Zimmer, Inc. recalls CPT Hip Joint Metal/Polymer Semi-Constrained Cemented Prosthesis, used in Hip Arthroplasty; Item Numbers 00811400000 00811400010…
Affected product has an increased risk of postoperative perisprosthetic femoral fracture (PFF). The IFU is being updated to reflect the risk of PFF.
Read the official notice →- Date
- 2024-08-07
- Country
- United States
- Brand
- Zimmer
- Product
- CPT Hip Joint Metal/Polymer Semi-Constrained Cemented Prosthesis, used in Hip Arthroplasty; Item Numbers 00811400000 00811400010…
- Classification
- Class II
- Units
- Worldwide - US Nationwide distribution.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Zimmer recalls
- 2025-02-05
US
Zimmer, Inc. recalls NexGen LCCK Legacy Articular Surface with Locking Screw - Striped Yellow/E,F - Height 10 mm, Item Number 00-5994-032-10
It was identified internally that two commingle events occurred where the metal support post within the polyethylene articular surface was assembled incorrectly for three units within the scope.
FDA · DeviceZimmer, Inc.Worldwide - US Nationwide distribution in the states of CA, FL, GA, KS, MO,…Official notice →- 2024-11-13
US
Zimmer, Inc. recalls Saphena Medical Venapax XL- Endoscopic Vessel Harvesting System for use in minimally invasive surgery allowing access for vessel…
A Gen 5 tail piece was assembled to a Gen 4 Finished Device, resulting in a 0.178" difference in the end of the endoscope slot to the end of the tail piece. This allows the scope to move further into the plastic sheath…
- 2024-09-18
US
Zimmer GmbH recalls Biolox Option Taper Sleeve, Type 1 Taper, -6mm Neck - Indicated for use in total hip replacement with cemented or non-cemented femoral and…
Mislabeled: Incorrect sleeve in the packaging. The outer packaging is labeled as a standard neck sleeve, however, the product within the box is a -6mm neck sleeve, and vice versa. The neck of the sleeve can be…
FDA · DeviceZimmer GmbHWorldwide distribution - US Nationwide and the countries of Japan, Netherlands.Official notice →- 2024-05-08
US
Zimmer, Inc. recalls ZPLP(Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Left, 6 Holes, 106 MM, Sterile- Indicated for temporary…
Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock,…
FDA · DeviceZimmer, Inc.Worldwide distribution - US Nationwide and the countries of Brazil, Canada,…Official notice →- 2024-05-01
US
Zimmer, Inc. recalls NexGen CR-Flex Femoral Component, Size E and Size E Minus (E-). Component of the NexGen CR-Flex Porous Femur System.
Out of specification violation of devices that results in a gap existing within the fixture that presses the fiber metal pad to the substrate, causing the pad to not fully bond to the substrate.
FDA · DeviceZimmer, Inc.Worldwide distribution - US Nationwide including the states of AK, AL, CA, FL,…Official notice →- 2023-09-13
US
Zimmer, Inc. recalls M/L Taper Hip Prosthesis Extended Offset, Reduced Neck Length, Size 4 (Item Number 00771100640) and Size 6 (Item Number 00771100440)
The outer package labeling and product etch are for a Size 6, however, the implant is a Size 4, and vice versa.
FDA · DeviceZimmer, Inc.Worldwide - US Nationwide distribution in the states of AL, CA, GA, MI, NJ,…Official notice →
Other drugs & medical recalls in the United States
- 2024-08-07
US
Beckman Coulter Mishima K.K. recalls DxC 500 AU Clinical Chemistry Analyzer, REF C63520 is an automated chemistry analyzer that measures analytes in samples, in…
There are two software bug issues with the analyzer. Issue 1: When a dedicated rack is reserved for calibration, after a calibration test has been placed and completed, the same rack may be used to place both a patient…
FDA · DeviceBeckman Coulter Mishima K.K.Worldwide - U.S. Nationwide distribution in the states of AL, CA, GA, HI, IL,…Official notice →US
Angiodynamics, Inc. recalls SOFT-VU KUMPE 5F X 40CM -Used for patients in need of angiographic diagnosis. Catalog Number: 10732702
Mislabeled. The shelf-box labeling (unit of sale) is accurate for the packaged devices, however the inner pouch label may contain information for a different model (catheter length)
US
Boston Scientific Corporation recalls Capio SLIM Suture Capturing Device, Box of 5, Outer Package UPN: M0068318261, Inner Package UPN: M0068318260. Capio SLIM Suture…
Some lots of Capio SLIM do not consistently catch the suture after activation possibly resulting in a prolonged procedure or additional medical intervention.
FDA · DeviceBoston Scientific CorporationDomestic: Nationwide Distribution. International: PendingOfficial notice →US
EVIDENT SCIENTIFIC INC recalls SlideView VS200 slide viewer,
A broken fuse inside the laser that was caused by a short circuit at PIN 4 of the DOOR CTRL. This short circuit was caused by a badly positioned cable that made contact with the housing/frame and caused a short circuit…
US
Eugia US LLC recalls Progesterone Injection, USP, 500 mg per 10 mL (50mg/mL), 10 mL Multiple-Dose Vial, Rx Only, Mfd. in India for: Eugia US LLC, E Windsor, NJ…
Presence of Particulate Matter: Complaint received of a glass particle in the vial.
US
FDC Limited recalls Ciprofloxacin ophthalmic solution USP, 0.3% as base, package in bottles: a) 10 mL (NDC 69315-308-10), b) 2.5 mL (NDC 69315-308-02), Rx…
Defective container: unable to get the solution out of the bottle as the spike of the cap was lodged in the nozzle of the product bottle.