Cardinal Health 200, LLC recalls Jackson-Pratt 3-Spring Reservoir Kit with PVC Round Drain, 400mL, 7Fr., REF SU130-401D
Cardinal Health is issuing a lot number specific voluntary removal of Jackson-Pratt 3-Spring Reservoir kits due to specific products being shipped to users before undergoing sterilization.
Read the official notice →- Date
- 2024-05-08
- Country
- United States
- Brand
- Cardinal Health
- Product
- Jackson-Pratt 3-Spring Reservoir Kit with PVC Round Drain, 400mL, 7Fr., REF SU130-401D
- Classification
- Class II
- Units
- Worldwide distribution - US Nationwide and the countries of Panama, Guatemala,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Cardinal Health recalls
- 2024-10-02
US
Cardinal Health 200, LLC recalls (a) CardinalHealth Salem Sump Dual Lumen Stomach Tube with Anti-Reflux Valve, 12 Fr/Ch (4.0 mm) x 48 in. (122 cm), Model #8888266114,…
The firm received reports of breakage in the Anti-Reflux Valve (ARV) due to improper use. As a result, changes to labeling were made.
FDA · DeviceCardinal Health 200, LLCDistribution was nationwide, including Puerto Rico and Guam. There was also…Official notice →- 2024-07-17
US
Cardinal Health 200, LLC recalls Presource kits and trays, labeled as: 1) Cardiovascular Procedure Kit, Catalog Number SAN29ANCC3; 2) Cardiovascular Procedure Kit,…
Cardinal Health is following the FDA's recommendations regarding the disposition of affected Jiangsu Shenli Medical Production Co. Ltd (Shenli) Syringes in the Merit Medical¿ supplied finished kits. Merit Medical¿…
- 2024-05-22
US
Cardinal Health 200, LLC recalls CardinalHealth Curity Flexible Adhesive Bandage X-Large 2 x 3.75 in. (5.1 x 9.5 cm), Model 44102, Sterile, Hypoallergenic.
Potential contamination of the product with latex adhesive residual.
- 2024-02-28
US
Cardinal Health 200, LLC recalls Dover" IRRIGATION 1200 CC TRAY, 60 CC PISTON SYRINGE, STERILE SALINE, TIP PROTECTOR, ALCOHOL, LIDDED
Convenience kits were manufactured and distributed with components which were subsequently recalled by Nurse Assist.
- 2024-02-14
US
Cardinal Health 200, LLC recalls Cardinal Health Monoject, 1 mL Purple Enteral Syringe with Enfit Connection Sterile, REF 401SE
Cardinal Health is expanding their previous product correction actions to a product removal of all sizes of Cardinal Health brand Monoject sterile Syringe Luer-Lock (1, 3, 6, 12, 20, 35, and 60 mL) and Cardinal Health…
FDA · DeviceCardinal Health 200, LLCDistribution US nationwide and Canada. Foreign distribution updated 9/6/2024…Official notice →US
Cardinal Health 200, LLC recalls Cardinal Health Monoject, 35mL Enteral Syringe (purple) - For Use with ENFit Connection System, REF 435SE
The listed lots of Cardinal Health Monoject Enteral Syringes with ENFit (6, 12, 35, 60 mL) have recognition and compatibility issues with certain syringe enteral feeding pumps. As a result, Cardinal Health recommends…
FDA · DeviceCardinal Health 200, LLCDistribution US nationwide and Canada. Foreign distribution updated 9/6/2024…Official notice →
Other drugs & medical recalls in the United States
- 2024-05-08
US
Innomed, Inc. recalls INNOMED CUP REMOVAL STARTER INSTRUMENT, REF 520054, Hip cutter blade cup
54 MM finisher blade was inadvertently affixed to the shaft of the CupX handle in lieu of the 54 MM starter blade.
FDA · DeviceInnomed, Inc.FL, GA, PA, NY, CA, VA, SD, MN, TX, UT, NC, and UK, Denmark, Germany…Official notice →US
RAYSEARCH LABORATORIES AB recalls RayStation 13.0.0.1547, 13.1.0.144 & 13.1.1.89. Radiation Therapy Treatment Planning System.
Potential for reported SSD to be too high.
FDA · DeviceRAYSEARCH LABORATORIES ABWorldwide distribution - US Nationwide and the countries of Algeria,…Official notice →US
Stryker Neurovascular recalls The Distal Access Catheter is a single-lumen, braided shaft, variable stiffness catheter with radiopaque marker on the distal end and a…
Stryker Neurovascular is recalling their DAC- Distal Access Catheter -038 136cm, a Distal Access Catheter by removal. The reason for the recall is: the DAC Lot #0000486382 was released with out-of-specification…
FDA · DeviceStryker NeurovascularUS Nationwide distribution in the states of TX, FL, NJ, CO, LA, NC, AL, PA,…Official notice →US
Noah Medical recalls Galaxy System REF GAL-001 (specifically GAL-019, Wired Controller) The Galaxy System and its accessories are intended to provide…
Due to a misaligned calibration of the wired controller that may result in unintended motion of the scope.
FDA · DeviceNoah MedicalU.S. Nationwide distribution in the states of CA, IN, MD, MO, OH, and PA.Official notice →US
Covidien recalls Medtronic Invos, Reusable Infant Sensor Adapter Cable for PM7100, REF: PMAC71RIC
manufacturing defects to sensor cable connectors may result in error codes or non-functioning channel
FDA · DeviceCovidienUS: AK AZ CA CT FL GA IA IN LA MA MD MI MN MO NC NY OH OK PA SC TN TX VA WA WI…Official notice →US
Zimmer, Inc. recalls ZPLP(Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Left, 6 Holes, 106 MM, Sterile- Indicated for temporary…
Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock,…
FDA · DeviceZimmer, Inc.Worldwide distribution - US Nationwide and the countries of Brazil, Canada,…Official notice →