Stryker GmbH recalls Stryker Hoffman LRF Wire Tensioner REF 4933-9-100, a component of the Hoffmann LRF System.
The adjustment ring on the device may become loose resulting in the wire tension not being able to be set correctly.
Read the official notice →- Date
- 2024-07-03
- Country
- United States
- Brand
- Stryker
- Product
- Stryker Hoffman LRF Wire Tensioner REF 4933-9-100, a component of the Hoffmann LRF System.
- Classification
- Class II
- Units
- Worldwide - US Nationwide distribution in the states of AZ, CA, CO, FL, GA,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Stryker recalls
- 2024-10-23
US
Stryker Corporation recalls 1.2mm Cross Cut Fissure - Intended for use in the cutting, drilling, reaming, decorticating, shaping and smoothing of bone, bone cement…
May exhibit endotoxin levels greater than 2.15 EU/device, fever and inflammation may occur from endotoxin exposure and meningitis may only occur if there is contact with cerebrospinal fluid and thus,…
FDA · DeviceStryker CorporationWorldwide distribution - US Nationwide and the countries of Canada, Columbia.Official notice →- 2024-09-04
US
Stryker Corporation recalls Stryker iBur 3.0mm Coarse Diamond Round, Distal Bend - intended to cut bone in the following manner: drilling, reaming, decorticating,…
Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tissue/structure…
FDA · DeviceStryker CorporationWorldwide distribution - US Nationwide and the countries of Australia, Canada,…Official notice →- 2024-07-10
US
Stryker Corporation recalls The Non-Sterile Flyte hood covers the user s head and neck region and is intended to be worn with commercially available gowns. The hood…
Expired Products distributed to customers
FDA · DeviceStryker CorporationUS Nationwide distribution in the states of Arkansas, California, Colorado,…Official notice →- 2024-02-21
US
Stryker, Inc. recalls Stryker Neurovascular, Trevo Trak 21 Microcatheter, REF 90338, 0.021in X162cm
The microcatheter IFU contains an intended use that has not been evaluated by the FDA.
FDA · DeviceStryker, Inc.US Nationwide distribution in the states of AL, FL, GA, HI, IL, MI, NC, NH,…Official notice →- 2023-11-15
US
US
Stryker Corporation recalls Disposable Spetzler-Malis Bipolar forceps, 23CM SM DISP BAYONET 1.5MM TIP, Catalog Number 6760-230-015
An error was identified on the expiration date of the product label. The shelf life for the products impacted is 36 months (3 years) however the product label represents a 54 month (4.5 year) shelf life.
FDA · DeviceStryker CorporationUS Nationwide. Australia, Canada, Japan, Korea, Singapore.Official notice →
Other drugs & medical recalls in the United States
- 2024-07-03
US
Cook Incorporated recalls Ultrathane Mac-Loc Locking Loop Multipurpose Drainage Catheter ULT16.0-38-25-P-6S-MCL-HC--Intended for percutaneous drainage in a variety…
Product labels state the incorrect expiration dates that extends beyond their true expiration date
US
ASPEN SURGICAL recalls Endoscopic Kittner Blunt Dissecting Instrument, Intended to be used for blunt dissection of tissue, Model Number KT-9101
The affected products may have been packaged with a defect in the sterile barrier seal. As a result, the integrity of the sterile barrier may be compromised.
US
Centers For Disease Control and Prevention recalls CDC, Influenza A/H5 Subtyping Kit, Model/Catalog Number: FluIVD03-11
The H5b component may fail to amplify, resulting in an inconclusive result.
US
Merit Medical Systems, Inc. recalls Prelude Roadster Guide Sheath REF: -PG4F45S -PG4F90S -PG5F45MP -PG5F45R -PG5F45S -PG5F45S018 -PG5F90MP -PG5F90S -PG5F90S018 -PG6F45HS…
Due to older version of Instruction for Use (IFU) instructing to ensure that the value and the sheath are tightly connected before use. This may cause users to over-tightening the luer which may result in the swivel…
FDA · DeviceMerit Medical Systems, Inc.Worldwide - U.S. Nationwide distribution including in the states of AK, AL,…Official notice →US
NovaSignal Corp. recalls NovaGuide 2 Traditional Ultrasound , REF: NSC-TCDNG2
Due to inadvertently disconnecting grounding cable of ultrasound system could result in electrical shock, electrical interference and/or equipment malfunction.
FDA · DeviceNovaSignal Corp.Worldwide - US Nationwide distribution in the states of MA, TX, NY, MN, AZ,…Official notice →US
Thoratec Switzerland GMBH recalls 2nd Generation CentriMag Primary Console, REF: 201-90401, 201-90701, 201-90411, 201-90421, 201-30300, L201-90401, L201-90411,…
Due to a component change, the console part of a circulatory support system may not conform to the IEC 61000-4-5 standard, so if subjected to a power surge above 1.8kV and up to 2.0 kV, the console may shut down…
FDA · DeviceThoratec Switzerland GMBHWorldwide - US Nationwide distribution included in the states of CA, PA, FL,…Official notice →