Drugs & medicalClass II

SunMed Holdings, LLC recalls Full Kit Pulset 3cc Syringe 23g x 1 Safety Tip 25u BH, REF Number 3373-91

The kits are being recalled due to a lack of sterility of a spare needle contained within the kits. There is a potential for skin infection or sepsis if the patient is exposed to a non-sterilized, unused needle.

Read the official notice →
Date
2022-10-05
Country
United States
Brand
SunMed Holdings
Product
Full Kit Pulset 3cc Syringe 23g x 1 Safety Tip 25u BH, REF Number 3373-91
Classification
Class II
Units
Domestic distribution to AL, FL, GA, IL, LA, MO, NJ, NY, TN, TX.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More SunMed Holdings recalls

  1. 2026-02-04
  2. US
    Drugs & medicalClass I

    SunMed Holdings, LLC recalls Brand Name: Broselow Product Name: Broselow Pediatric Emergency Rainbow Tape Model/Catalog Number: 7700RE Software Version: Not Applicable…

    Product contains potentially harmful dosing errors for Vecuronium, Flumazenil, and Ketamine.

    FDA · DeviceSunMed Holdings, LLCWorldwide distribution - US Nationwide and the countries of Canada, Hong Kong,…Official notice →
  3. 2025-07-02
  4. US
    Drugs & medicalClass I

    SunMed Holdings, LLC recalls Broselow Pediatric Emergency Rainbow Tape (distribution by AirLife); REF 7700REA AirLife brand, 2025 Edition, and 36-23446 Rev 2 Print…

    The impacted tape was manufactured with incorrect information on the tape. Incorrect values are printed in the Red zone, Orange zone, and Grey zone. Using the incorrect values in the Red zone could lead to shocking a…

    FDA · DeviceSunMed Holdings, LLCUS distribution nationwide. International distribution to Canada.Official notice →
  5. 2025-06-11
  6. US
    Drugs & medicalClass I

    SunMed Holdings, LLC recalls Adult Manual Resuscitator with Medium Adult Mask, Bag REservoir, Filter, Manometer and 7 ft Oxygen Tubing, REF CPRM1116F

    Affected lots were manufactured with B/V Filter incorrectly attached to the wrong port (patient port instead of the exhalation port). If not noticed prior to patient use, there would be interruption or delay in patient…

    FDA · DeviceSunMed Holdings, LLCUS Nationwide distribution via Medline.Official notice →
  7. 2023-11-29
  8. US
    Drugs & medicalClass II

    SunMed Holdings, LLC recalls Medline manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. Adult Resuscitator REF #s:…

    A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.

    FDA · DeviceSunMed Holdings, LLCUS Nationwide distribution including in the states of AR, AZ, CA¿CO, CT, FL,…Official notice →
  9. 2023-02-01
  10. US
    Drugs & medicalClass II

    SunMed Holdings, LLC recalls Air-Q3 AirQ3 Standard ILA with Manual Cuff Inflation and GA Size REF Number 0.0 50005 0.5 50055 1.0 50105 1.5 50155 2.0 50205 3.0 50305…

    There is an orogastric (OG) tube size discrepancy between the labeling in the IFU and the printed text on the device. If the clinician is unable to pass a gastric catheter through one of the channels, the clinician may…

    FDA · DeviceSunMed Holdings, LLCUS Nationwide distribution in the states of AL, CA, FL, MI, OH, TN, TX, WA.Official notice →
All 6 SunMed Holdings recalls →

Other drugs & medical recalls in the United States

  1. 2022-10-05
  2. US
    Drugs & medicalClass III

    Mckesson Medical-Surgical Inc. Corporate Office recalls SPS MEDICAL SUPPLY Crosstex International Sterilization Roll Sani-Roll Ethylene Oxide (EO) Gas / Steam 3 Inch X…

    Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.

    FDA · DeviceMckesson Medical-Surgical Inc. Corporate OfficeUS Nationwide distribution.Official notice →
  3. US
    Drugs & medicalClass II

    Micro-X Ltd. recalls MICRO-X ROVER MOBILE X-RAY SYSTEM, MXU-RV19

    Software calibration error with product equip with a Dose Area Product (DAP) meter. This results in the product displaying the incorrect DAP meter values which may impede radiation exposure management decisions

    FDA · DeviceMicro-X Ltd.Worldwide distribution - U.S. Nationwide distribution in the states of CA, NV,…Official notice →
  4. US
    Drugs & medicalClass II

    CareFusion 303, Inc. recalls BD Pyxis MedBank Cabinet Software v2.1.2.9 Legacy MedBank Cabinets with Matrix Drawers include: . CUBEX MedFlex operating with Windows 7 .…

    The automated dispensing cabinet software update included a .dll file that is not compatible with MedBank Cabinet stations with MedCart Matrix drawers. The cabinet system will stop functioning (Matrix drawers will stop…

    FDA · DeviceCareFusion 303, Inc.U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO,…Official notice →
  5. US
    Drugs & medicalClass II

    Howmedica Osteonics Corp. recalls Short tibial bearing MK4 - Std, STM, Catalog No. SMMLTB02. Knee prosthesis component.

    A MK4 Tibial Bearing was supplied with the intention to mate with the Tibial Metal Casing in-situ. However, the device in-situ was a MK3 Tibial Metal Casing. The supplied MK4 tibial bearing was not exchangeable for the…

    FDA · DeviceHowmedica Osteonics Corp.Distributed to one customer located in New York.Official notice →
  6. US
    Drugs & medicalClass II

    Collagen Matrix Inc recalls Collagen Dental Wound Dressing - Bovine Dermis Tap, 25mm x 75mm Product was distributed under the following brand names: 1) Cytoplast…

    The packaging may not be sealed, which could compromise sterility.

    FDA · DeviceCollagen Matrix IncUS distribution to distributors in IL, IN, MI, NC, OH, and TXOfficial notice →
  7. US
    Drugs & medicalClass I

    Cardiac Assist, Inc recalls TANDEM LIFE LivaNova LIFESPARC Controller; Controller and Accessories Model 5900-0000; Controller Only Model 5900-0001. For circulatory or…

    Critical Failure of the LifeSPARC controller when the software freezes or crashes and the screen does not display data.

    FDA · DeviceCardiac Assist, IncUS Nationwide distribution.Official notice →
See all drugs & medical recalls →