- 2024-03-27
US
Landauer recalls microStar Reader, Model Numbers: 10015-000 (marketed in EMEA as 100, 102, 110, and 201), 10016-000 (marketed in EMEA as 101), 10036-000,…
Permanent discontinuance of the microSTARii medical dosimetry system. Landauer has initiated this action to remove all reference to use of the microStar readers with the nanoDot or any otherbdosimeter for medical…
FDA · DeviceLandauerWorldwide distribution - United States Nationwide and the countries of…Official notice →- 2024-01-03
US
Landauer recalls Landauer microSTARii reader, Medical Dosimetry System, Model Numbers: 18000-000/18001-000/18007-000
MicroSTARii readers may be contributing to the nanodot dosimeter measurement inaccuracy in two possible ways. The first is related to the interaction of the reader s LED beam profile with the nanoDot dosimeter. The…
FDA · DeviceLandauerUS and Australia, Brazil, Canada, China, France, Germany, India, Japan,…Official notice →- 2023-09-20
US
Landauer recalls nanoDot D2DNS, Model Numbers: a) 18100-000; b) 18100-1MO; c) 18100-2MO; d) 18100-KIT; radiation monitoring dosimeter used with the…