RAYSEARCH LABORATORIES AB recalls RayStation 13.0.0.1547, 13.1.0.144 & 13.1.1.89. Radiation Therapy Treatment Planning System.
Potential for reported SSD to be too high.
Read the official notice →- Date
- 2024-05-08
- Country
- United States
- Brand
- Raysearch Laboratories Ab
- Product
- RayStation 13.0.0.1547, 13.1.0.144 & 13.1.1.89. Radiation Therapy Treatment Planning System.
- Classification
- Class II
- Units
- Worldwide distribution - US Nationwide and the countries of Algeria,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Raysearch Laboratories Ab recalls
- 2026-01-28
US
RAYSEARCH LABORATORIES AB recalls Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 17.0.0, 17.0.1 Software Version: RayStation v2025,…
Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material…
FDA · DeviceRAYSEARCH LABORATORIES ABRefer to the attached documents: Distribution list 167168 US - updated,…Official notice →- 2025-09-10
US
RAYSEARCH LABORATORIES AB recalls RayStation with the following product descriptions: 1. RayStation 11B, Software Version: 12.0.0.932. 2. RayStation 11B Service Pack 1,…
DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.
FDA · DeviceRAYSEARCH LABORATORIES ABWorldwide - US Nationwide distribution in the states of AL, CA, FL, GA, IL,…Official notice →- 2025-02-26
US
RAYSEARCH LABORATORIES AB recalls Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 1.0.0.951, 11.0.1.29,…
Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT…
FDA · DeviceRAYSEARCH LABORATORIES ABWorldwide distribution - US Nationwide and the countries of Australia,…Official notice →- 2024-04-17
US
RAYSEARCH LABORATORIES AB recalls RayStation Software Version Numbers version numbers 8B, 9A, 9B, 10A, 10B, 11A, 11B, 12A, 12B, 2023B, 2024A including some service packs.
A use error that has occurred with RayStation/RayPlan, where material override using Silicon, Si, was incorrectly selected for a silicone gel polymer implant.
FDA · DeviceRAYSEARCH LABORATORIES ABWorldwide distribution - US Nationwide and the countries of Australia,…Official notice →- 2023-11-08
US
RAYSEARCH LABORATORIES AB recalls RayCare 5B and 6A including service packs- An oncology information system used to support workflows and scheduling, as well as clinical…
An issue was found in RayCare 5B, 6A, including service packs, where values will not be visible for documents in a read-only state when opened outside the Documents workspace when using document fields configured with…
- 2023-09-06
US
RAYSEARCH LABORATORIES AB recalls RayCare software, include version number 5A, 5B, 6A, including service packs-An oncology information system used to support workflows…
An issue was found in Software RayCare 5A, 5B, 6A, including service packs, where an allergy warning against medication substance (drug ingredient) will not be displayed as expected under certain circumstances
FDA · DeviceRAYSEARCH LABORATORIES ABUS Nationwide distribution in the state of TN.Official notice →
Other drugs & medical recalls in the United States
- 2024-05-08
US
Johnson & Johnson Surgical Vision, Inc. recalls Johnson & Johnson VISION Reusable LAMINAR High Flow Irrigation Sleeve and Test Chamber 21 Gauge REF OPOHF21L The…
Due to a manufacturing variation of the irrigation sleeves which could result in a missing port hole, this non-conformity could lead to insufficient flow to cool the ultrasonic phaco tip, potentially leading to an…
FDA · DeviceJohnson & Johnson Surgical Vision, Inc.U.S. Nationwide distribution in the states of AZ, CA, CO, FL, GA, IL, LA, MA,…Official notice →US
SEDECAL SA recalls Battery Mobile X-ray Unit EASY MOVING Model with telescopic column 3.1, model numbers: a. SM-20HF-Batt; b. SM-32HF-Batt; c. SM-40HF-Batt;…
Potential interference with the anti-fall system.
FDA · DeviceSEDECAL SAWorldwide distribution - US Nationwide in the states of CA, IL, NJ, and areas…Official notice →US
Innomed, Inc. recalls INNOMED CUP REMOVAL STARTER INSTRUMENT, REF 520054, Hip cutter blade cup
54 MM finisher blade was inadvertently affixed to the shaft of the CupX handle in lieu of the 54 MM starter blade.
FDA · DeviceInnomed, Inc.FL, GA, PA, NY, CA, VA, SD, MN, TX, UT, NC, and UK, Denmark, Germany…Official notice →US
Stryker Neurovascular recalls The Distal Access Catheter is a single-lumen, braided shaft, variable stiffness catheter with radiopaque marker on the distal end and a…
Stryker Neurovascular is recalling their DAC- Distal Access Catheter -038 136cm, a Distal Access Catheter by removal. The reason for the recall is: the DAC Lot #0000486382 was released with out-of-specification…
FDA · DeviceStryker NeurovascularUS Nationwide distribution in the states of TX, FL, NJ, CO, LA, NC, AL, PA,…Official notice →US
Cardinal Health 200, LLC recalls Jackson-Pratt 3-Spring Reservoir Kit with PVC Round Drain, 400mL, 7Fr., REF SU130-401D
Cardinal Health is issuing a lot number specific voluntary removal of Jackson-Pratt 3-Spring Reservoir kits due to specific products being shipped to users before undergoing sterilization.
FDA · DeviceCardinal Health 200, LLCWorldwide distribution - US Nationwide and the countries of Panama, Guatemala,…Official notice →US
Noah Medical recalls Galaxy System REF GAL-001 (specifically GAL-019, Wired Controller) The Galaxy System and its accessories are intended to provide…
Due to a misaligned calibration of the wired controller that may result in unintended motion of the scope.
FDA · DeviceNoah MedicalU.S. Nationwide distribution in the states of CA, IN, MD, MO, OH, and PA.Official notice →