Stryker Corporation recalls Stryker iBur 3.0mm Coarse Diamond Round, Distal Bend - intended to cut bone in the following manner: drilling, reaming, decorticating,…
Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tissue/structure…
Read the official notice →- Date
- 2024-09-04
- Country
- United States
- Brand
- Stryker
- Product
- Stryker iBur 3.0mm Coarse Diamond Round, Distal Bend - intended to cut bone in the following manner: drilling, reaming, decorticating,…
- Classification
- Class II
- Units
- Worldwide distribution - US Nationwide and the countries of Australia, Canada,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Stryker recalls
- 2025-01-15
US
Stryker Corporation recalls 100V NEPTUNE 3 ROVER JAPANESE (0703-003-000) 120V NEPTUNE S ROVER (0711-001-000) 230V NEPTUNE 3 ROVER ENGLISH (0703-002-000ENG) 230V…
Bolts on the vacuum pump filter housing and the flange plate were not properly tightened, impacting the function of the Neptune rovers. Systems with loose bolts may fail to provide sufficient suction force or loss of…
FDA · DeviceStryker CorporationUS (Domestic) distribution to states of: WA, GA, and IL. OUS (International)…Official notice →- 2024-11-20
US
Stryker Corporation recalls Stryker Neptune SafeAir Rocker-Switch Smoke Evacuation Pencil, Coated, 70mm - Designed for general electrosurgical applications including…
There is a potential for the product to activate without manual input when plugged into a power source or remain active after buttons are released. Unintentional activation of the device may result in localized burns…
FDA · DeviceStryker CorporationWorldwide distribution - US Nationwide and the countries of Canada, Japan,…Official notice →- 2024-10-23
US
Stryker Corporation recalls 1.2mm Cross Cut Fissure - Intended for use in the cutting, drilling, reaming, decorticating, shaping and smoothing of bone, bone cement…
May exhibit endotoxin levels greater than 2.15 EU/device, fever and inflammation may occur from endotoxin exposure and meningitis may only occur if there is contact with cerebrospinal fluid and thus,…
FDA · DeviceStryker CorporationWorldwide distribution - US Nationwide and the countries of Canada, Columbia.Official notice →- 2024-07-10
US
Stryker Corporation recalls The Non-Sterile Flyte hood covers the user s head and neck region and is intended to be worn with commercially available gowns. The hood…
Expired Products distributed to customers
FDA · DeviceStryker CorporationUS Nationwide distribution in the states of Arkansas, California, Colorado,…Official notice →- 2024-07-03
US
Stryker GmbH recalls Stryker Hoffman LRF Wire Tensioner REF 4933-9-100, a component of the Hoffmann LRF System.
The adjustment ring on the device may become loose resulting in the wire tension not being able to be set correctly.
FDA · DeviceStryker GmbHWorldwide - US Nationwide distribution in the states of AZ, CA, CO, FL, GA,…Official notice →- 2024-02-21
US
Stryker, Inc. recalls Stryker Neurovascular, Trevo Trak 21 Microcatheter, REF 90338, 0.021in X162cm
The microcatheter IFU contains an intended use that has not been evaluated by the FDA.
FDA · DeviceStryker, Inc.US Nationwide distribution in the states of AL, FL, GA, HI, IL, MI, NC, NH,…Official notice →
Other drugs & medical recalls in the United States
- 2024-09-04
US
Schiller, Ag recalls CARDIOVIT AT-180. Model Number: 0A.110000; Catalog number: 3.920570
Potential for high-frequency signal artifacts is recorded during an ECG acquisition performed by CARDIOVIT AT-180 electrocardiographs.
FDA · DeviceSchiller, AgWorldwide - US distribution in the state of Florida and the countries of…Official notice →US
GE Healthcare (China) Co., Ltd. recalls GE Healthcare Proteus XR/a, stationary X-Ray System, Model/Catalog Numbers: 1) 2259973; 2) 2259988; 3) 2266999; 4) 2373519; 5)…
The XR system, Proteus XR/a, does not have a de-installation manual describing process steps for de-installing the devices. De-installation instruction process steps must be followed when de-installing these devices to…
US
Howmedica Osteonics Corp. recalls Custom Made Device - GMRS EXTENSION Piece 40MM LEFT, 35 Deg EXT Rotated Tabs (Patient Specific); Catalog Number: C-MM100434-00;
A patient specific, custom-made device was supplied for a surgical procedure that did not match the design proposal.
FDA · DeviceHowmedica Osteonics Corp.International distribution to the country of Australia.Official notice →US
Intuitive Surgical, Inc. recalls Ion Endoluminal System, REF: 380748-65
There is a potential that the robotic system's instrument cart arm may contain screws that may potentially fail which could result in uncontrolled motion of a catheter inside a patient's airways.
FDA · DeviceIntuitive Surgical, Inc.US Nationwide distribution in the states of AZ and CA.Official notice →US
Neurovision Medical Products Inc recalls Cobra, 3-Plate EMG ET Tube, 6.0mm, REF: LTE7003PM, not made with natural rubber latex, Rx Only, STERILEEO
mislabeling; EMG RLN monitoring kit labeling contains the incorrect tube size.
US
Treace Medical Concepts, Inc. recalls Nanoplasty 3D Bunion Correction System Implant, Model Number: SK57. System contains the 01-00066 1st Metatarsal Plate as a…
Potential issue causing a higher frequency of interference with locking screws and/or inserter rods than anticipated, resulting in screws stripping and/or not fully seating in the plates and/or interface issues between…
FDA · DeviceTreace Medical Concepts, Inc.US Nationwide distribution in the states of CO, FL, MD, MI, NC, SC, TX, VA.Official notice →