Drugs & medicalClass II

MicroPort Orthopedics Inc. recalls EVOLUTION MP FEMORAL COMPONENT CS/CR POROUS SIZE 5 RIGHT PRIMARY, REF: EFSRP5PR; EVOLUTION MP FEMORAL COMPONENT CS/CR NON-POROUS SIZE…

Femoral implant porous size 5 right primary and femoral implant non-porous size 6 left primary have correct markings on devices, but may have incorrect packaging.

Read the official notice →
Date
2024-12-25
Country
United States
Brand
MicroPort Orthopedics
Product
EVOLUTION MP FEMORAL COMPONENT CS/CR POROUS SIZE 5 RIGHT PRIMARY, REF: EFSRP5PR; EVOLUTION MP FEMORAL COMPONENT CS/CR NON-POROUS SIZE 6…
Classification
Class II
Units
OUS: CN, GR, ES, ZA, FR, JP, IT
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More MicroPort Orthopedics recalls

  1. 2026-01-28
  2. US
    Drugs & medicalClass II

    MicroPort Orthopedics Inc. recalls Brand Name: MicroPort Product Name: HIPTURN FEM HEAD TRIAL 36MM REUSABLE Model/Catalog Number: HTHT0036 Software Version: N/A Product…

    Due to products not having FDA Premarket authorization to be distributed within the United States.

    FDA · DeviceMicroPort Orthopedics Inc.The product was sent to distributors in multiple US states, including Texas,…Official notice →
  3. 2024-07-10
  4. US
    Drugs & medicalClass II

    MicroPort Orthopedics Inc. recalls EVOLUTION¿ MP Femoral CS/CR Non-Porous Non-porous MicroPort total knee replacement implants are for cemented use only.

    One confirmed incident has been received that EFSRN4PR, lot number MP1982170, Size 4 Evolution MP Non-Porous Femoral Right was opened during surgery and contained part number EFSRN3PL, lot number MP1982097, Size 3…

    FDA · DeviceMicroPort Orthopedics Inc.International distribution in the country of China. There is no field…Official notice →
  5. 2023-02-22
  6. US
    Drugs & medicalClass II

    MicroPort Orthopedics Inc. recalls EVOLUTION MP Tibial Bases, Size 7 LEFT, REF ETPKN7SL, knee tibial base

    One confirmed incident has been received that ETPKN2PL lot 1916559, size 2 Evolution MP Tibial Base, was opened during surgery and contained ETPKN7SL lot 1916715, size 7 Evolution MP Tibial Base, in the packaging.

    FDA · DeviceMicroPort Orthopedics Inc.International Distribution to countries of: Germany, Italy, France, Belgium,…Official notice →
  7. 2022-06-08
  8. US
    Drugs & medicalClass II

    MicroPort Orthopedics Inc. recalls PROFEMUR PLASMA Z CLASSIC STEM SIZE 7 STANDARD SHORT NECK

    One manufacturing lot is missing the plasma-spray coating in the proximal region of the stem body.

    FDA · DeviceMicroPort Orthopedics Inc.Worldwide distribution - US Nationwide distribution in the states of FL, IN,…Official notice →
All 5 MicroPort Orthopedics recalls →

Other drugs & medical recalls in the United States

  1. 2024-12-25
  2. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls Duravent Silicone Ventilation Tube. Model Number: 240075.

    Potential that Duravent Silicone Ventilation Tubes may contain incorrect product, specifically, 1.32 mm Tiny T Tubes (model: 70240076) instead of the intended Duravent Silicone Ventilation Tubes.

    FDA · DeviceOlympus Corporation of the AmericasAL, AZ, CA, IL, KY, MD, MN, MO, NC, ND, NY, OK, PA, SC, TN, TX, WA, WI.…Official notice →
  3. US
    Drugs & medicalClass II

    GE OEC Medical Systems, Inc recalls OEC 3D REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The OEC 3D is a mobile fluoroscopic C-Arm imaging system used to…

    Due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability.

    FDA · DeviceGE OEC Medical Systems, IncU.S.: AK, AZ, CA, CO, CT, DC, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA,…Official notice →
  4. US
    Drugs & medicalClass II

    Augustine Temperature Management, LLC recalls Augustine Surgical Inc. HOT DOG PATIENT WARMING, Temperature Management Controller, Multiport, Models: 1) Model # WC77, 2)…

    Instructions for use update on HotDog Temperature Management Controllers Models WC71, WC77, and Temperature Management Controller MP Model 2083516 pertaining to Power-on self-test error code "EA POST".

    FDA · DeviceAugustine Temperature Management, LLCUS and CanadaOfficial notice →
  5. US
    Drugs & medicalClass II

    MEDLINE INDUSTRIES, LP - Northfield recalls HUDSON RCI mBrace, ET Tube Holder with 4 Point Head Strap, REF DYNJMBRC4; tracheal tube fixation device

    Complaint defects have been noted regarding the adhesive elements of the device and the frame of the holder putting additional pressure on the patient's cheekbones and upper lip which may cause dermal injury depending…

    FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldUS and CanadaOfficial notice →
  6. US
    Drugs & medicalClass II

    Noah Medical recalls Galaxy System (GAL-001)

    Bronchoscopic system uninterruptable power may have defective component, may cause power supply short circuit/trip protective circuit breakers causing sudden, unrecoverable power loss, which could occur while system is…

    FDA · DeviceNoah MedicalUS: CA, MA, PA, NJ, ID, CT, MO, NC, MD, IL, TN, VA, OH, IN, WV, NYOfficial notice →
  7. US
    Drugs & medicalClass II

    CooperSurgical, Inc. recalls Brand Name: K-Systems Product Name: G85 Mini Incubator Model/Catalog Number: K22074 Software Version: N/A Product Description: An…

    The warming surfaces or incubator well temperatures of the listed products may exceed the desired set temperature. If temperatures exceed the desired set value, the device will alarm appropriately.

    FDA · DeviceCooperSurgical, Inc.States: NC, NJ, TX, VA, IL, NY, CA, PA Please note: The countries below,…Official notice →
See all drugs & medical recalls →