Drugs & medicalClass II

RAYSEARCH LABORATORIES AB recalls RayStation 9B SP1. For radiation treatment planning.

An issue with propagation of treatment course information from RayStation with RayTreat to RayCare has been identified. During some workflows, information may not be propagated to RayCare.

Read the official notice →
Date
2022-07-27
Country
United States
Brand
Raysearch Laboratories Ab
Product
RayStation 9B SP1. For radiation treatment planning.
Classification
Class II
Units
Worldwide distribution - US Nationwide distribution in the state of TN and the…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Raysearch Laboratories Ab recalls

  1. 2023-11-08
  2. US
    Drugs & medicalClass II

    RAYSEARCH LABORATORIES AB recalls RayCare 5B and 6A including service packs- An oncology information system used to support workflows and scheduling, as well as clinical…

    An issue was found in RayCare 5B, 6A, including service packs, where values will not be visible for documents in a read-only state when opened outside the Documents workspace when using document fields configured with…

    FDA · DeviceRAYSEARCH LABORATORIES ABMD, TNOfficial notice →
  3. 2023-09-06
  4. US
    Drugs & medicalClass II

    RAYSEARCH LABORATORIES AB recalls RayCare software, include version number 5A, 5B, 6A, including service packs-An oncology information system used to support workflows…

    An issue was found in Software RayCare 5A, 5B, 6A, including service packs, where an allergy warning against medication substance (drug ingredient) will not be displayed as expected under certain circumstances

    FDA · DeviceRAYSEARCH LABORATORIES ABUS Nationwide distribution in the state of TN.Official notice →
  5. 2023-03-29
  6. US
    Drugs & medicalClass II

    RAYSEARCH LABORATORIES AB recalls Raystation treatment planning system for radiation therapy, Model Numbers 9.0.0, 9.0.1, 9.1.0, 9.2.0, 10.0.0, 10.0.1, 10.0.2, 10.1.0,…

    It is possible to set a non-zero collimator angle in the Virtual Simulation module. The angle will be correct in the user interface but if the plan is DICOM exported by using the VSIM Export button, the angle will not…

    FDA · DeviceRAYSEARCH LABORATORIES ABUS NationwideOfficial notice →
  7. 2022-07-06
  8. US
    Drugs & medicalClass II

    RAYSEARCH LABORATORIES AB recalls RayStation 11B

    This notice concerns two issues found related to the display of Linear Energy Transfer (LET) in RayStation 11B including some service packs. First, when using a dose threshold for an evaluation dose, LET display might…

    FDA · DeviceRAYSEARCH LABORATORIES ABUS Nationwide distribution in the states of NJ and TX.Official notice →
  9. 2022-04-06
  10. US
    Drugs & medicalClass II

    RAYSEARCH LABORATORIES AB recalls RayStation/RayPlan versions 4 - 11A, including service packs. Build Stations: 4.0.0.14, 4.0.3.4, 4.3.0.14, 4.5.1.14, 4.7.2.5, 4.7.3.13,…

    An issue where the combined density in a dose grid voxel partially covered by the External ROI and also partially covered by an ROI of type Bolus, Support or Fixation may be unexpected.The density in the voxel can be…

    FDA · DeviceRAYSEARCH LABORATORIES ABUS Nationwide.Official notice →
  11. US
    Drugs & medicalClass II

    RAYSEARCH LABORATORIES AB recalls TomoHelical and TomoDirect plans in RayStation 6, 7, 8A, 8B, 9A, 9B, 10A, 10B and 11A. Build Numbers: 6.1.1.2, 6.2.0.7, 6.3.0.6,…

    If the dynamic jaw mode is used and an optimization is continued after changing the dose grid or modifying the target ROI, the positions of the jaws may change unexpectedly for some control points. User must be aware…

    FDA · DeviceRAYSEARCH LABORATORIES ABUS Nationwide distribution in the states of AL, CA, FL, GA, IA, IL, MO, MS,…Official notice →
All 14 Raysearch Laboratories Ab recalls →

Other drugs & medical recalls in the United States

  1. 2022-07-27
  2. US
    Drugs & medicalClass II

    ROi CPS LLC recalls Nonsterile convenience kits: (1) regard Item #830099008, CV01005H - CV Part 1 Bundle; and (2) regard Item #830019017, CV00214Q - CV.

    Some nonsterile components that were supposed to be sterile were packaged in nonsterile kits.

    FDA · DeviceROi CPS LLCDistribution was made to MO. There was no foreign/military/government…Official notice →
  3. US
    Drugs & medicalClass II

    Aesculap Implant Systems LLC recalls RIGHT CORONAL BENDER, TEK1940. Instrument used during spinal surgery.

    Customized coronal rod benders may deform the implant when used with the spinal system devices.

    FDA · DeviceAesculap Implant Systems LLCUnited States: COOfficial notice →
  4. US
    Drugs & medicalClass II

    Xstrahl Limited recalls X80 RADiant Photoelectric Therapy System

    There is a potential compatibility issue with the systems and replacement treatment applicators.

    FDA · DeviceXstrahl LimitedUS Distribution FL, CA, KYOfficial notice →
  5. US
    Drugs & medicalClass II

    Stryker Instruments Div. of Stryker Corporation recalls Triton Canister System (finished part numbers FG 12009, for 3L canisters, and FG 12019 for 2L canisters)…

    Some Canister Scanning Labels used to calibrate the Triton Canister System have a color variance. Based on a comprehensive analysis, blood loss estimates obtained using these Canister Scanning Labels may result in…

    FDA · DeviceStryker Instruments Div. of Stryker CorporationUS NationwideOfficial notice →
  6. US
    Drugs & medicalClass II

    Surgical Innovations Ltd recalls YelloPort Elite Universal Seal. For use in laparoscopic procedures.

    There is the potential that YelloPort Elite Universal Seals packaging may contain small holes which may affect the sterility of the product.

    FDA · DeviceSurgical Innovations LtdWorldwide distribution - US Nationwide distribution in the states of FL, IN,…Official notice →
  7. US
    Drugs & medicalClass III

    Stryker Instruments Div. of Stryker Corporation recalls Triton Sponge System (Software), GAUS-3, Software Versions 4.6.5, 4.8.7, 4.9.6

    Gauss Surgical is updating the Instructions for Use (IFU) for our Triton Sponge System in order to clarify proper technique for handling sponges and the calibration placard.

    FDA · DeviceStryker Instruments Div. of Stryker CorporationUS Nationwide.Official notice →
See all drugs & medical recalls →