Drugs & medicalClass II

Smiths Medical ASD Inc. recalls Medfusion Syringe Pump, Model 3500-500, v3, v4, v5, and v6

A force sensor in the occlusion detector may drift out of calibration leading to increased occlusion detection times, false occlusion alarms, or a System Failure Alarm. If the force sensor calibration shift is large…

Read the official notice →
Date
2023-10-18
Country
United States
Brand
Smiths Medical ASD
Product
Medfusion Syringe Pump, Model 3500-500, v3, v4, v5, and v6
Classification
Class II
Units
Worldwide distribution - US Nationwide and the country of Canada.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Smiths Medical ASD recalls

  1. 2024-01-24
  2. US
    Drugs & medicalClass II

    Smiths Medical Asd Inc recalls 48IN (121.9CM) INJ LINE ROTATING ADAPT (900psi), List Number MX694R

    There is a manufacturing defect affecting specific rotator lots. Within this population, the inner diameter of the devices O-ring may be oversized affecting seal integrity. An inadequate seal may lead to a leak during…

    FDA · DeviceSmiths Medical Asd IncWorldwide distribution - US Nationwide and the countries of Canada, UAE.Official notice →
  3. 2023-11-08
  4. US
    Drugs & medicalClass II

    Smiths Medical ASD Inc. recalls Smiths Medical Portex Bivona Pediatric Tracheostomy Tube, Cuffless, Model 60P045, Sterile.

    The unit box is labeled with an incorrect expiration date (the manufacturing date of 3/17/2021) but the individual packaging contains the correct expiration date (3/17/2026).

    FDA · DeviceSmiths Medical ASD Inc.Distribution was made to AZ, CA, FL, HI, IL, IN, MA, MN, MO, MS, NC, NJ, OH,…Official notice →
  5. 2023-11-01
  6. US
    Drugs & medicalClass II

    Smiths Medical ASD Inc. recalls Portex Pro-Vent Arterial Blood Sampling Kit with Dry Lithium Heparin for Gases and Electrolytes, REF 4599P-1

    The complaint involved the absence of a component called "ASM Filter-Pro ABG SYR CAP" from the packaging of Pro-Vent Arterial Blood Sampling (ABG) Kits. Specifically, this issue affected a total of 31,900 kits under…

    FDA · DeviceSmiths Medical ASD Inc.USOfficial notice →
  7. 2023-09-06
  8. US
    Drugs & medicalClass II

    Smiths Medical ASD Inc. recalls PORTEX 0.5u HYDROPHOBIC DISC FILTER, REF 002291 - Bacterial/Viral filter

    Smiths Medical identified a labeling error in which the barcode (UDI) was not correctly reflecting the expiration date of the product

    FDA · DeviceSmiths Medical ASD Inc.US Nationwide distribution.Official notice →
  9. 2023-05-24
  10. US
    Drugs & medicalClass II

    Smiths Medical ASD Inc. recalls DuraLife Autoclavable Double Swivel Elbow, REF 60-0010 (UltraSet Product Code 66-1991); Swivel Elbow with Suction Port for use with…

    DuraLife devices instructions for cleaning and sterilization have been determined to be inadequate. Specifically, for the DuraLife product, the IFU does not provide the cleaning method, duration of autoclave cycle or…

    FDA · DeviceSmiths Medical ASD Inc.Worldwide distribution - US Nationwide and the countries of Canada, New…Official notice →
  11. 2023-02-08
  12. US
    Drugs & medicalClass I

    Smiths Medical ASD Inc. recalls CADD Administration Set, List Numbers: 1) REF 21-7300-24, 2) REF 21-7301-24 3) REF 21-7302-24, 4) REF 21-7321-24, 5) REF 21-7322-24, 6)…

    There are two potential issues with CADD Infusion System Infusion Sets: (1) Lack of Delivery or Underdelivery related to Tubing Occlusion and (2) False "No Disposable Attached (NDA)" Alarms

    FDA · DeviceSmiths Medical ASD Inc.Worldwide Distribution: US (nationwide) and OUS countries of: CH, CA, ES, FR,…Official notice →
All 39 Smiths Medical ASD recalls →

Other drugs & medical recalls in the United States

  1. 2023-10-18
  2. US
    Drugs & medicalClass II

    Cardinal Health 200, LLC recalls Procedure packs: (1) Presource PBDS Cat. PN33BPCR2, Kit, Back, Sterile; (2) Presource PBDS Cat. POCGANA11, Kit, ARN Anterior Hip,…

    XXX

    FDA · DeviceCardinal Health 200, LLCWorldwide distribution - US Nationwide distribution in the states of AZ, CA,…Official notice →
  3. US
    Drugs & medicalClass II

    Cardinal Health 200, LLC recalls Procedure pack: Presource PBDS Cat. PLMHCSBHF, Hermann Hospital MHHS, Kit, C Section B MHHS, Sterile.

    Packs contain recalled components (equipment drapes, table cover, and surgical room turnover (SRT) kits).

    FDA · DeviceCardinal Health 200, LLCWorldwide distribution - US Nationwide distribution in the states of AZ, CA,…Official notice →
  4. US
    Drugs & medicalClass II

    RoyalVibe Health Ltd. recalls CellQuicken Analyzer (Smart-Watch and Software), RoyalVibe Ultrasound Generator, Brainvibe, Envirovibe Water Restructuring Pad, RoyalVibe…

    Ultrasound devices were not authorized, cleared, or approved by FDA, and thus their performance characteristics have not been adequately established. There is a possibility of customers' reliance on marketing claims…

    FDA · DeviceRoyalVibe Health Ltd.US: MI, PA, TX, GA, OH, CA, FL, NY, NH, UT, AL, WA. WI, MN, MA, NM, IA, MD,…Official notice →
  5. US
    Drugs & medicalClass II

    Flower Orthopedics Corporation recalls Coventus Flower Orthopedics Flex-Thread TM 3.2mm Cannulated Reamer-used to ream and subsequently open the intramedullary canal to…

    Device is breaking while in use, potential for patients to require a revision surgery.

    FDA · DeviceFlower Orthopedics CorporationUS Nationwide distribution.Official notice →
  6. US
    Drugs & medicalClass II

    LivaNova Deutschland GmbH recalls 3T Heater Cooler System, REF: 16-02-85

    The firm's labeling (Quick Reference Guide) contains incorrect information in the maintenance schedule for the Heater-Cooler device.

    FDA · DeviceLivaNova Deutschland GmbHUS Nationwide distribution in the states of CA, IN, GA, NC, TX, AZ, FL, MN,…Official notice →
  7. US
    Drugs & medicalClass I

    Fresenius Medical Care Holdings, Inc. recalls 2008T HD SYS. W/O CDX BLUESTAR

    Potential PCBA leaching from tubing of hemodialysis machines.

    FDA · DeviceFresenius Medical Care Holdings, Inc.Domestic: Nationwide Distribution.Official notice →
See all drugs & medical recalls →