Drugs & medicalClass II

B-K Medical A/S recalls bk3000, model #2300-11; bk3500, model #2300-41; bk5000, model #2300-61; and bkActive 2300, model #2300-66, ultrasound system with battery…

The internal power connections at the power inlet of the ultrasound devices can come loose resulting in the potential for excessive heat which could lead to a thermal event or the device becoming inoperable.

Read the official notice →
Date
2024-10-30
Country
United States
Brand
B-K Medical A/S
Product
bk3000, model #2300-11; bk3500, model #2300-41; bk5000, model #2300-61; and bkActive 2300, model #2300-66, ultrasound system with battery…
Classification
Class II
Units
Worldwide - US Nationwide - There was also government/military distribution…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More B-K Medical A/S recalls

  1. 2025-11-12
  2. US
    Drugs & medicalClass II

    B-K Medical A/S recalls Ultrasound System 1300, bkSpecto w/battery; Model No. 1300-21 or 1300-S1; System, Imaging, Pulsed Doppler, Ultrasonic

    GE HealthCare has become aware that the Instructions for Use within the user guide and service manual do not adequately state the required battery replacement intervals, end-of-life handling procedures, or general…

    FDA · DeviceB-K Medical A/SWorldwide distribution: US (nationwide) and OUS (foreign) to countries of:…Official notice →
  3. 2024-01-31
  4. US
    Drugs & medicalClass II

    B-K Medical A/S recalls bkActiv Ultrasound System 2300, Model numbers 2300-56 and 2300-66.

    An issue on the Dual Live Compare feature on the bkActiv system could result in a measurement error.

    FDA · DeviceB-K Medical A/SUS Nationwide distribution including in the states of AL, AZ, CA, FL, GA, ID,…Official notice →
All 3 B-K Medical A/S recalls →

Other drugs & medical recalls in the United States

  1. 2024-10-30
  2. US
    Drugs & medicalClass II

    SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION recalls Trinias unity, Digital Angiography System with Catheterization Table, Model: KS-100 This system consists of the…

    If digital angiography system is used with a specific catheterization table and the operator directs simultaneous movement of the table and the C-arm using direct memory, then direct memory button is released, the…

    FDA · DeviceSHIMADZU CORPORATION MEDICAL SYSTEMS DIVISIONUS Nationwide distribution in the states of MA, IL, SC, KY, SC, OH, TN.Official notice →
  3. US
    Drugs & medicalClass II

    Mckesson Medical-Surgical Inc. Corporate Office recalls Quantimetrix Urine Chemistry Control Set Dipper Urinalysis Dipstick Testing Levels 1 & 2 6 X 15 mL-IVD intended…

    The Quantimetrix Dipper Urinalysis Dipstick control was shipped without the validated summer packaging and may have been exposed to temperatures above 25¿C. This may lead to delay in patients' results.

    FDA · DeviceMckesson Medical-Surgical Inc. Corporate OfficeUS Nationwide distribution in the state of DC.Official notice →
  4. US
    Drugs & medicalClass II

    Medtronic Neuromodulation recalls SynchroMed II Infusion Pump. Product Number: 8637-20

    Potential for error and inability to interrogate pump due to data corruption in the pump memory.

    FDA · DeviceMedtronic NeuromodulationUS Nationwide distribution in the state of Mississippi.Official notice →
  5. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls Soltive Premium SuperPulsed Laser - Intended for incision, excision, resection, ablation, coagulation, hemostasis, and…

    The term Bladder Stone was incorrectly translated in both Spanish and Portuguese to Kidney Stone(Clculo renal) on the systems Graphical User Interface (GUI), a potential to deliver an incorrect or unintended amount of…

    FDA · DeviceOlympus Corporation of the AmericasWorldwide distribution - US Nationwide and the countries of AU, CA, CL, DE,…Official notice →
  6. US
    Drugs & medicalClass II

    MEDLINE INDUSTRIES, LP - Northfield recalls (1) Centurion Sterile CirClamp, Reorder #200CR; (2) Centurion CirClamp with 1.1cm Bell and Insert, Reorder #210CR, Sterile;…

    Product has been identified as having holes in the packaging.

    FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldWorldwide distribution - US Nationwide and the countries of India and Panama.Official notice →
  7. US
    Drugs & medicalClass I

    Datex-Ohmeda, Inc. recalls Aisys CS2 with Et Control. This anesthesia gas machine is intended to provide general inhalation anesthesia and ventilatory support to…

    Potential issue that may arise if the limb of a patient breathing circuit is mistakenly connected to the Auxiliary Common Gas Outlet (ACGO) port.

    FDA · DeviceDatex-Ohmeda, Inc.Worldwide distribution - US Nationwide and the countries of Aland Islands,…Official notice →