00-1007-R and 00-1007…

    software issue that causes an error screen on a portable vital signs monitor when attaching video Laryngoscope.

    FDA · DeviceRemote Diagnostic Technologies Ltd.Worldwide - US Nationwide distribution including in the states of AL, AZ, CA,…Official notice →
  3. 2024-12-04
  4. US
    Drugs & medicalClass III

    Remote Diagnostic Technologies Ltd. recalls Tempus LS-Manual Defibrillator Model 00-3020 The Tempus LS-Manual is a defibrillator that can deliver a shock in manual or…

    Due to incorrect labels applied to devices during servicing. Labels contain the incorrect product name and CAT number.

    FDA · DeviceRemote Diagnostic Technologies Ltd.U.S. Nationwide distribution in the states of AL, CA, FL, MO, NY, OH, TX and VA.Official notice →
  5. 2024-06-12
  6. US
    Drugs & medicalClass II

    Remote Diagnostic Technologies Ltd. recalls Philips Tempus LS-Manual Defibrillator, Part Number: 00-3020 when using Tempus LS-Manual User/Operator Manual

    Simultaneous ECG measurements with multiple devices, including a manual defibrillator, on the same patient may cause electrical noise or interference, which may lead to distortions of the ECG signal; distortions may…

    FDA · DeviceRemote Diagnostic Technologies Ltd.US Nationwide distribution including in the states of VA, DC, MA, WY, AZ, FL,…Official notice →
  7. 2023-06-21
  8. US
    Drugs & medicalClass II

    Remote Diagnostic Technologies Ltd. recalls Philips Tempus LS-Manual Defibrillator, Part Number: 00-3020

    Defibrillator/pacemaker module (DPM) may encounter a problem causing communication from the Mainboard to the DPM to fail, leading to no or ineffective pacing with displayed error message "HW failure DPM".

    FDA · DeviceRemote Diagnostic Technologies Ltd.US Nationwide distribution in the states of UT, CO, CA, NE, IL, TX, WV, KS, GA.Official notice →
  9. 2023-04-12
  10. US
    Drugs & medicalClass II

    Remote Diagnostic Technologies Ltd. recalls Tempus Pro Patient Monitor is a portable vital signs monitor intended to be used by clinicians and medically qualified…

    During internal testing, Regulatory Compliance issues were identified regarding the Monitor and Power Supplies for fluid ingress and basic safety issues.

    FDA · DeviceRemote Diagnostic Technologies Ltd.Worldwide Distribution: U.S.(nationwide): AL, AZ, CA, CO, DC, DE, FL, GA, HI,…Official notice →
  11. 2022-04-13
  12. US
    Drugs & medicalClass II

    Remote Diagnostic Technologies Ltd. recalls Tempus LS, Type DGE1, CAT 00-3010, REF 3.940590, Software version 1.3.5, UDI: (01)07613365001693

    Distribution of Defibrillators that are not approved or cleared for distribution US Market.

    FDA · DeviceRemote Diagnostic Technologies Ltd.US Nationwide distribution in the states of IL and KS.Official notice →
All 10 Remote Diagnostic Technologies recalls →

Other drugs & medical recalls in the United States

  1. 2023-10-04
  2. US
    Drugs & medicalClass II

    Sofie Co dba Sofie recalls Florbetaben F-18 (Neuraceq¿) Injection Solution 1.4 to 135 mCi/mL, 50 mL in 1 multi-dose glass vial, Diagnostic-For Intravenous Use Only,…

    Lack of Assurance of Sterility: out-of-specification test results observed for Filter Integrity Test (FIT).

    FDA · DrugSofie Co dba SofieProduct was released to one facility in VA.Official notice →
  3. 2023-09-27
  4. US
    Drugs & medicalClass II

    B. Braun Medical, Inc. recalls Perifix¿ / Epidural anesthesia set, medicated

    Product may be mislabeled with an incorrect lid stock label.

    FDA · DeviceB. Braun Medical, Inc.US Distributions to states of - AR, AZ, CA, FL, GA, IL, MA, MO, NC, NJ, PA,…Official notice →
  5. US
    Drugs & medicalClass II

    O&M HALYARD, INC. recalls FLUIDSHIELD 3 N95 Particulate Filter Respirator and Surgical Mask with SO SOFT* Lining, Orange, size small, Product Code 46827

    One respirator sample failed a filtration efficiency test. This respirator may not provide adequate protection to the user.

    FDA · DeviceO&M HALYARD, INC.US Distribution to States of: FL, GA, IA, IL, MO, MS, NE, OK, TN, TX, WI and…Official notice →
  6. US
    Drugs & medicalClass II

    Ortho-Clinical Diagnostics, INc. recalls VITROS XT Chemistry Products ALKP Slides

    Samples containing high levels of ALKP (or an interfering substance) could deplete the available substrate in the reagent layer of a VITROS ALKP Slide and not post the substrate deleption (DP) code, which may result in…

    FDA · DeviceOrtho-Clinical Diagnostics, INc.Worldwide Distribution: US (nationwide) and OUS (foreign) distribution to…Official notice →
  7. US
    Drugs & medicalClass II

    Avanos Medical, Inc. recalls COOLIEF RF Generator is an 80-watt Radio-Frequency (RF) Generator with four independent RF channels, allowing all RF channels to be…

    Avanos has received an increase in complaints from customers for COOLIEF RF Generators producing F100 and F101 faults.

    FDA · DeviceAvanos Medical, Inc.US Distribution to states of: AZ, CA, GA, IA, IN, KS, LA, MN, NC, NM, NY, OH,…Official notice →
  8. US
    Drugs & medicalClass II

    Smith & Nephew Inc recalls smith&nephew LEGION OXINUM POSTERIOR STABILIZED NONPOROUS FERMOAL COMPONENT, SIZE 5N, LEFT, NARROW, REF 71421265; knee joint

    The firm conducting a recall due to a complaint received indicating that a package contained a LEGION NARROW PS OXIN SZ 4N LT implant instead of a LEGION NARROW PSvOXIN SZ 5N LT as described on the product label.

    FDA · DeviceSmith & Nephew IncOH, CT, TX, NC, CA, NE, NVOfficial notice →
See all drugs & medical recalls →