Landauer recalls Landauer microSTARii reader, Medical Dosimetry System, Model Numbers: 18000-000/18001-000/18007-000
MicroSTARii readers may be contributing to the nanodot dosimeter measurement inaccuracy in two possible ways. The first is related to the interaction of the reader s LED beam profile with the nanoDot dosimeter. The…
Read the official notice →- Date
- 2024-01-03
- Country
- United States
- Brand
- Landauer
- Product
- Landauer microSTARii reader, Medical Dosimetry System, Model Numbers: 18000-000/18001-000/18007-000
- Classification
- Class II
- Units
- US and Australia, Brazil, Canada, China, France, Germany, India, Japan,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Landauer recalls
- 2024-03-27
US
Landauer recalls microStar Reader, Model Numbers: 10015-000 (marketed in EMEA as 100, 102, 110, and 201), 10016-000 (marketed in EMEA as 101), 10036-000,…
Permanent discontinuance of the microSTARii medical dosimetry system. Landauer has initiated this action to remove all reference to use of the microStar readers with the nanoDot or any otherbdosimeter for medical…
FDA · DeviceLandauerWorldwide distribution - United States Nationwide and the countries of…Official notice →- 2023-09-20
US
Landauer recalls nanoDot D2DNS, Model Numbers: a) 18100-000; b) 18100-1MO; c) 18100-2MO; d) 18100-KIT; radiation monitoring dosimeter used with the…
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the…
Other drugs & medical recalls in the United States
- 2024-01-03
US
Fresenius Medical Care Holdings, Inc. recalls Liberty Select Cycler, Model numbers RTLR180343 and RTLR180111
Affected Liberty Select cyclers may display an incorrect time stamp on the treatment data report sent to the clinic.
FDA · DeviceFresenius Medical Care Holdings, Inc.US Nationwide distribution in the states of AL, CA, FL, IL, IN, LA, MI, MO,…Official notice →US
Acumed LLC recalls RibLoc U Plus 90 Instrument PRIMARY GUIDE ASSY, LOW PROFILE REF RBL2320 The RBL2320, Low Profile Primary Guide is an instrument used with…
Due to potential breakage during use.
FDA · DeviceAcumed LLCWorldwide - U.S. Nationwide distribution including in the states of AL, AZ,…Official notice →US
Steris Corporation recalls Hookup Accessories used with MEDIVATORS ADVANTAGE PLUS and ADVANTAGE PLUS Pass Thru Automated Endoscope Reprocessors Model # Model…
Identification label on affected hookups could become illegible over time due to the label's ink coming off. This may potentially prevent users from being able to effectively identify the hookup and may ultimately…
FDA · DeviceSteris CorporationDomestic distribution nationwide. Foreign distribution to Australia Canada…Official notice →US
Medtronic Neuromodulation recalls Medtronic SynchroMed II, Model 8637-20, Programmable pump
Medtronic recently identified that if the SynchroMed II pump switches into telemetry mode due to electromagnetic interference (EMI) from an MRI scan, while the pump is sounding an alarm, the pump will not resume drug…
US
Abbott Laboratories recalls Alinity ci-series System Control Module, REF: 03R70-01
There are potential performance issues found in the Alinity ci-series System software versions 3.4.0 and lower, including: 1) SCC Reagent load error during daily maintenance; 2) Sample Laboratory Report error; 3)…
FDA · DeviceAbbott LaboratoriesUS: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA,…Official notice →US