Drugs & medicalClass II

EVIDENT SCIENTIFIC INC recalls SlideView VS200 slide viewer,

A broken fuse inside the laser that was caused by a short circuit at PIN 4 of the DOOR CTRL. This short circuit was caused by a badly positioned cable that made contact with the housing/frame and caused a short circuit…

Read the official notice →
Date
2024-08-07
Country
United States
Brand
EVIDENT SCIENTIFIC INC
Product
SlideView VS200 slide viewer,
Classification
Class II
Units
US
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

Other drugs & medical recalls in the United States

  1. 2024-08-07
  2. US
    Drugs & medicalClass II

    Angiodynamics, Inc. recalls SOFT-VU KUMPE 5F X 40CM -Used for patients in need of angiographic diagnosis. Catalog Number: 10732702

    Mislabeled. The shelf-box labeling (unit of sale) is accurate for the packaged devices, however the inner pouch label may contain information for a different model (catheter length)

    FDA · DeviceAngiodynamics, Inc.US Nationwide distribution.Official notice →
  3. US
    Drugs & medicalClass II

    Boston Scientific Corporation recalls Capio SLIM Suture Capturing Device, Box of 5, Outer Package UPN: M0068318261, Inner Package UPN: M0068318260. Capio SLIM Suture…

    Some lots of Capio SLIM do not consistently catch the suture after activation possibly resulting in a prolonged procedure or additional medical intervention.

    FDA · DeviceBoston Scientific CorporationDomestic: Nationwide Distribution. International: PendingOfficial notice →
  4. US
    Drugs & medicalClass II

    Zimmer, Inc. recalls CPT Hip Joint Metal/Polymer Semi-Constrained Cemented Prosthesis, used in Hip Arthroplasty; Item Numbers 00811400000 00811400010…

    Affected product has an increased risk of postoperative perisprosthetic femoral fracture (PFF). The IFU is being updated to reflect the risk of PFF.

    FDA · DeviceZimmer, Inc.Worldwide - US Nationwide distribution.Official notice →
  5. US
    Drugs & medicalClass II

    Eugia US LLC recalls Progesterone Injection, USP, 500 mg per 10 mL (50mg/mL), 10 mL Multiple-Dose Vial, Rx Only, Mfd. in India for: Eugia US LLC, E Windsor, NJ…

    Presence of Particulate Matter: Complaint received of a glass particle in the vial.

    FDA · DrugEugia US LLCNationwide in the USAOfficial notice →
  6. US
    Drugs & medicalClass II

    FDC Limited recalls Ciprofloxacin ophthalmic solution USP, 0.3% as base, package in bottles: a) 10 mL (NDC 69315-308-10), b) 2.5 mL (NDC 69315-308-02), Rx…

    Defective container: unable to get the solution out of the bottle as the spike of the cap was lodged in the nozzle of the product bottle.

    FDA · DrugFDC LimitedU.S. A. NationwideOfficial notice →
  7. US
    Drugs & medicalClass I

    Endo Pharmaceuticals, Inc. recalls Clonazepam Orally Disintegrating Tablets, USP (C-IV) 0.125mg, 60 tablets per carton (10 blister cards containing 6 tablets each), Rx…

    Labeling: Label Error on Declared Strength; Some cartons were incorrectly labeled as 0.125 mg instead of 0.25 mg. The blister strips inside the product carton reflect the correct strength of 0.25 mg.

    FDA · DrugEndo Pharmaceuticals, Inc.Nationwide in the USA and Puerto RicoOfficial notice →