RAYSEARCH LABORATORIES AB recalls RayStation/RayPlan- designed for treatment planning and analysis of radiation therapy. Software Version: 6.0.0.24 , 6.1.0.26 , 6.1.1.2…
If a new primary image set is selected while the cine loop is running, the primary image set will be displayed as both primary and secondary image set in all side-by-side views. This will also be true for any new…
Read the official notice →- Date
- 2022-03-16
- Country
- United States
- Brand
- Raysearch Laboratories Ab
- Product
- RayStation/RayPlan- designed for treatment planning and analysis of radiation therapy. Software Version: 6.0.0.24 , 6.1.0.26 , 6.1.1.2 ,…
- Classification
- Class II
- Units
- US Nationwide distribution.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Raysearch Laboratories Ab recalls
- 2023-09-06
US
RAYSEARCH LABORATORIES AB recalls RayCare software, include version number 5A, 5B, 6A, including service packs-An oncology information system used to support workflows…
An issue was found in Software RayCare 5A, 5B, 6A, including service packs, where an allergy warning against medication substance (drug ingredient) will not be displayed as expected under certain circumstances
FDA · DeviceRAYSEARCH LABORATORIES ABUS Nationwide distribution in the state of TN.Official notice →- 2023-03-29
US
RAYSEARCH LABORATORIES AB recalls Raystation treatment planning system for radiation therapy, Model Numbers 9.0.0, 9.0.1, 9.1.0, 9.2.0, 10.0.0, 10.0.1, 10.0.2, 10.1.0,…
It is possible to set a non-zero collimator angle in the Virtual Simulation module. The angle will be correct in the user interface but if the plan is DICOM exported by using the VSIM Export button, the angle will not…
- 2022-07-27
US
RAYSEARCH LABORATORIES AB recalls RayStation 9B SP1. For radiation treatment planning.
An issue with propagation of treatment course information from RayStation with RayTreat to RayCare has been identified. During some workflows, information may not be propagated to RayCare.
FDA · DeviceRAYSEARCH LABORATORIES ABWorldwide distribution - US Nationwide distribution in the state of TN and the…Official notice →- 2022-07-06
US
RAYSEARCH LABORATORIES AB recalls RayStation 11B
This notice concerns two issues found related to the display of Linear Energy Transfer (LET) in RayStation 11B including some service packs. First, when using a dose threshold for an evaluation dose, LET display might…
FDA · DeviceRAYSEARCH LABORATORIES ABUS Nationwide distribution in the states of NJ and TX.Official notice →- 2022-04-06
US
RAYSEARCH LABORATORIES AB recalls RayStation/RayPlan versions 4 - 11A, including service packs. Build Stations: 4.0.0.14, 4.0.3.4, 4.3.0.14, 4.5.1.14, 4.7.2.5, 4.7.3.13,…
An issue where the combined density in a dose grid voxel partially covered by the External ROI and also partially covered by an ROI of type Bolus, Support or Fixation may be unexpected.The density in the voxel can be…
US
RAYSEARCH LABORATORIES AB recalls TomoHelical and TomoDirect plans in RayStation 6, 7, 8A, 8B, 9A, 9B, 10A, 10B and 11A. Build Numbers: 6.1.1.2, 6.2.0.7, 6.3.0.6,…
If the dynamic jaw mode is used and an optimization is continued after changing the dose grid or modifying the target ROI, the positions of the jaws may change unexpectedly for some control points. User must be aware…
FDA · DeviceRAYSEARCH LABORATORIES ABUS Nationwide distribution in the states of AL, CA, FL, GA, IA, IL, MO, MS,…Official notice →
Other drugs & medical recalls in the United States
- 2022-03-16
US
Bolder Surgical, LLC recalls CoolSeal Generator, REF: CSL-200-50
A potential manufacturing/assembly defect could result in the generator not meeting IEC 60601 standards for medical electrical equipment, and it can potentially, under specific conditions, lead to the delivery of…
US
Ortho-Clinical Diagnostics, Inc. recalls VITROS Chemistry Products VersaTips, Catalog No. 6801715 A VersaTip sales unit is composed of 1000 tips per box.
The lot has a shape deformity which may lead to a 16% reduction in the volume of the sample or reagent dispensed. In cases where the error is not detected and the affected results are not suppressed, clinical impact…
FDA · DeviceOrtho-Clinical Diagnostics, Inc.Domestic distribution nationwide. Foreign distribution to Canada.Official notice →US
Brainlab AG recalls ExacTrac Dynamic software model numbers: 20910-01, ETD Positioning and Moni. 1.0.0 INST; 20910-01A, ETD Positioning and Moni. 1.0.1 INST;…
The yaw angle may be incorrect for CBCT positioning workflows using setup beams with Varian LINACs.
FDA · DeviceBrainlab AGDistribution was made to AZ, AR, CA, CO, FL, GA, IN, KY, MD, MS, MO, NJ, NY,…Official notice →US
CELLTRION USA INC recalls Celltrion DiaTrust COVID-19 Ag Rapid Test, Reference No. CT-P60 D-2 02
Kits were labeled with incorrect expiration dates.
FDA · DeviceCELLTRION USA INCThe products were distributed to the following US states: CA, CT, FL, HI, MA,…Official notice →US
Maquet Cardiovascular, LLC recalls HEMASHIELD GOLD KNITTED MICROVEL DOUBLE VELOUR VASCULAR GRAFT- D: 7 mm; L: 60 cm Indicated for use in the replacement or repair of…
Outer labelling (HEMASHIELD GOLD Knitted Microvel Double Velour Bifurcated Vascular Graft, D: 16 x 8 mm; L: 40 cm) does not correspond to the (inner) sterile packaging labelling (HEMASHIELD GOLD Knitted Microvel Double…
FDA · DeviceMaquet Cardiovascular, LLCUS Nationwide distribution in the state of MO.Official notice →US
Stryker Corporation recalls 1688 Camera Control Unit (CCU), Catalog number:1688010000. Used in general laparoscopy, nasopharyngoscopy, ear endoscopy, sinuscopy, and…
A software defect in the camera control unit (CCU) will cause the image on the monitor to flip upside-down into an incorrect orientation. The potential of harms are conversion to open procedure, additional medical…
FDA · DeviceStryker CorporationWorldwide distribution. US nationwide including Puerto Rico. OUS: BR, AE, NL,…Official notice →