Micro-X recalls
8 official recalls and safety alerts involving Micro-X since 2022-10, from FDA · Device. Newest first. Check the official notice for exact models and batch codes.
This list only includes recalls Recall Atlas has collected since 2021-10. A product not listed here may still be affected; check with Micro-X or the relevant agency.
- 2026-04-29
US
Micro-X Ltd. recalls Rover Mobile X-ray System Model/Catalog Number: MXU-RV35 and MXU-RV71
X-Ray generator may malfunction resulting in the X-Ray being inoperable.
FDA · DeviceMicro-X Ltd.US Nationwide distribution in the states and territories of WA, NJ, CA, Puerto…Official notice →- 2025-04-30
US
Micro-X Ltd. recalls Brand Name: Rover Product Name: Mobile X-ray System Model/Catalog Number:MXU-RV35 Product Description: Rover Mobile X-rays System…
Potential for internal fastening mechanism within generator of mobile x-Ray to fail.
- 2025-04-23
US
Micro-X Ltd. recalls Brand Name: Rover Product Name: Mobile X-ray System Model/Catalog Number: MXU-RV35 Product Description: Rover Mobile X-rays System…
Potential for early life x-ray tube failure for mobile x-ray system due to excessive thermal load on the anode.
FDA · DeviceMicro-X Ltd.US Nationwide distribution in the states of CA, NJ, and Puerto Rico (US…Official notice →- 2024-12-18
US
Micro-X Ltd. recalls MICRO-X Rover Mobile X-ray System, # MXU-RV19
During internal testing it was found that if the system is Ready to expose and the hand switch is then dropped into the stowage unit such that it lands on the prep/expose switch, the prep switch can be activated for a…
US
Micro-X Ltd. recalls MICRO-X Rover Mobile X-ray System, # MXU-RV35; MICRO-X Rover Mobile X-ray System, # MXU-RV71
Mobile x-ray system can report and deliver a lower post-exposure mAs value than what set by the operator.
US
Micro-X Ltd. recalls MICRO-X Rover Mobile X-ray System, # MXU-RV35
Shots were terminated by the mAs integrator; however, the 4 ms exposure time did not allow enough tolerance to achieve diagnostic exposures due to variations in hardware and how different timers are started.
- 2024-12-04
US
Micro-X Ltd. recalls MICRO-X Rover Mobile X-ray System, # MXU-RV35
Mobile x-ray system receiving/storage(operation) requirements to be followed: Temp:14 to 131(50 to 86) degrees F, RH:10% to 86%(30% to 60%) allow condensation dry time, Atmospheric Pressure:50kPa to 106 kPa(70kPa to…
- 2022-10-05
US
Micro-X Ltd. recalls MICRO-X ROVER MOBILE X-RAY SYSTEM, MXU-RV19
Software calibration error with product equip with a Dose Area Product (DAP) meter. This results in the product displaying the incorrect DAP meter values which may impede radiation exposure management decisions
FDA · DeviceMicro-X Ltd.Worldwide distribution - U.S. Nationwide distribution in the states of CA, NV,…Official notice →