MicroPort Orthopedics recalls
5 official recalls and safety alerts involving MicroPort Orthopedics since 2022-06, from FDA · Device. Newest first. Check the official notice for exact models and batch codes.
This list only includes recalls Recall Atlas has collected since 2021-10. A product not listed here may still be affected; check with MicroPort Orthopedics or the relevant agency.
- 2026-01-28
US
MicroPort Orthopedics Inc. recalls Brand Name: MicroPort Product Name: HIPTURN FEM HEAD TRIAL 36MM REUSABLE Model/Catalog Number: HTHT0036 Software Version: N/A Product…
Due to products not having FDA Premarket authorization to be distributed within the United States.
FDA · DeviceMicroPort Orthopedics Inc.The product was sent to distributors in multiple US states, including Texas,…Official notice →- 2024-12-25
US
MicroPort Orthopedics Inc. recalls EVOLUTION MP FEMORAL COMPONENT CS/CR POROUS SIZE 5 RIGHT PRIMARY, REF: EFSRP5PR; EVOLUTION MP FEMORAL COMPONENT CS/CR NON-POROUS SIZE…
Femoral implant porous size 5 right primary and femoral implant non-porous size 6 left primary have correct markings on devices, but may have incorrect packaging.
- 2024-07-10
US
MicroPort Orthopedics Inc. recalls EVOLUTION¿ MP Femoral CS/CR Non-Porous Non-porous MicroPort total knee replacement implants are for cemented use only.
One confirmed incident has been received that EFSRN4PR, lot number MP1982170, Size 4 Evolution MP Non-Porous Femoral Right was opened during surgery and contained part number EFSRN3PL, lot number MP1982097, Size 3…
FDA · DeviceMicroPort Orthopedics Inc.International distribution in the country of China. There is no field…Official notice →- 2023-02-22
US
MicroPort Orthopedics Inc. recalls EVOLUTION MP Tibial Bases, Size 7 LEFT, REF ETPKN7SL, knee tibial base
One confirmed incident has been received that ETPKN2PL lot 1916559, size 2 Evolution MP Tibial Base, was opened during surgery and contained ETPKN7SL lot 1916715, size 7 Evolution MP Tibial Base, in the packaging.
FDA · DeviceMicroPort Orthopedics Inc.International Distribution to countries of: Germany, Italy, France, Belgium,…Official notice →- 2022-06-08
US
MicroPort Orthopedics Inc. recalls PROFEMUR PLASMA Z CLASSIC STEM SIZE 7 STANDARD SHORT NECK
One manufacturing lot is missing the plasma-spray coating in the proximal region of the stem body.
FDA · DeviceMicroPort Orthopedics Inc.Worldwide distribution - US Nationwide distribution in the states of FL, IN,…Official notice →