Drugs & medicalClass II

SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION recalls Trinias unity, Digital Angiography System with Catheterization Table, Model: KS-100 This system consists of the…

If digital angiography system is used with a specific catheterization table and the operator directs simultaneous movement of the table and the C-arm using direct memory, then direct memory button is released, the…

Read the official notice →
Date
2024-10-30
Country
United States
Brand
Shimadzu Corporation Medical Systems Division
Product
Trinias unity, Digital Angiography System with Catheterization Table, Model: KS-100 This system consists of the following main components:…
Classification
Class II
Units
US Nationwide distribution in the states of MA, IL, SC, KY, SC, OH, TN.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Shimadzu Corporation Medical Systems Division recalls

  1. 2025-11-19
  2. US
    Drugs & medicalClass II

    SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION recalls Ceiling Type X-Ray Tube Support Models CH-200 and CH-200M used with RADspeed, SonialVision and FLUOROspeed Systems…

    Due to an issue (potential breakage) with the shaft for the mounting of the X-ray tube assembly that could potentially result assembly coming off the rotating shaft.

    FDA · DeviceSHIMADZU CORPORATION MEDICAL SYSTEMS DIVISIONWorldwide - U.S. Nationwide distribution including in the states of AK, AL,…Official notice →
  3. 2024-11-13
  4. US
    Drugs & medicalClass II

    SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION recalls SONIALVISION G4 is a multi-purpose X-ray R/F system. The equipment is suitable for radiographic and fluoroscopic…

    X-ray R/F system software with certain versions of the control board may occasionally fail to display images when switching from Fluoroscopy mode to Radiography mode, which may lead to the possibility of delay or…

    FDA · DeviceSHIMADZU CORPORATION MEDICAL SYSTEMS DIVISIONUS Nationwide distribution in the states of CA, CT, MA, WV, OH, NY, IA, NE,…Official notice →
  5. 2024-03-20
  6. US
    Drugs & medicalClass II

    SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION recalls Shimadzu, Trinias, Digital Angiography System

    Due to capacitor manufacturing issue, their is a potential that Flat Panel Device may have abnormal display or no function.

    FDA · DeviceSHIMADZU CORPORATION MEDICAL SYSTEMS DIVISIONUS Nationwide distribution in the states of OH.Official notice →
  7. 2023-08-09
  8. US
    Drugs & medicalClass II

    SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION recalls Over-Head Ceiling Tube Support, CH-200, CH-200M, Used with RadSpeed and RadSpeed PRO, Sonial Vision safire, Versa…

    Eight fixing bolts at the base of the X-ray ceiling tube column assembly may be loose or missing so service is required to install a cable, bracket, and label behind the bracket, indicating when service is required, to…

    FDA · DeviceSHIMADZU CORPORATION MEDICAL SYSTEMS DIVISIONU.S. Nationwide distribution.Official notice →
All 5 Shimadzu Corporation Medical Systems Division recalls →

Other drugs & medical recalls in the United States

  1. 2024-10-30
  2. US
    Drugs & medicalClass II

    MEDLINE INDUSTRIES, LP - Northfield recalls Medline brand medical procedure convenience kits, labeled as: 1) THORACENTESIS TRAY, REF 00-400616J; 2) HH PACK SONICATION…

    XXX

    FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldWorldwide distribution - US Nationwide and the countries of Canada, Panama.Official notice →
  3. US
    Drugs & medicalClass II

    Smiths Medical ASD Inc. recalls smiths medical portex, 'ORATOR' SPEAKING VALVE FOR TRACHEOSTOMY TUBE, REF 100/550/000

    Uncertainty in the seal integrity of the sterile packaging.

    FDA · DeviceSmiths Medical ASD Inc.Worldwide distribution.Official notice →
  4. US
    Drugs & medicalClass II

    Physio-Control, Inc. recalls STERILIZABLE INTERNAL DEFIBRILLATION PADDLES REF LP20 STERILIZABLE INTERNAL DEFIBRILLATION PADDLES REF LP15 Defibrillation is indicated…

    Due to customer complaints and out of box failures, the firm is conducting a voluntary recall as a result of product failures (cracks and over-molding separation of the sterilizable internal defibrillation paddles.

    FDA · DevicePhysio-Control, Inc.Worldwide - U.S. Nationwide distribution including in the states of AL, AR,…Official notice →
  5. US
    Drugs & medicalClass II

    Mckesson Medical-Surgical Inc. Corporate Office recalls Quantimetrix Urine Chemistry Control Set Dipper Urinalysis Dipstick Testing Levels 1 & 2 6 X 15 mL-IVD intended…

    The Quantimetrix Dipper Urinalysis Dipstick control was shipped without the validated summer packaging and may have been exposed to temperatures above 25¿C. This may lead to delay in patients' results.

    FDA · DeviceMckesson Medical-Surgical Inc. Corporate OfficeUS Nationwide distribution in the state of DC.Official notice →
  6. US
    Drugs & medicalClass II

    Medtronic Neuromodulation recalls SynchroMed II Infusion Pump. Product Number: 8637-20

    Potential for error and inability to interrogate pump due to data corruption in the pump memory.

    FDA · DeviceMedtronic NeuromodulationUS Nationwide distribution in the state of Mississippi.Official notice →
  7. US
    Drugs & medicalClass II

    B-K Medical A/S recalls bk3000, model #2300-11; bk3500, model #2300-41; bk5000, model #2300-61; and bkActive 2300, model #2300-66, ultrasound system with battery…

    The internal power connections at the power inlet of the ultrasound devices can come loose resulting in the potential for excessive heat which could lead to a thermal event or the device becoming inoperable.

    FDA · DeviceB-K Medical A/SWorldwide - US Nationwide - There was also government/military distribution…Official notice →
See all drugs & medical recalls →