SunMed Holdings, LLC recalls Medline manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. Adult Resuscitator REF #s:…
A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
Read the official notice →- Date
- 2023-11-29
- Country
- United States
- Brand
- SunMed Holdings
- Product
- Medline manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. Adult Resuscitator REF #s:…
- Classification
- Class II
- Units
- US Nationwide distribution including in the states of AR, AZ, CA¿CO, CT, FL,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More SunMed Holdings recalls
- 2026-02-04
US
SunMed Holdings, LLC recalls Brand Name: Broselow Product Name: Broselow Pediatric Emergency Rainbow Tape Model/Catalog Number: 7700RE Software Version: Not Applicable…
Product contains potentially harmful dosing errors for Vecuronium, Flumazenil, and Ketamine.
FDA · DeviceSunMed Holdings, LLCWorldwide distribution - US Nationwide and the countries of Canada, Hong Kong,…Official notice →- 2025-07-02
US
SunMed Holdings, LLC recalls Broselow Pediatric Emergency Rainbow Tape (distribution by AirLife); REF 7700REA AirLife brand, 2025 Edition, and 36-23446 Rev 2 Print…
The impacted tape was manufactured with incorrect information on the tape. Incorrect values are printed in the Red zone, Orange zone, and Grey zone. Using the incorrect values in the Red zone could lead to shocking a…
FDA · DeviceSunMed Holdings, LLCUS distribution nationwide. International distribution to Canada.Official notice →- 2025-06-11
US
SunMed Holdings, LLC recalls Adult Manual Resuscitator with Medium Adult Mask, Bag REservoir, Filter, Manometer and 7 ft Oxygen Tubing, REF CPRM1116F
Affected lots were manufactured with B/V Filter incorrectly attached to the wrong port (patient port instead of the exhalation port). If not noticed prior to patient use, there would be interruption or delay in patient…
- 2023-02-01
US
SunMed Holdings, LLC recalls Air-Q3 AirQ3 Standard ILA with Manual Cuff Inflation and GA Size REF Number 0.0 50005 0.5 50055 1.0 50105 1.5 50155 2.0 50205 3.0 50305…
There is an orogastric (OG) tube size discrepancy between the labeling in the IFU and the printed text on the device. If the clinician is unable to pass a gastric catheter through one of the channels, the clinician may…
FDA · DeviceSunMed Holdings, LLCUS Nationwide distribution in the states of AL, CA, FL, MI, OH, TN, TX, WA.Official notice →- 2022-10-05
US
SunMed Holdings, LLC recalls Full Kit Pulset 3cc Syringe 23g x 1 Safety Tip 25u BH, REF Number 3373-91
The kits are being recalled due to a lack of sterility of a spare needle contained within the kits. There is a potential for skin infection or sepsis if the patient is exposed to a non-sterilized, unused needle.
FDA · DeviceSunMed Holdings, LLCDomestic distribution to AL, FL, GA, IL, LA, MO, NJ, NY, TN, TX.Official notice →
Other drugs & medical recalls in the United States
- 2023-11-29
US
Boston Scientific Corporation recalls EXALT Model D Single-Use Duodenoscope: Outer box UPN# M00542420 & Outer box UPN# M0054242CE0.
Poor image quality due to fluid ingress in the lens.
FDA · DeviceBoston Scientific CorporationWorldwide - US Nationwide distribution including in the states of AL, AR, AZ,…Official notice →US
Philips North America Llc recalls SmartPath to dStream for 1.5T
Potential component failure in the Gradient Coil could product smoke and/or fire.
FDA · DevicePhilips North America LlcDomestic: Nationwide Distribution; Foreign: Albania, Andorra, Argentina,…Official notice →US
Biomet, Inc. recalls Compress Device Short Anchor Plug, 26 MM-For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement.…
Anchor plugs potentially have metal burrs in the transverse holes of the device, preventing the drill or pins from passing through, may result in extension of surgery
FDA · DeviceBiomet, Inc.Worldwide distribution - US Nationwide and the countries of ARGENTINA,…Official notice →US
Avanos Medical, Inc. recalls AVANOS BALLARD Closed Suction System for Adults, 14 Fr, DSE, MDI, Product code 221038, provide suction on patients on mechanical ventilation
Certain lots of BALLARD Closed Suction System for Adults, 14 Fr, DSE, MDI were labeled as not being made with Di(2-ethylhexyl) phthalate (DEHP); however, Avanos has determined that the tubing components of the product…
US
Instrumentation Laboratory recalls GEM Premier 3000 PAK - pH, pCO2, pO2, Hct, Na+, K+, Ca++, Part Numbers 00024307507, 00024315007, 00024330007, 00024345007, 00024360007
Internal testing identified that the GEM Premier PAKs (cartridges) may exhibit an over-recovery (i.e., positive bias) of pO2 results in the low range (at pO2 levels less than 60 mmHg) that are outside of labeled…
US
Sage Products Inc recalls stryker Sage PrimaFit External Urine Management System for the Female Anatomy
A small percentage of devices in three lots of product may contain a tape with a natural rubber latex adhesive which is not declared on the label