Drugs & medicalClass II

Siemens Healthcare Diagnostics, Inc. recalls IMMULITE 2000 Systems HCG [200 Tests, material number 10381206, catalog number L2KCG] and [600 Tests, material number…

Potential for falsely elevated hCG results due to sample carryover in serum and urine patient samples, as well as quality control samples and adjustors.

Read the official notice →
Date
2022-05-04
Country
United States
Brand
Siemens Healthcare Diagnostics
Product
IMMULITE 2000 Systems HCG [200 Tests, material number 10381206, catalog number L2KCG] and [600 Tests, material number 10381194, catalog…
Classification
Class II
Units
Worldwide distribution - US Nationwide distribution in the states of AZ, CA,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Siemens Healthcare Diagnostics recalls

  1. 2022-07-13
  2. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Dimension LTNI -In vitro diagnostic test intended to quantitively measure cardiac troponin-I levels in human serum and…

    Positive bias with Lithium Heparin plasma samples from individuals that are expected to be below the 99th percentile of 0.07 ng/mL [0.07 ¿g/L], may lead to inappropriate intervention for myocardial infarction

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide distribution in the states of AL, AR,…Official notice →
  3. 2022-06-15
  4. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Atellica IM 1300 Analyzer - automated, immunoassay analyzer designed to perform in vitro diagnostic tests on clinical…

    Lots incompatible with Test Definition (TDef) Version 1.4, potentially lead to the generation of erroneous total T3 results

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide and the countries of Albania,…Official notice →
  5. 2022-05-11
  6. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls ADVIA Centaur Enhanced Estradiol (eE2) 500T - In vitro diagnostic use in the quantitative determination of estradiol SMN:…

    Falsely Elevated Results Observed with Plasma Specimens, could lead a clinician to misinterpret patient results

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide and the countries of Austria, Belgium,…Official notice →
  7. 2022-04-20
  8. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Fructosamine (Fruc), Siemens Material Number 11097637

    Siemens Healthcare Diagnostics Inc. has confirmed the potential for reagent carryover impacting quality control (QC), patient samples, and calibrator results.

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Global Distribution. US Nationwide.Official notice →
  9. 2022-04-06
  10. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Siemens Dimension Vista Magnesium Flex reagent cartridge (MG) Material # 10445158, Catalog # K3057 used in the Dimension…

    Dimension Vista Magnesium Flex reagent cartridge lots exhibiting imprecision and intermittent Abnormal Assay flags. When this issue occurs, there is a potential for QC failures or erroneous patient results which may…

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution: US (nationwide) 47 States including: AK, AL, AR, AZ,…Official notice →
  11. 2022-01-12
  12. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Atellica CH 930 Analyzer with Software version: V1.25.1 and lower- An automated, clinical chemistry analyzer designed to…

    (1)Software (SW) versions V1.25.1 and lower may result in TDef (Test Definition) Parameters for Open Channel Assays reverting to Default Value would shift test results by the magnitude of the implemented Assay…

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Nationwide Foreign: AU DE AE TR BG CL IN CZ BR ES KE CN PL SA BE IQ IT CW SE…Official notice →
All 80 Siemens Healthcare Diagnostics recalls →

Other drugs & medical recalls in the United States

  1. 2022-05-11
  2. US
    Drugs & medicalClass II

    Macleods Pharma Usa Inc recalls Losartan Potassium Tablets, USP, 50 mg, a) 30-count bottles (NDC # 33342-045-07), b) 90-count bottles (NDC # 33342-045-10), c) 1000-count…

    CGMP Deviations- AZIDO Impurity levels observed to be above acceptable limits

    FDA · DrugMacleods Pharma Usa IncProduct was distributed nationwide.Official notice →
  3. US
    Drugs & medicalClass II

    SCA Pharmaceuticals recalls Vancomycin HCl 1.5 g in 0.9% Sodium Chloride, 500 mL bags, Rx only, SCA Pharmaceuticals Windsor, CT 06095, NDC 70004-0924-44

    CGMP Deviations

    FDA · DrugSCA PharmaceuticalsProduct was distributed nationwide within the United StatesOfficial notice →
  4. 2022-05-04
  5. US
    Drugs & medicalClass II

    Freudenberg Medical, Llc recalls Empower Introducer Sheath with Hydrophilic Coating 21F x 56cm Model: 668782-100 (US)

    Non-intended glue from the hub/shaft bond on the inside ID of the catheter shaft may cause Embolism leading to peripheral ischemia

    FDA · DeviceFreudenberg Medical, LlcUS distribution: GA, OH Foreign distribution: NetherlandsOfficial notice →
  6. US
    Drugs & medicalClass II

    Instrumentation Laboratory recalls ACL TOP 300 CTS/350 CTS/500 CTS/550 CTS/970 CL: Name Part Number 1.ACL TOP 300 CTS 00000280060 2. ACL TOP 300 CTS 00000280060R…

    Updated and revised user instructions issued 7/14/22. Potential carryover issue may cause elevated quality control and sample results for heparin, apixaban or rivaroxaban when HemosIL Liquid Anti-Xa reagent is used in…

    FDA · DeviceInstrumentation LaboratoryWorldwide distribution - US Nationwide and the countries of Argentina,…Official notice →
  7. US
    Drugs & medicalClass II

    Waters Medical Systems LLC recalls WAVES WCM Kidney Cassette Module

    There is a potential for the WCM cassette pump head leaking solution.

    FDA · DeviceWaters Medical Systems LLCWCM cassettes have been distributed to one domestic customer with two…Official notice →
  8. US
    Drugs & medicalClass II

    Draegar Medical Systems, Inc. recalls Drager Infinity CentralStation Software versions VG2.1.2 and VG2.1.2 SU12. For patient monitoring with Infinity M300 wireless…

    Software issue resulting in temporary loss of central monitoring functionality.

    FDA · DeviceDraegar Medical Systems, Inc.Worldwide distribution - US Nationwide distribution in the states of AL, AR,…Official notice →
See all drugs & medical recalls →