Drugs & medicalClass II

Siemens Healthcare Diagnostics, Inc. recalls Dimension LTNI -In vitro diagnostic test intended to quantitively measure cardiac troponin-I levels in human serum and…

Positive bias with Lithium Heparin plasma samples from individuals that are expected to be below the 99th percentile of 0.07 ng/mL [0.07 ¿g/L], may lead to inappropriate intervention for myocardial infarction

Read the official notice →
Date
2022-07-13
Country
United States
Brand
Siemens Healthcare Diagnostics
Product
Dimension LTNI -In vitro diagnostic test intended to quantitively measure cardiac troponin-I levels in human serum and heparinized plasma…
Classification
Class II
Units
Worldwide distribution - US Nationwide distribution in the states of AL, AR,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Siemens Healthcare Diagnostics recalls

  1. 2022-09-07
  2. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Dimension Vista LOCI CA 15-3 Flex reagent cartridge. In Vitro Diagnostic for the quantitative measurement of CA 15-3.…

    Negative bias with patient samples. The mean patient sample bias range from -26% to -33%,potential risk of delay in appropriate treatment and management decisions.

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide and the countries of Canada, Austria,…Official notice →
  3. 2022-08-10
  4. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Iron_2 (Iron_2) is for in vitro diagnostic use in the quantitative determination of iron in human serum and…

    Falsely Elevated Atellica CH Microalbumin_2 (¿ALB_2) Results due to Reagent Carryover from the Iron_2 Assay. This defect occurs only when a microalbumin test is run immediately after an iron test.

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide and the countries of AR, AU, AT, BS,…Official notice →
  5. 2022-07-20
  6. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics Inc recalls epoc BGEM Test Card 25 Test Cards/box. In vitro diagnostic device for the quantitative testing of samples of heparinized or…

    Discrepant (low bias) pO2 results, could be an incorrect diagnosis of hypoxemia and initiation of unnecessary or inappropriate respiratory support.

    FDA · DeviceSiemens Healthcare Diagnostics IncAL CA CO FL IA ID KS NM OK TXOfficial notice →
  7. 2022-06-15
  8. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Atellica IM 1300 Analyzer - automated, immunoassay analyzer designed to perform in vitro diagnostic tests on clinical…

    Lots incompatible with Test Definition (TDef) Version 1.4, potentially lead to the generation of erroneous total T3 results

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide and the countries of Albania,…Official notice →
  9. 2022-05-11
  10. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls ADVIA Centaur Enhanced Estradiol (eE2) 500T - In vitro diagnostic use in the quantitative determination of estradiol SMN:…

    Falsely Elevated Results Observed with Plasma Specimens, could lead a clinician to misinterpret patient results

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide and the countries of Austria, Belgium,…Official notice →
  11. 2022-05-04
  12. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls IMMULITE 2000 Systems HCG [200 Tests, material number 10381206, catalog number L2KCG] and [600 Tests, material number…

    Potential for falsely elevated hCG results due to sample carryover in serum and urine patient samples, as well as quality control samples and adjustors.

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide distribution in the states of AZ, CA,…Official notice →
All 80 Siemens Healthcare Diagnostics recalls →

Other drugs & medical recalls in the United States

  1. 2022-07-13
  2. US
    Drugs & medicalClass II

    Abbott recalls Abbott TactiCath Sensor Enabled, Contact Force Ablation Catheter, 8F 115cm FJ, REF A-TCSE-FJ. For mapping of the heart chambers during…

    When connected to the EnSite Precision Navigation System, an affected TactiCath Contact Force Ablation Catheter, Sensor Enabled may present the error message: "invalid catheter,, or "expired catheter."

    FDA · DeviceAbbottUS Nationwide distribution.Official notice →
  3. US
    Drugs & medicalClass II

    Siemens Medical Solutions USA, Inc recalls Artis zeego (Model no. 10280959) and Artis Q.zeego (Model no. 10848283), Interventional Fluoroscopic X-Ray System

    It may occur that after system startup no stand movement is possible any longer in the event of a discharged BIOS battery of the robotic stand control PC. If this problem occurs during startup, all stand movements are…

    FDA · DeviceSiemens Medical Solutions USA, IncWorldwide distribution - US Nationwide distribution in the states of AL, AR,…Official notice →
  4. US
    Drugs & medicalClass II

    Siemens Medical Solutions USA, Inc recalls Artis Icono, Interventional Fluoroscopic X-Ray System, Model No. 11327600

    During system tests, an increased wearing of the Image acquisition system fans has been observed. This could lead to an electrical failure which may cause malfunction of the image system during regular system operation.

    FDA · DeviceSiemens Medical Solutions USA, IncDomestic distribution to AL, AR, AZ, CA, CO, CT, DE, FL, ID, IL, IN, KY, LA,…Official notice →
  5. US
    Drugs & medicalClass II

    Siemens Medical Solutions USA, Inc recalls syngo Application software VE20 (Material Number 10848815) installed on the Artis pheno and Artis icono systems as follows:…

    After CT image data from Toshiba is loaded, image mirroring can occur along the horizontal and vertical image axes. If this error occurs, the patient orientation/position may be misinterpreted and result in…

    FDA · DeviceSiemens Medical Solutions USA, IncUS Nationwide.Official notice →
  6. US
    Drugs & medicalClass II

    Rayner Intraocular Lenses Ltd recalls RayOne Preloaded Hydrophilic Acrylic IOL Injection Systems containing one MICS injection system with Intraocular Lens: (1) Outer…

    The outer package is mislabeled and the package contains a different IOL strength.

    FDA · DeviceRayner Intraocular Lenses LtdUS Nationwide distribution in the states of AZ, CA, FL, GA, IN, MI, MN, MO,…Official notice →
  7. US
    Drugs & medicalClass II

    Chromsystems Instruments & Chemicals GmbH recalls MassCheck Amino Acid, Acylcarnitines Dried Blood Spot Control Level I

    Too low concentration of glycine was detected in the dried blood controls of the batch 2821 of the products.

    FDA · DeviceChromsystems Instruments & Chemicals GmbHUS: NY, OH, CT.Official notice →