Micro-X Ltd. recalls MICRO-X Rover Mobile X-ray System, # MXU-RV35; MICRO-X Rover Mobile X-ray System, # MXU-RV71
Mobile x-ray system can report and deliver a lower post-exposure mAs value than what set by the operator.
Read the official notice →- Date
- 2024-12-18
- Country
- United States
- Brand
- Micro-X
- Product
- MICRO-X Rover Mobile X-ray System, # MXU-RV35; MICRO-X Rover Mobile X-ray System, # MXU-RV71
- Classification
- Class II
- Units
- US Nationwide Distribution: NJ, CA, PR
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Micro-X recalls
- 2025-04-30
US
Micro-X Ltd. recalls Brand Name: Rover Product Name: Mobile X-ray System Model/Catalog Number:MXU-RV35 Product Description: Rover Mobile X-rays System…
Potential for internal fastening mechanism within generator of mobile x-Ray to fail.
- 2025-04-23
US
Micro-X Ltd. recalls Brand Name: Rover Product Name: Mobile X-ray System Model/Catalog Number: MXU-RV35 Product Description: Rover Mobile X-rays System…
Potential for early life x-ray tube failure for mobile x-ray system due to excessive thermal load on the anode.
FDA · DeviceMicro-X Ltd.US Nationwide distribution in the states of CA, NJ, and Puerto Rico (US…Official notice →- 2024-12-18
US
Micro-X Ltd. recalls MICRO-X Rover Mobile X-ray System, # MXU-RV19
During internal testing it was found that if the system is Ready to expose and the hand switch is then dropped into the stowage unit such that it lands on the prep/expose switch, the prep switch can be activated for a…
US
Micro-X Ltd. recalls MICRO-X Rover Mobile X-ray System, # MXU-RV35
Shots were terminated by the mAs integrator; however, the 4 ms exposure time did not allow enough tolerance to achieve diagnostic exposures due to variations in hardware and how different timers are started.
- 2024-12-04
US
Micro-X Ltd. recalls MICRO-X Rover Mobile X-ray System, # MXU-RV35
Mobile x-ray system receiving/storage(operation) requirements to be followed: Temp:14 to 131(50 to 86) degrees F, RH:10% to 86%(30% to 60%) allow condensation dry time, Atmospheric Pressure:50kPa to 106 kPa(70kPa to…
- 2022-10-05
US
Micro-X Ltd. recalls MICRO-X ROVER MOBILE X-RAY SYSTEM, MXU-RV19
Software calibration error with product equip with a Dose Area Product (DAP) meter. This results in the product displaying the incorrect DAP meter values which may impede radiation exposure management decisions
FDA · DeviceMicro-X Ltd.Worldwide distribution - U.S. Nationwide distribution in the states of CA, NV,…Official notice →
Other drugs & medical recalls in the United States
- 2024-12-18
US
CareFusion 303, Inc. recalls BD Pyxis CII Safe ES, REF: 1116-00
Due to a software issue, automated dispensing cabinet devices may open the wrong cubie pocket or position, which can lead to: 1) Unintended and incorrect medication removal, or 2) Potential for a different medication…
FDA · DeviceCareFusion 303, Inc.Worldwide - US Nationwide distribution including in the states of OK, LA, FL,…Official notice →US
Hologic, Inc recalls Hologic Sizer Set- sterilization tray and 6 individual sizers, intended as a Surgical instrument guide Model/Catalog Number: F0101
IFU was not provided with the Sizer Set sterilization tray Revised Sterilization Instructions for Sizer Set
US
Micro-X Ltd. recalls MICRO-X Rover Mobile X-ray System, # MXU-RV19
During internal testing it was found that if the system is Ready to expose and the hand switch is then dropped into the stowage unit such that it lands on the prep/expose switch, the prep switch can be activated for a…
US
Micro-X Ltd. recalls MICRO-X Rover Mobile X-ray System, # MXU-RV35
Shots were terminated by the mAs integrator; however, the 4 ms exposure time did not allow enough tolerance to achieve diagnostic exposures due to variations in hardware and how different timers are started.
US
Microline Surgical, Inc. recalls Microscissor Tip (Disposable); Microline Catalog #: 3122.
A lot of ReNew Microscissor Tips exhibited a 2% rate of failure in terms of the heat shrink breaking off.
FDA · DeviceMicroline Surgical, Inc.Worldwide - US Nationwide in the state of OK and the country of Japan.Official notice →US
ICU Medical, Inc. recalls ChemoLock Close Vial Spike REF CL-80 The Chemolock Closed System Transfer Device prevents the transfer of environmental contaminants. The…
Due to finished goods incorrectly packaged and labeled.
FDA · DeviceICU Medical, Inc.Worldwide - U.S. Nationwide distribution in the state of IL and the country of…Official notice →