Smiths Medical ASD Inc. recalls DuraLife Autoclavable Double Swivel Elbow, REF 60-0010 (UltraSet Product Code 66-1991); Swivel Elbow with Suction Port for use with…
DuraLife devices instructions for cleaning and sterilization have been determined to be inadequate. Specifically, for the DuraLife product, the IFU does not provide the cleaning method, duration of autoclave cycle or…
Read the official notice →- Date
- 2023-05-24
- Country
- United States
- Brand
- Smiths Medical ASD
- Product
- DuraLife Autoclavable Double Swivel Elbow, REF 60-0010 (UltraSet Product Code 66-1991); Swivel Elbow with Suction Port for use with…
- Classification
- Class II
- Units
- Worldwide distribution - US Nationwide and the countries of Canada, New…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Smiths Medical ASD recalls
- 2023-11-01
US
Smiths Medical ASD Inc. recalls Portex Pro-Vent Arterial Blood Sampling Kit with Dry Lithium Heparin for Gases and Electrolytes, REF 4599P-1
The complaint involved the absence of a component called "ASM Filter-Pro ABG SYR CAP" from the packaging of Pro-Vent Arterial Blood Sampling (ABG) Kits. Specifically, this issue affected a total of 31,900 kits under…
- 2023-10-18
US
Smiths Medical ASD Inc. recalls Medfusion Syringe Pump, Model 3500-500, v3, v4, v5, and v6
A force sensor in the occlusion detector may drift out of calibration leading to increased occlusion detection times, false occlusion alarms, or a System Failure Alarm. If the force sensor calibration shift is large…
FDA · DeviceSmiths Medical ASD Inc.Worldwide distribution - US Nationwide and the country of Canada.Official notice →- 2023-09-06
US
Smiths Medical ASD Inc. recalls PORTEX 0.5u HYDROPHOBIC DISC FILTER, REF 002291 - Bacterial/Viral filter
Smiths Medical identified a labeling error in which the barcode (UDI) was not correctly reflecting the expiration date of the product
- 2023-02-08
US
Smiths Medical ASD Inc. recalls CADD Administration Set, List Numbers: 1) REF 21-7300-24, 2) REF 21-7301-24 3) REF 21-7302-24, 4) REF 21-7321-24, 5) REF 21-7322-24, 6)…
There are two potential issues with CADD Infusion System Infusion Sets: (1) Lack of Delivery or Underdelivery related to Tubing Occlusion and (2) False "No Disposable Attached (NDA)" Alarms
FDA · DeviceSmiths Medical ASD Inc.Worldwide Distribution: US (nationwide) and OUS countries of: CH, CA, ES, FR,…Official notice →- 2022-11-30
US
Smiths Medical ASD Inc. recalls Medfusion Model 4000 Syringe Infusion Pump, with software versions V1.6.0 or V1.6.1 that have implemented the Smart Pump Programming…
Potential for delay in sending Infusion Start Messages to third-party Electronic Health Record systems which may affect infusion auto-programming on the pump and documentation
FDA · DeviceSmiths Medical ASD Inc.US distribution to states of: Indiana, Texas and VirginiaOfficial notice →- 2022-06-29
US
Smiths Medical ASD Inc. recalls Medfusion Syringe Pump Models: 3500, 3500-0600-00, 3500-0600-01, 3500-0600-249, 3500-0600-50, 3500-0600-51, 3500-0600-82, 3500-306,…
Multiple issues with the potential for interruption of therapy or over-infusion: 1. Primary Audible Alarm (PAA), 2. Unanticipated Depleted Battery Alarms, 3. Time Base Alarm, 4. Intermittent Volume Over Time (IVOT) -…
FDA · DeviceSmiths Medical ASD Inc.Worldwide distribution. US nationwide including US Virgin Islands, Countries…Official notice →
Other drugs & medical recalls in the United States
- 2023-05-24
US
Philips North America recalls Philips Incisive CT -Computed Tomography X-Ray System Model Number: 72813
Multi-Function Foot Switch Unload Pedal Foot Entrapment-Pressing the Unload Pedal of the foot switch to perform the unload function may cause entrapment of the operator's foot and injury
FDA · DevicePhilips North AmericaWorldwide - US Nationwide distribution in the states of AR, CA, FL, IL, IN,…Official notice →US
Philips North America Llc recalls Spectral CT 7500 Computed Tomography X-Ray System, Model Number 728333
Pressing the Unload Pedal of the foot switch to perform the unload function may cause entrapment of the operators foot. If foot or leg entrapment occurs while pressing Unload Pedal, the risk to operators may include:…
FDA · DevicePhilips North America LlcWorldwide - US Nationwide distribution including in the states of States: AZ,…Official notice →US
Greiner Bio-One North America, Inc. recalls greiner bio-one VACUETTE TUBE 6 ml K3E K3EDTA, 13x100 pink cap-black ring, non-ridged, 24 racks of 50 pcs., 1200 pcs in…
Some of the tubes may be incorrectly labeled.
FDA · DeviceGreiner Bio-One North America, Inc.US Nationwide distribution in the states of WI and GA.Official notice →US
ICU Medical Inc recalls Replacement Battery List Number SUB0000864, found in Plum 360 Infusion System, List Number 30010
Due to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery runtime may decrease earlier in the…
US
Draeger Medical, Inc. recalls VentStar Basic (P)180, MP00351
Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.